Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where E. Vandervoort is active.

Publication


Featured researches published by E. Vandervoort.


Medical Physics | 2015

WE-AB-303-05: Breathing Motion of Liver Segments From Fiducial Tracking During Robotic Radiosurgery and Comparison with 4D-CT-Derived Fiducial Motion

J Sutherland; J Pantarotto; V Nair; G Cook; M Plourde; E. Vandervoort

Purpose: To quantify respiratory-induced motion of liver segments using the positions of implanted fiducials during robotic radiosurgery. This study also compared fiducial motion derived from four-dimensional computed tomography (4D-CT) maximum intensity projections (MIP) with motion derived from imaging during treatment. Methods: Forty-two consecutive liver patients treated with liver ablative radiotherapy were accrued to an ethics approved retrospective study. The liver segment in which each fiducial resided was identified. Fiducial positions throughout each treatment fraction were determined using orthogonal kilovoltage images. Any data due to patient repositioning or motion was removed. Mean fiducial positions were calculated. Fiducial positions beyond two standard deviations of the mean were discarded and remaining positions were fit to a line segment using least squares minimization (LSM). For eight patients, fiducial motion was derived from 4D-CT MIPs by calculating the CT number weighted mean position of the fiducial on each slice and fitting a line segment to these points using LSM. Treatment derived fiducial trajectories were corrected for patient rotation and compared to MIP derived trajectories. Results: The mean total magnitude of fiducial motion across all liver segments in left-right, anteroposterior, and superoinferior (SI) directions were 3.0 ± 0.2 mm, 9.3 ± 0.4 mm, and 20.5 ± 0.5 mm, respectively. Differences in per-segment mean fiducial motion were found with SI motion ranging from 12.6 ± 0.8 mm to 22.6 ± 0.9 mm for segments 3 and 8, respectively. Large, varied differences between treatment and MIP derived motion at simulation were found with the mean difference for SI motion being 2.6 mm (10.8 mm standard deviation). Conclusion: The magnitude of liver fiducial motion was found to differ by liver segment. MIP derived liver fiducial motion differed from motion observed during treatment, implying that 4D-CTs may not accurately capture the range of liver motion across fractions and during treatment. Author V. Nair was funded by the Cushing estate for a SABR clinical research fellowship


Journal of Applied Clinical Medical Physics | 2018

COMP Report: CPQR technical quality control guidelines for CyberKnife® Technology

E. Vandervoort; Horacio Patrocinio; Tom Chow; E Soisson; Dominic Béliveau Nadeau

Abstract The Canadian Organization of Medical Physicists (COMP), in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR) has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. This particular TQC contains detailed performance objectives and safety criteria for CyberKnife® Technology. The quality control recommendations in this document are based upon previously published guidelines and the collective experience of all Canadian sites using this technology. This TQC guideline has been field tested at the newest Canadian CyberKnife installation site and includes recommendations for quality control of the Iris™ and InCise™ MLC collimation systems.


Journal of Applied Clinical Medical Physics | 2018

Comparison of four techniques for spine stereotactic body radiotherapy: Dosimetric and efficiency analysis

Saif Aljabab; Balamurugan Vellayappan; E. Vandervoort; Jamie Bahm; Robert Zohr; John Sinclair; Jean-Michel Caudrelier; J. Szanto; Shawn Malone

