Bryan R. Muir
National Research Council
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by Bryan R. Muir.
Medical Physics | 2015
Bryan R. Muir
PURPOSE To analyze absorbed dose calibration coefficients, ND,w, measured at accredited dosimetry calibration laboratories (ADCLs) for client ionization chambers to study (i) variability among ND,w coefficients for chambers of the same type calibrated at each ADCL to investigate ion chamber volume fluctuations and chamber manufacturing tolerances; (ii) equivalency of ion chamber calibration coefficients measured at different ADCLs by intercomparing ND,w coefficients for chambers of the same type; and (iii) the long-term stability of ND,w coefficients for different chamber types by investigating repeated chamber calibrations. METHODS Large samples of ND,w coefficients for several chamber types measured over the time period between 1998 and 2014 were obtained from the three ADCLs operating in the United States. These are analyzed using various graphical and numerical statistical tests for the four chamber types with the largest samples of calibration coefficients to investigate (i) and (ii) above. Ratios of calibration coefficients for the same chamber, typically obtained two years apart, are calculated to investigate (iii) above and chambers with standard deviations of old/new ratios less than 0.3% meet stability requirements for accurate reference dosimetry recommended in dosimetry protocols. RESULTS It is found that ND,w coefficients for a given chamber type compared among different ADCLs may arise from differing probability distributions potentially due to slight differences in calibration procedures and/or the transfer of the primary standard. However, average ND,w coefficients from different ADCLs for given chamber types are very close with percent differences generally less than 0.2% for Farmer-type chambers and are well within reported uncertainties. CONCLUSIONS The close agreement among calibrations performed at different ADCLs reaffirms the Calibration Laboratory Accreditation Subcommittee process of ensuring ADCL conformance with National Institute of Standards and Technology standards. This study shows that ND,w coefficients measured at different ADCLs are statistically equivalent, especially considering reasonable uncertainties. This analysis of ND,w coefficients also allows identification of chamber types that can be considered stable enough for accurate reference dosimetry.
Journal of Applied Clinical Medical Physics | 2017
Bryan R. Muir; Wesley S. Culberson; Stephen Davis; G Kim; Y Huang; S Lee; J Lowenstein; A Sarfehnia; J Siebers; Naresh Tolani
Purpose To present the results and discuss potential insights gained through surveys on reference dosimetry practices. Methods Two surveys were sent to medical physicists to learn about the current state of reference dosimetry practices at radiation oncology clinics worldwide. A short survey designed to maximize response rate was made publicly available and distributed via the AAPM website and a medical physics list server. Another, much more involved survey, was sent to a smaller group of physicists to gain insight on detailed dosimetry practices. The questions were diverse, covering reference dosimetry practices on topics like measurements required for beam quality specification, the actual measurement of absorbed dose and ancillary equipment required like electrometers and environment monitoring measurements. Results There were 190 respondents to the short survey and seven respondents to the detailed survey. The diversity of responses indicates nonuniformity in reference dosimetry practices and differences in interpretation of reference dosimetry protocols. Conclusions The results of these surveys offer insight on clinical reference dosimetry practices and will be useful in identifying current and future needs for reference dosimetry.
