Edward G. Solomon
Stanford University
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by Edward G. Solomon.
Medical Physics | 2004
Taly Gilat Schmidt; Rebecca Fahrig; Norbert J. Pelc; Edward G. Solomon
We propose an inverse-geometry volumetric CT system for acquiring a 15-cm volume in one rotation with negligible cone-beam artifacts. The system uses a large-area scanned source and a smaller detector array. This note describes two feasibility investigations. The first examines data sufficiency in the transverse planes. The second predicts the signal-to-noise ratio (SNR) compared to a conventional scanner. Results showed sufficient sampling of the full volume in less than 0.5 s and, when compared to a conventional scanner operating at 24 kW with a 0.5-s voxel illumination time (e.g., 0.5-s gantry rotation and pitch of one), predicted a relative SNR of 76%.
Medical Imaging 1999: Physics of Medical Imaging | 1999
Edward G. Solomon; Brian P. Wilfley; Michael S. Van Lysel; Aaron W. Joseph; Joseph A. Heanue
An advanced Scanning-Beam Digital X-ray (SBDX) system for cardiac angiography has been constructed. The 15-kW source operates at 70 - 120 kVp and has an electron beam that is electromagnetically scanned across a 23-cm X 23-cm transmission target. The target is directly liquid cooled for continuous full-power operation and is located behind a focused source collimator. The collimator is a rectangular grid of 100 X 100 apertures whose axes are aligned with the center of the detector array. X-ray beam divergence through the collimator apertures is matched to the 5.4-cm X 5.4 cm detector, which is 150 cm from the source. The detector is a 48 X 48 element CdZnTe direct-conversion photon-counting detector. A narrow x-ray beam scans the full field of view at up to 30 frames per second. A custom digital processor simultaneously reconstructs sixteen 1,0002 pixel tomographic images in real time. The slices are spaced 1.2 cm apart and cover the entire cardiac anatomy. The small detector area and large patient-detector distance result in negligible detected x-ray scatter. Image signal-to-noise ratio is calculated to be equal to conventional fluoroscopic systems at only 12% of the patient exposure and 25% of the staff exposure. Exposure reduction is achieved by elimination of detected scatter, elimination of the anti-scatter grid, increased detector DQE, and increased patient entrance area.
Medical Physics | 2013
Scott S. Hsieh; Joseph A. Heanue; Tobias Funk; Waldo Hinshaw; Brian P. Wilfley; Edward G. Solomon; Norbert J. Pelc
PURPOSE Inverse geometry computed tomography (IGCT) has been proposed as a new system architecture that combines a small detector with a large, distributed source. This geometry can suppress cone-beam artifacts, reduce scatter, and increase dose efficiency. However, the temporal resolution of IGCT is still limited by the gantry rotation time. Large reductions in rotation time are in turn difficult due to the large source array and associated power electronics. We examine the feasibility of using stationary source arrays for IGCT in order to achieve better temporal resolution. We anticipate that multiple source arrays are necessary, with each source array physically separated from adjacent ones. METHODS Key feasibility issues include spatial resolution, artifacts, flux, noise, collimation, and system timing clashes. The separation between the different source arrays leads to missing views, complicating reconstruction. For the special case of three source arrays, a two-stage reconstruction algorithm is used to estimate the missing views. Collimation is achieved using a rotating collimator with a small number of holes. A set of equally spaced source spots are designated on the source arrays, and a source spot is energized when a collimator hole is aligned with it. System timing clashes occur when multiple source spots are scheduled to be energized simultaneously. We examine flux considerations to evaluate whether sufficient flux is available for clinical applications. RESULTS The two-stage reconstruction algorithm suppresses cone-beam artifacts while maintaining resolution and noise characteristics comparable to standard third generation systems. The residual artifacts are much smaller in magnitude than the cone-beam artifacts eliminated. A mathematical condition is given relating collimator hole locations and the number of virtual source spots for which system timing clashes are avoided. With optimization, sufficient flux may be achieved for many clinical applications. CONCLUSIONS IGCT with stationary source arrays could be an imaging platform potentially capable of imaging a complete 16-cm thick volume within a tenth of a second.