Abstract Purpose The aim of this study is to compare the dosimetric differences between four techniques for spine stereotactic body radiotherapy (SBRT): CyberKnife (CK), volumetric modulated arc therapy (VMAT), and helical tomotherapy (HT) with dynamic jaws (HT‐D) and fixed jaws (HT‐F). Materials/methods Data from 10 patients were utilized. All patients were planned for 24 Gy in two fractions, with the primary objectives being: (a) restricting the maximum dose to the cord to ≤ 17 Gy and/or cauda equina to ≤ 20 Gy, and (b) to maximize the clinical target volume (CTV) to receive the prescribed dose. Treatment plans were generated by separate dosimetrists and then compared using velocity AI. Parameters of comparison include target volume coverage, conformity index (CI), gradient index (GI), homogeneity index (HI), treatment time (TT) per fraction, and monitor units (MU) per fraction. Results PTV D2 and D5 were significantly higher for CK compared to VMAT, HT‐F, and HT‐D (P < 0.001). The average volume of CTV receiving the prescription dose (CTV D95) was significantly less for VMAT compared to CK, HT‐F and HT‐D (P = 0.036). CI improved for CK (0.69), HT‐F (0.66), and HT‐D (0.67) compared to VMAT (0.52) (P = 0.013). CK (41.86) had the largest HI compared to VMAT (26.99), HT‐F (20.69), and HT‐D (21.17) (P < 0.001). GI was significantly less for CK (3.96) compared to VMAT (6.76) (P = 0.001). Likewise, CK (62.4 min, 14059 MU) had the longest treatment time and MU per fraction compared to VMAT (8.5 min, 9764 MU), HT‐F (13 min, 10822 MU), and HT‐D (13.5 min, 11418 MU) (P < 0.001). Conclusion Both CK and HT plans achieved conformal target coverage while respecting cord tolerance. Dose heterogeneity was significantly larger in CK. VMAT required the least treatment time and MU output, but had the least steep GI, CI, and target coverage.


Physics in Medicine and Biology | 2017

Small composite field correction factors for the CyberKnife radiosurgery system: clinical and PCSR plans

E Christiansen; Bryan R. Muir; J Belec; E. Vandervoort

A formalism has been proposed for small and non-standard photon fields in which [Formula: see text] correction factors are used to correct dosimeter response in small fields (indiviual or composite) relative to that in a larger machine-specific reference (MSR) field. For clinical plans consisting of several fields, a plan-class specific reference (PCSR) plan can also be defined, serving as an intermediate calibration field between the MSR and clinical plans within a certain plan-class. In this work, the formalism was applied in the calculation of [Formula: see text] for 21 clinical plans delivered by the [Formula: see text] radiosurgery system, each plan employing one or two of the smallest diameter collimators: 5 mm, 7.5 mm, and 10 mm. Three detectors were considered: the Exradin A16 and A26 micro chambers, and the W1 plastic scintillator. The clinical plans were grouped into 7 plan-classes according to commonly shared characteristics. The suitability of using a PCSR plan to represent the detector response of each plan within the plan-class was investigated. Total and intermediate correction factors were calculated using the [Formula: see text] Monte Carlo user code. The corrections for the micro chambers were large, primarily due to the presence of the low-density air cavity and the volume averaging effect. The correction for the scintillator was found to be close to unity for most plans, indicating that this detector may be used to measure small clinical plan correction factors in any plan except for those using the 5 mm collimator. The PCSR plan was shown to be applicable to plan-classes comprising isocentric plans only, with plan-classes divided according to collimator size. For non-isocentric plans, the variation of [Formula: see text] as a function of the point of measurement within a single plan, as well as the high inter-plan-class variability of the correction factor, precludes the use of a PCSR plan.


Medical Physics | 2016

SU-F-T-577: Comparison of Small Field Dosimetry Measurements in Fields Shaped with Conical Applicators On Two Different Accelerating Systems

Bryan R. Muir; M McEwen; J Belec; E Christiansen; E. Vandervoort

PURPOSE To investigate small field dosimetry measurements and associated uncertainties when conical applicators are used to shape treatment fields from two different accelerating systems. METHODS Output factor measurements are made in water in beams from the CyberKnife radiosurgery system, which uses conical applicators to shape fields from a (flattening filter-free) 6 MV beam, and in a 6 MV beam from the Elekta Precise linear accelerator (with flattening filter) with BrainLab external conical applicators fitted to shape the field. The measurements use various detectors: (i) an Exradin A16 ion chamber, (ii) two Exradin W1 plastic scintillation detectors, (iii) a Sun Nuclear Edge diode, and (iv) two PTW microDiamond synthetic diamond detectors. Profiles are used for accurate detector positioning and to specify field size (FWHM). Output factor measurements are corrected with detector specific correction factors taken from the literature where available and/or from Monte Carlo simulations using the EGSnrc code system. RESULTS Differences in measurements of up to 1.7% are observed with a given detector type in the same beam (i.e., intra-detector variability). Corrected results from different detectors in the same beam (inter-detector differences) show deviations up to 3 %. Combining data for all detectors and comparing results from the two accelerators results in a 5.9% maximum difference for the smallest field sizes (FWHM=5.2-5.6 mm), well outside the combined uncertainties (∼1% for the smallest beams) and/or differences among detectors. This suggests that the FWHM of a measured profile is not a good specifier to compare results from different small fields with the same nominal energy. CONCLUSION Large differences in results for both intra-detector variability and inter-detector differences suggest potentially high uncertainties in detector-specific correction factors. Differences between the results measured in circular fields from different accelerating systems provide insight into sources of variability in small field dosimetric measurements reported in the literature.