Medical Physics | 2017
Bryan R. Muir; C Cojocaru; M McEwen; C. K. Ross
Purpose: To provide results of water calorimetry and ion chamber measurements in high‐energy electron beams carried out at the National Research Council Canada (NRC). There are three main aspects to this work: (a) investigation of the behavior of ionization chambers in electron beams of different energies with focus on long‐term stability, (b) water calorimetry measurements to determine absorbed dose to water in high‐energy beams for direct calibration of ion chambers, and (c) using measurements of chamber response relative to reference ion chambers, determination of beam quality conversion factors, kQ, for several ion chamber types. Methods: Measurements are made in electron beams with energies between 8 MeV and 22 MeV from the NRC Elekta Precise clinical linear accelerator. Ion chamber measurements are made as a function of depth for cylindrical and plane‐parallel ion chambers over a period of five years to investigate the stability of ion chamber response and for indirect calibration. Water calorimetry measurements are made in 18 MeV and 22 MeV beams. An insulated enclosure with fine temperature control is used to maintain a constant temperature (drifts less than 0.1 mK/min) of the calorimeter phantom at 4°C to minimize effects from convection. Two vessels of different designs are used with calibrated thermistor probes to measure radiation induced temperature rise. The vessels are filled with high‐purity water and saturated with H2 or N2 gas to minimize the effect of radiochemical reactions on the measured temperature rise. A set of secondary standard ion chambers are calibrated directly against the calorimeter. Finally, several other ion chambers are calibrated in the NRC 60Co reference field and then cross‐calibrated against the secondary standard chambers in electron beams to realize kQ factors. Results: The long‐term stability of the cylindrical ion chambers in electron beams is better (always Symbol%) than plane‐parallel chambers (0.2% to 0.4%). Calorimetry measurements made at 22 MeV with two different vessel geometries are consistent within 0.2% after correction for the vessel perturbation. Measurements of absorbed dose calibration coefficients for the same secondary standard chamber separated in time by 10 yr are within 0.2%. Drifts in linac output that would affect the transfer of the standard are mitigated to the 0.1% level by performing daily ion chamber normalization measurements. Calibration coefficients for secondary standard ion chambers can be achieved with uncertainties less than 0.4% (k = 1) in high‐energy electron beams. The additional uncertainty in deriving calibration coefficients for well‐behaved chambers indirectly against the secondary standard reference chambers is negligible. The kQ factors measured here differ by up to 1.3% compared to those in TG‐51, an important change for reference dosimetry measurements. Symbol. No Caption available. Conclusions: The measurements made here of kQ factors for eight plane‐parallel and six cylindrical ion chambers will impact future updates of reference dosimetry protocols by providing some of the highest quality measurements of this crucial dosimetric parameter.
Medical Physics | 2014
Bryan R. Muir; M McEwen
PURPOSE To investigate the use of cylindrical Farmer-type ionization chambers to improve the accuracy of low-energy electron beam calibration. Historically, these chamber types have not been used in beams with incident energies less than 10 MeV (R5 0 < 4.3 cm) because early investigations suggested large (up to 5 %) fluence perturbation factors in these beams, implying that a significant component of uncertainty would be introduced if used for calibration. More recently, the assumptions used to determine perturbation corrections for cylindrical chambers have been questioned. METHODS Measurements are made with cylindrical chambers in Elekta Precise 4, 8 and 18 MeV electron beams. Several chamber types are investigated that employ graphite walls and aluminum electrodes with very similar specifications (NE2571, NE2505/3, FC65-G). Depth-ionization scans are measured in water in the 8 and 18 MeV beams. To reduce uncertainty from chamber positioning, measurements in the 4 MeV beam are made at the reference depth in Virtual Water™. The variability of perturbation factors is quantified by comparing normalized response of various chambers. RESULTS Normalized ion chamber response varies by less than 0.7 % for similar chambers at average electron energies corresponding to that at the reference depth from 4 or 6 MeV beams. Similarly, normalized measurements made with similar chambers at the reference depth in the 4 MeV beam vary by less than 0.4 %. Absorbed dose calibration coefficients derived from these results are stable within 0.1 % on average over a period of 6 years. CONCLUSION These results indicate that the uncertainty associated with differences in fluence perturbations for cylindrical chambers with similar specifications is only 0.2 %. The excellent long-term stability of these chambers in both photon and electron beams suggests that these chambers might offer the best performance for all reference dosimetry applications.
Physics in Medicine and Biology | 2017
E Christiansen; Bryan R. Muir; J Belec; E. Vandervoort
A formalism has been proposed for small and non-standard photon fields in which [Formula: see text] correction factors are used to correct dosimeter response in small fields (indiviual or composite) relative to that in a larger machine-specific reference (MSR) field. For clinical plans consisting of several fields, a plan-class specific reference (PCSR) plan can also be defined, serving as an intermediate calibration field between the MSR and clinical plans within a certain plan-class. In this work, the formalism was applied in the calculation of [Formula: see text] for 21 clinical plans delivered by the [Formula: see text] radiosurgery system, each plan employing one or two of the smallest diameter collimators: 5 mm, 7.5 mm, and 10 mm. Three detectors were considered: the Exradin A16 and A26 micro chambers, and the W1 plastic scintillator. The clinical plans were grouped into 7 plan-classes according to commonly shared characteristics. The suitability of using a PCSR plan to represent the detector response of each plan within the plan-class was investigated. Total and intermediate correction factors were calculated using the [Formula: see text] Monte Carlo user code. The corrections for the micro chambers were large, primarily due to the presence of the low-density air cavity and the volume averaging effect. The correction for the scintillator was found to be close to unity for most plans, indicating that this detector may be used to measure small clinical plan correction factors in any plan except for those using the 5 mm collimator. The PCSR plan was shown to be applicable to plan-classes comprising isocentric plans only, with plan-classes divided according to collimator size. For non-isocentric plans, the variation of [Formula: see text] as a function of the point of measurement within a single plan, as well as the high inter-plan-class variability of the correction factor, precludes the use of a PCSR plan.
