Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where A Ghebremedhin is active.

Publication


Featured researches published by A Ghebremedhin.


Medical Physics | 1997

Microdosimetry spectra of the Loma Linda proton beam and relative biological effectiveness comparisons

G. Coutrakon; J. Cortese; A Ghebremedhin; J. Hubbard; J. Johanning; P. Koss; G. Maudsley; C. R. Slater; C. Zuccarelli; J. Robertson

Protons have long been recognized as low LET radiation in radiotherapy. However, a detailed account of LET (linear energy transfer) and RBE (relative biological effectiveness) changes with incident beam energy and depth in tissue is still unresolved. This issue is particularly important for treatment planning, where the physical dose prescription is calculated from a RBE using cobalt as the reference radiation. Any significant RBE changes with energy or depth will be important to incorporate in treatment planning. In this paper we present microdosimetry spectra for the proton beam at various energies and depths and compare the results to cell survival studies performed at Loma Linda. An empirically determined biological weighting function that depends on lineal energy is used to correlate the microdosimetry spectra with cell survival data. We conclude that the variations in measured RBE with beam energy and depth are small until the distal edge of the beam is reached. On the distal edge, protons achieve stopping powers as high as 100 keV/micron, which is reflected in the lineal energy spectra taken there. Lineal energy spectra 5 cm beyond the distal edge of the Bragg peak also show a high LET component but at a dose rate 600 times smaller than observed inside the proton field.


Medical Physics | 2007

Leakage and scatter radiation from a double scattering based proton beamline

Michael F. Moyers; E.R. Benton; A Ghebremedhin; G. Coutrakon

Proton beams offer several advantages over conventional radiation techniques for treating cancer and other diseases. These advantages might be negated if the leakage and scatter radiation from the beamline and patient are too large. Although the leakage and scatter radiation for the double scattering proton beamlines at the Loma Linda University Proton Treatment Facility were measured during the acceptance testing that occurred in the early 1990s, recent discussions in the radiotherapy community have prompted a reinvestigation of this contribution to the dose equivalent a patient receives. The dose and dose equivalent delivered to a large phantom patient outside a primary proton field were determined using five methods: simulations using Monte Carlo calculations, measurements with silver halide film, measurements with ionization chambers, measurements with rem meters, and measurements with CR-39 plastic nuclear track detectors. The Monte Carlo dose distribution was calculated in a coronal plane through the simulated patient that coincided with the central axis of the beam. Measurements with the ionization chambers, rem meters, and plastic nuclear track detectors were made at multiple locations within the same coronal plane. Measurements with the film were done in a plane perpendicular to the central axis of the beam and coincident with the surface of the phantom patient. In general, agreement between the five methods was good, but there were some differences. Measurements and simulations also tended to be in agreement with the original acceptance testing measurements and results from similar facilities published in the literature. Simulations illustrated that most of the neutrons entering the patient are produced in the final patient-specific aperture and precollimator just upstream of the aperture, not in the scattering system. These new results confirm that the dose equivalents received by patients outside the primary proton field from primary particles that leak through the nozzle are below the accepted standards for x-ray and electron beams. The total dose equivalent outside of the field is similar to that received by patients undergoing treatments with intensity modulated x-ray therapy. At the center of a patient for a whole course of treatment, the dose equivalent is comparable to that delivered by a single whole-body XCT scan.


Medical Physics | 2012

Water-equivalent path length calibration of a prototype proton CT scanner

R. F. Hurley; Reinhard W. Schulte; V. Bashkirov; A Wroe; A Ghebremedhin; H. Sadrozinski; V. Rykalin; G. Coutrakon; P. Koss; B Patyal

