Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Brian P. Wilfley is active.

Publication


Featured researches published by Brian P. Wilfley.


Medical Physics | 2006

Scanning-beam digital x-ray (SBDX) technology for interventional and diagnostic cardiac angiography

Michael A. Speidel; Brian P. Wilfley; Josh Star-Lack; Joseph A. Heanue; Michael S. Van Lysel

The scanning-beam digital x-ray (SBDX) system is designed for x-ray dose reduction in cardiac angiographic applications. Scatter reduction, efficient detection of primary x-rays, and an inverse beam geometry are the main components of the entrance dose reduction strategy. This paper reports the construction of an SBDX prototype, image reconstruction techniques, and measurements of spatial resolution and x-ray output. The x-ray source has a focal spot that is electronically scanned across a large-area transmission target. A multihole collimator beyond the target defines a series of x-ray beams directed at a distant small-area detector array. The prototype has a 23 cm X 23 cm target, 100 X 100 focal spot positions, and a 5 cm X 5 cm CdTe detector positioned 150 cm from the target. With this nonmechanical method of beam scanning, patient images with low detected scatter are generated at up to 30 frame/s. SBDX data acquisition is tomosynthetic. The prototype simultaneously reconstructs 16 planes spaced throughout the cardiac volume using shift-and-add backprojection. Image frames analogous to conventional projection images are generated with a multiplane compositing algorithm. Single-plane versus multiplane reconstruction of contrast-filled coronary arteries is demonstrated with images of the porcine heart. Phantom and porcine imaging studies show multiplane reconstruction is practicable under clinically realistic levels of patient attenuation and cardiac motion. The modulation transfer function for an in-plane slit at mechanical isocenter measured 0.41-0.56 at 1 cycle/mm, depending on the detector element to image pixel interpolation technique. Modeling indicates that desired gains in spatial resolution are achievable by halving the detector element width. The x-ray exposure rate 15 cm below isocenter, without table or patient in the beam, measured 11.5 R/min at 120 kVp, 24.3 kWp and 3.42 R/min at 70 kVp, 14.2 kWp.


Medical Imaging 1999: Physics of Medical Imaging | 1999

Scanning-beam digital x-ray (SBDX) system for cardiac angiography

Edward G. Solomon; Brian P. Wilfley; Michael S. Van Lysel; Aaron W. Joseph; Joseph A. Heanue

An advanced Scanning-Beam Digital X-ray (SBDX) system for cardiac angiography has been constructed. The 15-kW source operates at 70 - 120 kVp and has an electron beam that is electromagnetically scanned across a 23-cm X 23-cm transmission target. The target is directly liquid cooled for continuous full-power operation and is located behind a focused source collimator. The collimator is a rectangular grid of 100 X 100 apertures whose axes are aligned with the center of the detector array. X-ray beam divergence through the collimator apertures is matched to the 5.4-cm X 5.4 cm detector, which is 150 cm from the source. The detector is a 48 X 48 element CdZnTe direct-conversion photon-counting detector. A narrow x-ray beam scans the full field of view at up to 30 frames per second. A custom digital processor simultaneously reconstructs sixteen 1,0002 pixel tomographic images in real time. The slices are spaced 1.2 cm apart and cover the entire cardiac anatomy. The small detector area and large patient-detector distance result in negligible detected x-ray scatter. Image signal-to-noise ratio is calculated to be equal to conventional fluoroscopic systems at only 12% of the patient exposure and 25% of the staff exposure. Exposure reduction is achieved by elimination of detected scatter, elimination of the anti-scatter grid, increased detector DQE, and increased patient entrance area.


Medical Physics | 2006

Comparison of entrance exposure and signal-to-noise ratio between an SBDX prototype and a wide-beam cardiac angiographic system

Michael A. Speidel; Brian P. Wilfley; Josh Star-Lack; Joseph A. Heanue; Timothy D. Betts; Michael S. Van Lysel