Medical Physics | 2017
Bryan R. Muir; M McEwen
Purpose: To investigate the use of cylindrical chambers for electron beam dosimetry independent of energy by studying the variability of relative ion chamber perturbation corrections, one of the main concerns for electron beam dosimetry with cylindrical chambers. Methods: Measurements are made with sets of cylindrical and plane‐parallel reference‐class chambers as a function of depth in water in 8 MeV and 18 MeV electron beams. The ratio of chamber readings for similar chambers is normalized in a high‐energy electron beam and can be thought of as relative perturbation corrections. Data are plotted as a function of mean electron energy at depth for a range of depths close to the phantom surface to R80, the depth at which the ionization falls to 80% of its maximum value. Additional, similar measurements are made in a Virtual Water® phantom with cylindrical chambers at the reference depth in a 4 MeV electron beam. Results: The variability of relative ion chamber perturbation corrections for nominally identical cylindrical Farmer‐type chambers is found to be less than 0.4%, no worse than plane‐parallel chambers with similar specifications. Conclusions: This work discusses several issues related to the use of plane‐parallel ion chambers and suggests that reference‐class cylindrical chambers may be appropriate for reference dosimetry of all electron beams. This would simplify the reference dosimetry procedure and improve accuracy of beam calibration.
Medical Physics | 2016
Eric Vandervoort; Eric Christiansen; Jason Belec; Bryan R. Muir
Purpose: The purpose of this work is to investigate the utility of plan class specific reference (PCSR) fields for predicting dosimeter response within isocentric and non-isocentric composite clinical fields using the smallest fields employed by the CyberKnife radiosurgery system. Methods: Monte Carlo dosimeter response correction factors (CFs) were calculated for a plastic scintillator and microchamber dosimeter in 21 clinical fields and 9 candidate plan-class PCSR fields which employ the 5, 7.5 and 10 mm diameter collimators. Measurements were performed in 5 PCSR fields to confirm the predicted relative response of detectors in the same field. Results: Ratios of corrected measured dose in the PCSR fields agree to within 1% of unity. Calculated CFs for isocentric fields agree within 1.5% of those for PCSR fields. Large and variable microchamber CFs are required for non-isocentric fields, with differences as high as 5% between different clinical fields in the same plan class and 4% within the same field depending on the point of measurement. Non-isocentric PCSR fields constructed to have relatively homogenous dose over a region larger than the detector have very different ion chamber CFs from clinical fields. The plastic scintillator detector has much more consistent response within each plan class but still require 3–4% corrections in some fields. Conclusions: While the PCSR field concept is useful for small isocentric fields, this approach may not be appropriate for non-isocentric clinical fields which exhibit large and variable ion chamber CFs which differ significantly from CFs for homogenous field PCSRs.
Medical Physics | 2016
Bryan R. Muir; M McEwen; J Belec; E Christiansen; E. Vandervoort
PURPOSE To investigate small field dosimetry measurements and associated uncertainties when conical applicators are used to shape treatment fields from two different accelerating systems. METHODS Output factor measurements are made in water in beams from the CyberKnife radiosurgery system, which uses conical applicators to shape fields from a (flattening filter-free) 6 MV beam, and in a 6 MV beam from the Elekta Precise linear accelerator (with flattening filter) with BrainLab external conical applicators fitted to shape the field. The measurements use various detectors: (i) an Exradin A16 ion chamber, (ii) two Exradin W1 plastic scintillation detectors, (iii) a Sun Nuclear Edge diode, and (iv) two PTW microDiamond synthetic diamond detectors. Profiles are used for accurate detector positioning and to specify field size (FWHM). Output factor measurements are corrected with detector specific correction factors taken from the literature where available and/or from Monte Carlo simulations using the EGSnrc code system. RESULTS Differences in measurements of up to 1.7% are observed with a given detector type in the same beam (i.e., intra-detector variability). Corrected results from different detectors in the same beam (inter-detector differences) show deviations up to 3 %. Combining data for all detectors and comparing results from the two accelerators results in a 5.9% maximum difference for the smallest field sizes (FWHM=5.2-5.6 mm), well outside the combined uncertainties (∼1% for the smallest beams) and/or differences among detectors. This suggests that the FWHM of a measured profile is not a good specifier to compare results from different small fields with the same nominal energy. CONCLUSION Large differences in results for both intra-detector variability and inter-detector differences suggest potentially high uncertainties in detector-specific correction factors. Differences between the results measured in circular fields from different accelerating systems provide insight into sources of variability in small field dosimetric measurements reported in the literature.