PURPOSE The authors present a calibration method for a prototype proton computed tomography (pCT) scanner. The accuracy of these measurements depends upon careful calibration of the energy detector used to measure the residual energy of the protons that passed through the object. METHODS A prototype pCT scanner with a cesium iodide (CsI(Tl)) crystal calorimeter was calibrated by measuring the calorimeter response for protons of 200 and 100 MeV initial energies undergoing degradation in polystyrene plates of known thickness and relative stopping power (RSP) with respect to water. Calibration curves for the two proton energies were obtained by fitting a second-degree polynomial to the water-equivalent path length versus calorimeter response data. Using the 100 MeV calibration curve, the RSP values for a variety of tissue-equivalent materials were measured and compared to values obtained from a standard depth-dose range shift measurement using a water-tank. A cylindrical water phantom was scanned with 200 MeV protons and its RSP distribution was reconstructed using the 200 MeV calibration. RESULTS It is shown that this calibration method produces measured RSP values of various tissue-equivalent materials that agree to within 0.5% of values obtained using an established water-tank method. The mean RSP value of the water phantom reconstruction was found to be 0.995 ± 0.006. CONCLUSIONS The method presented provides a simple and reliable procedure for calibration of a pCT scanner.


Medical Physics | 2007

Calibration of a proton beam energy monitor.

Michael F. Moyers; G. Coutrakon; A Ghebremedhin; K. Shahnazi; P. Koss; E. Sanders

Delivery of therapeutic proton beams requires an absolute energy accuracy of +/-0.64 to 0.27 MeV for patch fields and a relative energy accuracy of +/-0.10 to 0.25 MeV for tailoring the depth dose distribution using the energy stacking technique. Achromatic switchyard tunes, which lead to better stability of the beam incident onto the patient, unfortunately limit the ability of switchyard magnet tesla meters to verify the correct beam energy within the tolerances listed above. A new monitor to measure the proton energy before each pulse is transported through the switchyard has been installed into a proton synchrotron. The purpose of this monitor is to correct and/or inhibit beam delivery when the measured beam energy is outside of the tolerances for treatment. The monitor calculates the beam energy using data from two frequency and eight beam position monitors that measure the revolution frequency of the proton bunches and the effective offset of the orbit from the nominal radius of the synchrotron. The new energy monitor has been calibrated by measuring the range of the beam through water and comparing with published range-energy tables for various energies. A relationship between depth dose curves and range-energy tables was first determined using Monte Carlo simulations of particle transport and energy deposition. To reduce the uncertainties associated with typical scanning water phantoms, a new technique was devised in which the beam energy was scanned while fixed thickness water tanks were sandwiched between two fixed parallel plate ionization chambers. Using a multitude of tank sizes, several energies were tested to determine the nominal accelerator orbit radius. After calibration, the energy reported by the control system matched the energy derived by range measurements to better than 0.72 MeV for all nine energies tested between 40 and 255 MeV with an average difference of -0.33 MeV. A study of different combinations of revolution frequency and radial offsets to test the envelope of algorithm accuracy demonstrated a relative accuracy of +/-0.11 MeV for small energy changes between 126 and 250 MeV. These new measurements may serve as a data set for benchmarking range-energy relationships.


Medical Physics | 2013

Evaluation of the dosimetric properties of a synthetic single crystal diamond detector in high energy clinical proton beams

A Mandapaka; A Ghebremedhin; B Patyal; M. Marinelli; G. Prestopino; C. Verona; G. Verona-Rinati

PURPOSE To investigate the dosimetric properties of a synthetic single crystal diamond Schottky diode for accurate relative dose measurements in large and small field high-energy clinical proton beams. METHODS The dosimetric properties of a synthetic single crystal diamond detector were assessed by comparison with a reference Markus parallel plate ionization chamber, an Exradin A16 microionization chamber, and Exradin T1a ion chamber. The diamond detector was operated at zero bias voltage at all times. Comparative dose distribution measurements were performed by means of Fractional depth dose curves and lateral beam profiles in clinical proton beams of energies 155 and 250 MeV for a 14 cm square cerrobend aperture and 126 MeV for 3, 2, and 1 cm diameter circular brass collimators. ICRU Report No. 78 recommended beam parameters were used to compare fractional depth dose curves and beam profiles obtained using the diamond detector and the reference ionization chamber. Warm-up∕stability of the detector response and linearity with dose were evaluated in a 250 MeV proton beam and dose rate dependence was evaluated in a 126 MeV proton beam. Stem effect and the azimuthal angle dependence of the diode response were also evaluated. RESULTS A maximum deviation in diamond detector signal from the average reading of less than 0.5% was found during the warm-up irradiation procedure. The detector response showed a good linear behavior as a function of dose with observed deviations below 0.5% over a dose range from 50 to 500 cGy. The detector response was dose rate independent, with deviations below 0.5% in the investigated dose rates ranging from 85 to 300 cGy∕min. Stem effect and azimuthal angle dependence of the diode signal were within 0.5%. Fractional depth dose curves and lateral beam profiles obtained with the diamond detector were in good agreement with those measured using reference dosimeters. CONCLUSIONS The observed dosimetric properties of the synthetic single crystal diamond detector indicate that its behavior is proton energy independent and dose rate independent in the investigated energy and dose rate range and it is suitable for accurate relative dosimetric measurements in large as well as in small field high energy clinical proton beams.