Medical Physics | 2016

Poster - 11: Radiation barrier thickness calculations for the GammaPod

Daniel La Russa; E. Vandervoort; David Wilkins

A consortium of radiotherapy centers in North America is in the process of evaluating a novel new 60Co teletherapy device, called the GammaPod™ (Xcision Medical Systems, Columbia Maryland), designed specifically for breast SBRT. The GammaPod consists of 36 collimated 60Co sources with a total activity of 4320 Ci. The sources are housed in a hemispherical source carrier that rotates during treatment to produce a cylindrically symmetric cone of primary beam spanning 16° – 54° degrees from the horizontal. This unique beam geometry presents challenges when designing or evaluating room shielding for the purposes of meeting regulatory requirements, and for ensuring the safety of staff and the public in surrounding areas. Conventional methods for calculating radiation barrier thicknesses have been adapted so that barrier transmission factors for the GammaPod can be determined from a few relevant distances and characteristics of the primary beam. Simple formalisms have been determined for estimating shielding requirements for primary radiation (with a rotating and non-rotating source carrier), patient-scattered radiation, and leakage radiation. When making worst case assumptions, it was found that conventional barrier thicknesses associated with linac treatment suites are sufficient for shielding all sources of radiation from the GammaPod.


Medical Physics | 2014

Poster — Thur Eve — 19: Performance assessment of a 160‐leaf beam collimation system

Elsayed Ali; D La Russa; E. Vandervoort

In this study, the performance of the new beam collimation system with 160 leaves, each with a 5 mm leaf width projected at isocenter, is evaluated in terms of positional accuracy and plan/delivery quality. Positional accuracy was evaluated using a set of static and dynamic MLC/jaw delivery patterns at different gantry angles, dose rates, and MLC/jaw speeds. The impact on IMRT plan quality was assessed by comparing against a previous generation collimation system using the same optimization parameters, while delivery quality was quantified using a combination of patient-specific QA measurements with ion chambers, film, and a bi-planar diode array. Positional accuracy for four separate units was comparable. The field size accuracy, junction width, and total displacement over 16 cm leaf travel are 0.3 ± 0.2 mm, 0.4 ± 0.3 mm, and 0.5 ± 0.2 mm, respectively. The typical leaf minor offset is 0.05 ± 0.04 mm, and MLC hysteresis effects are 0.2 ± 0.1 mm over 16 cm travel. The dynamic output is linear with MU and MLC/jaw speed, and is within 0.7 ± 0.3 % of the planning system value. Plan quality is significantly improved both in terms of target coverage and OAR sparing due, in part, to the larger allowable MLC and jaw speeds. γ-index pass rates for the patient-specific QA measurements exceeded 97% using criteria of 2%/2 mm. In conclusion, the performance of the Agility system is consistent among four separate installations, and is superior to its previous generations of collimation systems.


Medical Physics | 2013

SU‐E‐T‐424: Improved Dosimetric Accuracy for Cyberknife Patient Plans Using a Dual‐Detector Measurement Method for Relative Output Factors