Medical Imaging 1996: Physics of Medical Imaging | 1996
Edward G. Solomon; Michael S. Van Lysel; Robert E. Melen; Jack W. Moorman; Brian Skillicorn
A prototype scanning-beam digital x-ray system for cardiac fluoroscopy has been constructed. Source-to-detector distance is 94 cm with the subject positioned near the source. The 4-kW source operates at 70-110 kVp and has an electromagnetically-scanned 25-cm-diameter transmission target. The target is at ground potential and is directly liquid cooled for continuous full-power operation. The source collimator has 22,000 holes whose axes are aligned with the center of the detector array. Beam divergence through the 0.38-mm-diameter collimator holes is matched to the 1.8-cm diameter of the detector array. The detector is a 96- element scintillator array optically coupled to a 96-channel photomultiplier tube. A narrow (0.6 degree half-angle) x-ray beam scans the 19-cm-diameter field of view at 30 frames/sec. A two-dimensional shift-and-add reconstruction algorithm produces a narrow-angle classical tomographic view of the subject in real time. The small detector area and large patient- detector distance result in negligible detected x-ray scatter. Signal-to-noise ratio is calculated to be equal to conventional fluoroscopic systems with ten times less patient skin exposure and better than four times less patient integral dose. Exposure reduction is due to the elimination of x-ray scatter and the anti-scatter grid, increased detector DQE, and geometric considerations.
Medical Imaging 2003: Physics of Medical Imaging | 2003
Rebecca Fahrig; Angel R. Pineda; Edward G. Solomon; Ann N. Leung; Norbert J. Pelc
Radiology-based lung-cancer detection is a high-contrast imaging task, consisting of the detection of a small mass of tissue within much lower density lung parenchyma. This imaging task requires removal of confounding image details, fast image acquisition (< 0.1 s for pericardial region), low dose (comparable to a chest x-ray), high resolution (< 0.25 mm in-plane) and patient positioning flexibility. We present an investigation of tomosynthesis, implemented using the Scanning-Beam Digital X-ray System (SBDX), to achieve these goals. We designed an image-based computer model of tomosynthesis using a high-resolution (0.15-mm isotropic voxels), low-noise CT volume image of a lung phantom, numerically added spherical lesions and convolution-based tomographic blurring. Lesion visibility was examined as a function of half-tomographic angle for 2.5 and 4.0 mm diameter lesions. Gaussian distributed noise was added to the projected images. For lesions 2.5 mm and 4.0 mm in diameter, half-tomographic angles of at least 6° and 9° respectively were necessary before visualization of the lesions improved. The addition of noise for a dose equivalent to 1/10 that used for a standard chest radiograph did not significantly impair lesion detection. The results are promising, indicating that lung-cancer detection using a modified SBDX system is possible.
Medical Imaging 1997: Physics of Medical Imaging | 1997
Michael S. Van Lysel; Edward G. Solomon; Brian P. Wilfley; April Dutta; Michael A. Speidel
A prototype scanning-beam digital x-ray (SBDX) system for cardiac fluoroscopy has been constructed. The unique geometry and absence of detected x-ray scatter in the SBDX image promises to provide image quality equivalent to a conventional image-intensifier-based fluoroscopic system at substantially reduced x-ray exposure to patient and staff. In order to measure the SBDX exposure advantage, a contrast- detail study was performed comparing SBDX and a conventional cardiac fluoroscopic system. Low-contrast deductibility as a function of the phantom entrance exposure was determined. The expected SBDX exposure advantage was 3.0 to 3.4, for low-contrast objects ranging in diameter from 2 to 10 mm. This exposure advantage is applicable to the AP projection through an average-size cardiac patient. Based on these results, calculations show that angulated views and larger patients will experience significantly greater exposure reductions. In addition, the results also indicate that SBDX system design modifications can provide a greater exposure reduction from that measured with this prototype.
Proceedings of SPIE | 2011
Steve Burion; Anne Sandman; Kate Leeann Bechtel; Edward G. Solomon; Tobias Funk
Radiation dose is particularly a concern in pediatric cardiac fluoroscopy procedures, which account for 7% of all cardiac procedures performed. The Scanning-Beam Digital X-ray (SBDX) fluoroscopy system has already demonstrated reduced dose in adult patients owing to its high-DQE photon-counting detector, reduced detected scatter, and the elimination of the anti-scatter grid. Here we show that the unique flexible illumination platform of the SBDX system will enable further dose area product reduction, which we are currently developing for pediatric patients, but which will ultimately benefit all patients. The SBDX system has a small-area detector array and a large-area X-ray source with up to 9,000 individually-controlled X-ray focal spots. Each focal spot illuminates a small fraction of the full field of view. To acquire a frame, each focal spot is activated for a fixed number of 1-microsecond periods. Dose reduction is made possible by reducing the number of activations of some of the X-ray focal spots during each frame time. This can be done dynamically to reduce the exposure in areas of low patient attenuation, such as the lung field. This spatially-adaptive illumination also reduces the dynamic range in the full image, which is visually pleasing. Dose can also be reduced by the user selecting a region of interest (ROI) where full image quality is to be maintained. Outside the ROI, the number of activations of each X-ray focal spot is reduced and the image gain is correspondingly increased to maintain consistent image brightness. Dose reduction is dependent on the size of the ROI and the desired image quality outside the ROI. We have developed simulation software that is based on real data and can simulate the performance of the equalization and ROI filtration. This software represents a first step toward real-time implementation of these dose-reduction methods. Our simulations have shown that dose area product reductions of 40% are possible using equalization, and dose savings as high as 74% are possible with the ROI approach. The dose reduction achieved in clinical use will depend on patient anatomy.