Medical Physics | 2016
G Kim; Bryan R. Muir; W Culberson; Stephen Davis; Y Huang; S Lee; J Lowenstein; A Sarfehnia; N Tolani; J Siebers
PURPOSE The working group on the review and extension of the TG-51 protocol (WGTG51) collected data from American Association of Physicists in Medicine (AAPM) members with respect to their current TG-51 and associated addendum usage in the interest of considering future protocol addenda and guidance on reference dosimetry best practices. This study reports an overview of this survey on dosimetry of external beams. METHODS Fourteen survey questions were developed by WGTG51 and released in November 2015. The questions collected information on reference dosimetry, beam quality specification, and ancillary calibration equipment. RESULTS Of the 190 submissions completed worldwide (U.S. 70%), 83% were AAPM members. Of the respondents, 33.5% implemented the TG-51 addendum, with the maximum calibration difference for any photon beam, with respect to the original TG-51 protocol, being <1% for 97.4% of responses. One major finding is that 81.8% of respondents used the same cylindrical ionization chamber for photon and electron dosimetry, implying that many clinics are foregoing the use of parallel-plate chambers. Other evidence suggests equivalent dosimetric results can be obtained with both cylindrical and parallel-plate chambers in electron beams. This, combined with users comfort with cylindrical chambers for electrons will likely impact recommendations put forward in an upcoming electron beam addendum to the TG-51 protocol. Data collected on ancillary equipment showed 58.2% (45.0%) of the thermometers (barometers) in use for beam calibration had NIST traceable calibration certificates, but 48.4% (42.7%) were never recalibrated. CONCLUSION This survey provides a snapshot of TG-51 external beam reference dosimetry practice in radiotherapy centers. Findings demonstrate the rapid take-up of the TG-51 photon beam addendum and raise issues for the WGTG51 to focus on going forward, including guidelines on ancillary equipment and the choice of chamber for electron beam dosimetry.
Medical Physics | 2015
Bryan R. Muir; M McEwen
Purpose: To investigate uncertainties in small field output factors and detector specific correction factors from variations in field size for nominally identical fields using measurements and Monte Carlo simulations. Methods: Repeated measurements of small field output factors are made with the Exradin W1 (plastic scintillation detector) and the PTW microDiamond (synthetic diamond detector) in beams from the Elekta Precise linear accelerator. We investigate corrections for a 0.6x0.6 cm2 nominal field size shaped with secondary photon jaws at 100 cm source to surface distance (SSD). Measurements of small field profiles are made in a water phantom at 10 cm depth using both detectors and are subsequently used for accurate detector positioning. Supplementary Monte Carlo simulations with EGSnrc are used to calculate the absorbed dose to the detector and absorbed dose to water under the same conditions when varying field size. The jaws in the BEAMnrc model of the accelerator are varied by a reasonable amount to investigate the same situation without the influence of measurements uncertainties (such as detector positioning or variation in beam output). Results: For both detectors, small field output factor measurements differ by up to 11 % when repeated measurements are made in nominally identical 0.6x0.6 cm2 fields. Variations in the FWHM of measured profiles are consistent with field size variations reported by the accelerator. Monte Carlo simulations of the dose to detector vary by up to 16 % under worst case variations in field size. These variations are also present in calculations of absorbed dose to water. However, calculated detector specific correction factors are within 1 % when varying field size because of cancellation of effects. Conclusion: Clinical physicists should be aware of potentially significant uncertainties in measured output factors required for dosimetry of small fields due to field size variations for nominally identical fields.