E. Vandervoort; D La Russa; N Ploquin; J Szanto; E Henderson; P Francescon

PURPOSE The measurement of output factors (OFs) for small fields can lead to large dosimetric errors if detector effects are not accounted for. With its high spatial resolution and tissue equivalence, GAFCHROMIC film provides a correction free measure of OFs. We recently changed the OFs used in our Cyberknife planning system from uncorrected diode values to a dual detector method employing a diode with Monte-Carlo corrections for the smallest collimators and a micro ion chamber for collimators >10 mm in diameter. METHODS We measured OFs for the CyberKnife G4 fixed collimators (5 to 60 mm) using an A16 microchamber and an Edge diode detector. The diode measured OFs for collimator sizes <10 mm were corrected using Monte-Carlo correction factors. OFs were also measured using GAFCHROMIC film. We evaluated how this change in OFs influenced the dosimetric accuracy of patient specific QA measurements for 13 patient plans (9 before and 4 after the OF change) using film and the A16 chamber. RESULTS The OFs measured using the dual-detector method agree with film to within two standard deviations for the full range of collimator sizes. When the dual detector method OFs are used, we achieve better dosimetric agreement (<2 sigma for pixels within the 80% isodose) than with uncorrected diode OFs for all patient specific QA plans measured using film. For patient specific QA using the microchamber, we get good agreement (<3%) for collimator sizes >5 mm, with differences observed for the 5 mm collimator consistent with volume averaging and a 1 mm setup uncertainty. CONCLUSIONS OFs can be determined consistently using the dual-detector method and verified using film. For patient specific QA, we achieve good agreement with microchambers for collimators >5 mm in diameter but film is the most appropriate detector for patient specific QA using the 5 mm collimator.


Medical Physics | 2012

Sci—Thur AM: Planning ‐ 10: Improved dosimetric accuracy for patient specific quality assurance using a dual‐detector measurement method for cyberknife output factors

E. Vandervoort; D La Russa; N Ploquin; I Ait El Kadir; J Szanto

The measurement of output factors for small fields is challenging and can lead to large dose errors in patient treatments if corrections for detector size and scatter from high-Z material are not applied. Due to its high spatial resolution and near tissue equivalence, GAFCHROMIC® film potentially provides a correction free measure of output factors but it can be challenging to obtain high quality dosimetric results using this film. We propose minimizing errors in the clinical determination of small field output factors by employing diode measurements with Monte-Carlo generated corrections for small fields ≤10 mm diameter and using small volume ion chambers for apertures >10 mm diameter with independent validation using radiochromic film. We performed patient specific quality assurance (QA) measurements for 9 patients using GAFCHROMIC® film and an A16 small volume ion chamber in a head-shaped phantom, employing this hybrid dual detector method for relative output factor measurements within the Multiplan treatment planning system. Our results suggest that consistent output factors can be determined using this method with experimental verification using GAFCHROMIC® film dosimetry. For the patient specific QA using film, we achieve good dosimetric agreement (<2σ) of the measured and calculated average dose for pixels within the 80% isodose line. For patient specific QA using the micro-ion chamber, we get good agreement (<3%) for cone sizes greater than 5 mm. The differences observed for the 5 mm cone plans are consistent with a 1 mm radial setup uncertainty for patient positioning using the Cyberknife system.


Medical Physics | 2010

Sci—Sat AM(1): Planning — 10: Evaluation of a New Commercial Monte-Carlo Treatment Planning System for Electrons

E. Vandervoort; Joanna E. Cygler

It has long been understood that the Monte Carlo (MC) method is the most effective means for accurately computing the dose delivered by clinical electron beams. Every commercial implementation of the MC method involves design compromises and the possibility of error. It is important, therefore, that each implementation is independently validated under conditions similar to those found in the clinic. In this abstract, we present the initial stages of validation for the XiO electron Monte Carlo (XiO eMC) software, a new treatment planning system for electron beams developed and commercialized by CMS incorporated. In this abstract we present a limited set of comparisons of calculated and experimental data for homogeneous water phantoms and for a 3D heterogeneous phantom meant to approximate the geometry of a trachea and spine. All Monte Carlo calculated and measured output factors agree within the estimated standard error for standard and extended SSD for open applicators and cerrobend cutouts with the exception of the smallest cutout size (2×2cm2) for 17 MeV at extended SSD. We also found good agreement between calculated and experimental depth dose curves and dose profiles. Dose calculations in heterogeneous phantoms are also in a very good agreement with measurements, given an estimated positional uncertainty of ±0.1cm in the depth direction, and provided that appropriate calculation voxel sizes are used for a given geometry. Acknowledgment: The authors would like to acknowledge the excellent technical support provided by Dr. J C Satterthwaite of Elekta CMS Software.

Collaboration


Dive into the E. Vandervoort's collaboration.

Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Bryan R. Muir

National Research Council

View shared research outputs
Researchain Logo
Decentralizing Knowledge