Proceedings of the 1999 Particle Accelerator Conference (Cat. No.99CH36366) | 1999

Design considerations for medical proton accelerators

George Coutrakon; James M. Slater; A Ghebremedhin

The design requirements for current heavy-particle accelerators operated within a hospital to deliver radiation therapy must satisfy both clinical and research needs. Advances in dedicated beam delivery systems for clinical utilization and biological studies add requirements that previous accelerators did not have. Eight years experience using the Loma Linda University proton facility has emphasized that the most important requirements are safety, reliability, beam stability, low energy consumption, and efficiency of beam delivery to the treatment rooms. In the future, raster scanning techniques will add further demands on the control of beam energy, intensity, and position stability. Rapid and precise flexibility in changing beam parameters is essential for satisfying clinical needs; electronic rather than mechanical control is clearly preferable for clinical use. Biological research increases the need to expand the margins of some clinical requirements, such as beam size, intensity, and energy ranges. Both clinical and research activities require a totally integrated control system, beginning with the ion source and continuing through the accelerator and switch yard to multiple rooms and each beam delivery system therein. Accordingly, designing the clinical accelerator requires a highly orchestrated design effort, involving the entire facility. Detailed design requirements addressing these issues will be presented.


Technology in Cancer Research & Treatment | 2013

Water Equivalent Thickness Analysis of Immobilization Devices for Clinical Implementation in Proton Therapy

A Wroe; A Ghebremedhin; I. R. Gordon; Reinhard W. Schulte; Jerry D. Slater

Immobilization devices can impact not only the inter- and intra-fraction motion of the patient, but also the range uncertainty of the treatment beam in proton therapy. In order to limit additional range uncertainty, the water equivalent thickness (WET) of the immobilization device needs to be well known and accurately reflected in the calculations by the treatment planning system (TPS). The method presented here focusses on the use of a nozzle-mounted variable range shifter and precision-machined polystyrene blocks of known WET to evaluate commercial immobilization devices prior to clinical implementation. CT studies were also completed to evaluate the internal uniformity of the immobilization devices under study. Multiple inserts of the kVue platform (Qfix Systems, Avondale, PA) were evaluated as part of this study. The results indicate that the inserts are largely interchangeable across a given design type and that the measured WET values agree with those generated by the TPS with a maximum difference less than 1 mm. The WET of the devices, as determined by the TPS, was not impacted by CT beam hardening normally experienced during clinical use. The reproducibility of the WET method was also determined to be better than 60.02 mm. In conclusion, the testing of immobilization prior to implementation in proton therapy is essential in order to ascertain their impact on the proton treatment and the methodology described here can also be applied to other immobilization systems.


Radiation Oncology | 2013

Evaluation and comparison of New 4DCT based strategies for proton treatment planning for lung tumors