The scanning-beam digital x-ray (SBDX) system uses an inverse geometry, narrow x-ray beam, and a 2-mm thick CdTe detector to improve the dose efficiency of the coronary angiographic procedure. Entrance exposure and large-area iodine signal-to-noise ratio (SNR) were measured with the SBDX prototype and compared to that of a clinical cardiac interventional system with image intensifier (II) and charge coupled device (CCD) camera (Philips H5000, MRC-200 x-ray tube, 72 kWp max). Phantoms were 18.6-35.0 cm acrylic with an iohexol-equivalent disk placed at midthickness (35 mg/cm2 iodine radiographic density). Imaging was performed at 15 frame/s, with the disk at mechanical isocenter and an 11-cm object-plane field width. The II/CCD system was operated in cine mode with automatic exposure control. With the SBDX prototype at maximum x-ray output (120 kVp, 24.3 kWp), the SBDX SNR was 107%-69% of the II/CCD SNR, depending on phantom thickness, and the SBDX entrance exposure rate was 10.7-9.3 R/min (9.4-8.2 cGy/min air kerma). For phantoms where an equal-kVp imaging comparison was possible (> or = 23.3 cm), the SBDX SNR ranged from 47% to 69% of the II/CCD SNR while delivering 6% to 9% of the II/CCD entrance exposure rate. From these measurements it was determined that the relative SBDX entrance exposure at equal SNR would be 31%-16%. Results were consistent with a model for relative entrance exposure at equal SNR, which predicted a 3-7 times reduction in entrance exposure due to SBDXs comparatively low scatter fraction (5.5%-8.1% measured, including off-focus radiation), high detector detective quantum efficiency (66%-73%, measured from 70 to 120 kVp), and large entrance field area (1.7x - 2.3x, for the same object-plane field width). With improvements to the system geometry, detector, and x-ray source, SBDX technology is projected to achieve conventional cine-quality SNR over a full range of patient thicknesses, with 5-10 times lower skin dose.


Medical Physics | 2013

The feasibility of an inverse geometry CT system with stationary source arrays.

Scott S. Hsieh; Joseph A. Heanue; Tobias Funk; Waldo Hinshaw; Brian P. Wilfley; Edward G. Solomon; Norbert J. Pelc

PURPOSE Inverse geometry computed tomography (IGCT) has been proposed as a new system architecture that combines a small detector with a large, distributed source. This geometry can suppress cone-beam artifacts, reduce scatter, and increase dose efficiency. However, the temporal resolution of IGCT is still limited by the gantry rotation time. Large reductions in rotation time are in turn difficult due to the large source array and associated power electronics. We examine the feasibility of using stationary source arrays for IGCT in order to achieve better temporal resolution. We anticipate that multiple source arrays are necessary, with each source array physically separated from adjacent ones. METHODS Key feasibility issues include spatial resolution, artifacts, flux, noise, collimation, and system timing clashes. The separation between the different source arrays leads to missing views, complicating reconstruction. For the special case of three source arrays, a two-stage reconstruction algorithm is used to estimate the missing views. Collimation is achieved using a rotating collimator with a small number of holes. A set of equally spaced source spots are designated on the source arrays, and a source spot is energized when a collimator hole is aligned with it. System timing clashes occur when multiple source spots are scheduled to be energized simultaneously. We examine flux considerations to evaluate whether sufficient flux is available for clinical applications. RESULTS The two-stage reconstruction algorithm suppresses cone-beam artifacts while maintaining resolution and noise characteristics comparable to standard third generation systems. The residual artifacts are much smaller in magnitude than the cone-beam artifacts eliminated. A mathematical condition is given relating collimator hole locations and the number of virtual source spots for which system timing clashes are avoided. With optimization, sufficient flux may be achieved for many clinical applications. CONCLUSIONS IGCT with stationary source arrays could be an imaging platform potentially capable of imaging a complete 16-cm thick volume within a tenth of a second.


Medical Physics | 2001

Comparison of vessel contrast measured with a scanning-beam digital x-ray system and an image intensifier/television system.

Michael A. Speidel; Brian P. Wilfley; Joseph A. Heanue; Timothy D. Betts; Michael S. Van Lysel

Vessel contrast was measured in the fluoroscopic images produced by a scanning-beam digital x-ray (SBDX) system and an image intensifier/television (II/TV) based system. The SBDX system electronically scans a series of pencil x-ray beams across the patient, each of which is directed at a distant small-area detector array. The reduction in detected scatter achieved with this geometry was expected to provide an increase in image contrast. Vessel contrast was evaluated from images of a phantom containing iodinated tubes. The vessels were inserted into an acrylic stack to provide a patient-mimicking scattering medium. Vessel diameter ranged from 0.3 to 3.1 mm. Images were acquired at 100 kVp with the SBDX and II/TV systems and averaged to reduce x-ray noise. The II/TV system was operated in the 6-in. image intensifier mode with an anti-scatter grid. The increase in contrast in the SBDX images, expressed as a ratio of the measured SBDX and II/TV contrasts, ranged from 1.63 to 1.79 for individual vessels. This agreed well with a prediction of the contrast improvement ratio for this experiment, based on measurements of the scatter fraction, object-plane line spread functions, and consideration of the source spectrum and detector absorption properties. The predicted contrast improvement ratio for SBDX relative to II/TV images was 1.62 to 1.77.