Medical Imaging 2004: Physics of Medical Imaging | 2004
Samuel R. Mazin; Taly Gilat Schmidt; Edward G. Solomon; Rebecca Fahrig; Norbert J. Pelc
An inverse-geometry volumetric CT (IGCT) system for imaging in a single fast rotation without cone-beam artifacts is being developed. It employs a large scanned source array and a smaller detector array. For a single-source/single-detector implementation, the FOV is limited to a fraction of the source size. Here we explore options to increase the FOV without increasing the source size by using multiple detectors spaced apart laterally to increase the range of radial distances sampled. We also look at multiple source array systems for faster scans. To properly reconstruct the FOV, Radon space must be sufficiently covered and sampled in a uniform manner. Optimal placement of the detectors relative to the source was determined analytically given system constraints (5cm detector width, 25cm source width, 45cm source-to-isocenter distance). For a 1x3 system (three detectors per source) detector spacing (DS) was 18deg and source-to-detector distances (SDD) were 113, 100 and 113cm to provide optimum Radon sampling and a FOV of 44cm. For multiple-source systems, maximum angular spacing between sources cannot exceed 125deg since detectors corresponding to one source cannot be occluded by a second source. Therefore, for 2x3 and 3x3 systems using the above DS and SDD, optimum spacing between sources is 115deg and 61deg respectively, requiring minimum scan rotations of 115deg and 107deg. Also, a 3x3 system can be much faster for full 360deg dataset scans than a 2x3 system (120deg vs. 245deg). We found that a significantly increased FOV can be achieved while maintaining uniform radial sampling as well as a substantial reduction in scan time using several different geometries. Further multi-parameter optimization is underway.
Filtration & Separation | 2004
Sungwon Yoon; Angel R. Pineda; Edward G. Solomon; Josh Star-Lack; Rebecca Fahrig
We are currently investigating the application of tomosynthesis to lung nodule detection using technology developed for the Scanning-Beam Digital X-ray (SBDX) system[1]. For system understanding and optimization, the interplay of various parameters must be investigated via simulations. We present a fast image-based SBDX system simulation model that produces equivalent tomosynthesis reconstructions to those from a physics-based model. Comparison between the two models were made using the central 75% of the reconstructed images. After applying geometric corrections arising from the SBDX system geometry, image-based model results were different by less than 3% and computed more than 10 times faster than physics-based model with comparable quality results. This work provides groundwork for SBDX system optimization for lung nodule detection. Furthermore, such analysis can be generalized to any tomosynthesis system for which the acquisition geometry is well known.
international conference of the ieee engineering in medicine and biology society | 2000
Rebecca Fahrig; Grant M. Stevens; Edward G. Solomon; Norbert J. Pelc
An ideal imaging technique for lung nodule screening would allow the visualization of small nodules within a complex anatomical background, use a low radiation dose technique, acquire images in <0.25 s, and retain patient positioning flexibility. A novel C-arm mounted scanning beam X-ray source and digital detector system (SBDX) can acquire tomosynthesis images in real-time. We investigate, using numerical simulation, this approach for lung nodule detection. A high-resolution CT volume (0.5 mm isotropic voxels) of a plastinated dog lung was acquired. Spherical nodules (40 HU) and overlying ribs (cortical bone1000 HU) were added numerically, providing a detection task with typical anatomic complexity. Tomographic blurring was modeled by convolving each slice with a normalized cylindrical blur function (edges rolled off using cosines). Lesion visibility was examined as a function of tome-angle and lesion size. For lesions 4.5 mm and 2.5 mm in diameter, half-tomo-angles of at least 3/spl deg/ and 4.5/spl deg/ respectively are necessary before visualization of the lesions improves. Modification of the SBDX system (current half-tome-angle=1.5/spl deg/) is therefore desired before optimal lung nodule detection is feasible. Possible approaches include increasing the size of the digital detector, and decreasing the object-to-detector distance.