N Wang; B Patyal; A Ghebremedhin; David A. Bush

PurposeTo evaluate different strategies for proton lung treatment planning based on four-dimensional CT (4DCT) scans.Methods and MaterialsTwelve cases, involving only gross tumor volumes (GTV), were evaluated. Single image sets of (1) maximum intensity projection (MIP3) of end inhale (EI), middle exhale (ME) and end exhale (EE) images; (2) average intensity projection (AVG) of all phase images; and (3) EE images from 4DCT scans were selected as primary images for proton treatment planning. Internal target volumes (ITVs) outlined by a clinician were imported into MIP3, AVG, and EE images as planning targets. Initially, treatment uncertainties were not included in planning. Each plan was imported into phase images of 4DCT scans. Relative volumes of GTVs covered by 95% of prescribed dose and mean ipsilateral lung dose of a phase image obtained by averaging the dose in inspiration and expiration phases were used to evaluate the quality of a plan for a particular case. For comparing different planning strategies, the mean of the averaged relative volumes of GTVs covered by 95% of prescribed dose and its standard deviation for each planning strategy for all cases were used. Then, treatment uncertainties were included in planning. Each plan was recalculated in phase images of 4DCT scans. Same strategies were used for plan evaluation except dose-volume histograms of the planning target volumes (PTVs) instead of GTVs were used and the mean and standard deviation of the relative volumes of PTVs covered by 95% of prescribed dose and the ipsilateral lung dose were used to compare different planning strategies.ResultsMIP3 plans without treatment uncertainties yielded 96.7% of the mean relative GTV covered by 95% of prescribed dose (standard deviations of 5.7% for all cases). With treatment uncertainties, MIP3 plans yielded 99.5% of mean relative PTV covered by 95% of prescribed dose (standard deviations of 0.7%). Inclusion of treatment uncertainties improved PTV dose coverage but also increased the ipsilateral mean lung dose in general, and reduced the variations of the PTV dose coverage among different cases. Plans based on conventional axial CT scan (CVCT) gave the poorest PTV dose coverage (about 96% of mean relative PTV covered by 95% isodose) compared to MIP3 and EE plans, which resulted in 100% of PTV covered by 95% isodose for tumors with relatively large motion. AVG plans demonstrated PTV dose coverage of 89.8% and 94.4% for cases with small tumors. MIP3 plans demonstrated superior tumor coverage and were least sensitive to tumor size and tumor location.ConclusionMIP3 plans based on 4DCT scans were the best planning strategy for proton lung treatment planning.


Medical Physics | 2008

Spill-to-spill and daily proton energy consistency with a new accelerator control system.

Michael F. Moyers; A Ghebremedhin

The Loma Linda University proton accelerator has had several upgrades installed including synchrotron dipole power supplies and a system for monitoring the beam energy. The consistency of the energy from spill-to-spill has been tested by measuring the depth ionization at the distal edge as a function of time. These measurements were made with a minimally equipped beamline to reduce interference from confounding factors. The consistency of the energy over several months was measured in a treatment room beamline using an ionization chamber based daily quality assurance device. The results showed that the energy of protons delivered from the accelerator (in terms of water equivalent range) was consistent from spill-to-spill to better than +/-0.03 mm at 70, 155, and 250 MeV and that the energy check performed each day in the treatment room over a several month period was within +/-0.11 mm (+/-0.06 MeV) at 149 MeV. These results are within the tolerances required for the energy stacking technique.


The fourteenth international conference on the application of accelerators in research and industry | 1997

Spill uniformity measurements for a raster scanned proton beam

G. Coutrakon; A Ghebremedhin; J. Johanning; P. Koss

The method of scanning a proton beam across a target region for radiation therapy requires a uniform beam intensity throughout the beam spill time. Achieving uniform intensity using feedback to an air core quadrupole in the Loma Linda synchrotron accelerator is described in this paper. Frequency domain transfer functions and time domain intensity ripple measurements are presented followed with results and discussion of issues requiring additional work.

Collaboration


Dive into the A Ghebremedhin's collaboration.

Top Co-Authors

Avatar

B Patyal

Loma Linda University Medical Center

View shared research outputs
Top Co-Authors

Avatar

N Wang

Loma Linda University Medical Center

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

A Wroe

Loma Linda University Medical Center

View shared research outputs
Top Co-Authors

Avatar

G. Coutrakon

Northern Illinois University

View shared research outputs
Top Co-Authors

Avatar

Jerry D. Slater

Loma Linda University Medical Center

View shared research outputs
Top Co-Authors

Avatar

P. Koss

Loma Linda University Medical Center

View shared research outputs
Top Co-Authors

Avatar

A Mandapaka

Loma Linda University Medical Center

View shared research outputs
Top Co-Authors

Avatar

David A. Bush

Loma Linda University Medical Center

View shared research outputs
Top Co-Authors

Avatar

Michael F. Moyers

Loma Linda University Medical Center

View shared research outputs
Researchain Logo
Decentralizing Knowledge