Medical Imaging 1997: Physics of Medical Imaging | 1997

Performance assessment of the scanning beam digital x-ray (SBDX) system

Michael S. Van Lysel; Edward G. Solomon; Brian P. Wilfley; April Dutta; Michael A. Speidel

A prototype scanning-beam digital x-ray (SBDX) system for cardiac fluoroscopy has been constructed. The unique geometry and absence of detected x-ray scatter in the SBDX image promises to provide image quality equivalent to a conventional image-intensifier-based fluoroscopic system at substantially reduced x-ray exposure to patient and staff. In order to measure the SBDX exposure advantage, a contrast- detail study was performed comparing SBDX and a conventional cardiac fluoroscopic system. Low-contrast deductibility as a function of the phantom entrance exposure was determined. The expected SBDX exposure advantage was 3.0 to 3.4, for low-contrast objects ranging in diameter from 2 to 10 mm. This exposure advantage is applicable to the AP projection through an average-size cardiac patient. Based on these results, calculations show that angulated views and larger patients will experience significantly greater exposure reductions. In addition, the results also indicate that SBDX system design modifications can provide a greater exposure reduction from that measured with this prototype.


Medical Physics | 2017

Feasibility of external beam radiation therapy to deep‐seated targets with kilovoltage x‐rays

Magdalena Bazalova-Carter; Michael D. Weil; Dylan Y. Breitkreutz; Brian P. Wilfley; Edward E. Graves

Purpose: Radiation therapy to deep‐seated targets is typically delivered with megavoltage x‐ray beams generated by medical linear accelerators or 60Co sources. Here, we used computer simulations to design and optimize a lower energy kilovoltage x‐ray source generating acceptable dose distributions to a deep‐seated target. Methods: The kilovoltage arc therapy (KVAT) x‐ray source was designed to treat a 4‐cm diameter target located at a 10‐cm depth in a 40‐cm diameter homogeneous cylindrical phantom. These parameters were chosen as an example of a clinical scenario for testing the performance of the kilovoltage source. A Monte Carlo (MC) model of the source was built in the EGSnrc/BEAMnrc code and source parameters, such as beam energy, tungsten anode thickness, beam filtration, number of collimator holes, collimator hole size and thickness, and source extent were varied. Dose to the phantom was calculated in the EGSnrc/DOSXYZnrc code for varying treatment parameters, such as the source‐to‐axis distance and the treatment arc angle. The quality of dose distributions was quantified by means of target‐to‐skin ratio and dose output expressed in D50 (50% isodose line) for a 30‐min irradiation in the homogeneous phantom as well as a lung phantom. Additionally, a patient KVAT dose distribution to a left pararenal lesion (˜1.6 cm in diameter) was calculated and compared to a 15 MV volumetric modulated arc therapy (VMAT) plan. Results: In the design of the KVAT x‐ray source, the beam energy, beam filtration, collimator hole size, source‐to‐isocenter distance, and treatment arc had the largest effect on the source output and the quality of dose distributions. For the 4‐cm target at 10‐cm depth, the optimized KVAT dose distribution generated a conformal plan with target‐to‐skin ratio of 5.1 and D50 in 30 min of 24.1 Gy in the homogeneous phantom. In the lung phantom, a target‐to‐skin ratio of 7.5 and D50 in 30 min of 25.3 Gy were achieved. High dose conformity of the 200 kV KVAT left pararenal plan was comparable to the 15 MV VMAT plan. The volume irradiated to at least 10% (<240 cGy) of the prescription dose was 2.2 × larger in the 200 kV KVAT plan than in the 15 MV VMAT plan, but considered clinically insignificant. Conclusions: This study demonstrated that conformal treatments of deep‐seated targets were achievable with kilovoltage x‐rays with dose distributions comparable to MV beams. However, due to the larger volumes irradiated to clinically tolerated low doses, KVAT x‐ray source usage for deep‐seated lesions will be further evaluated to determine optimal target size.


Medical Physics | 2012

Feasibility of low-dose single-view 3D fiducial tracking concurrent with external beam delivery.

Michael A. Speidel; Brian P. Wilfley; A Hsu; Dimitre Hristov

PURPOSE In external-beam radiation therapy, existing on-board x-ray imaging chains orthogonal to the delivery beam cannot recover 3D target trajectories from a single view in real-time. This limits their utility for real-time motion management concurrent with beam delivery. To address this limitation, the authors propose a novel concept for on-board imaging based on the inverse-geometry Scanning-Beam Digital X-ray (SBDX) system and evaluate its feasibility for single-view 3D intradelivery fiducial tracking. METHODS A chest phantom comprising a posterior wall, a central lung volume, and an anterior wall was constructed. Two fiducials were placed along the mediastinal ridge between the lung cavities: a 1.5 mm diameter steel sphere superiorly and a gold cylinder (2.6 mm length × 0.9 mm diameter) inferiorly. The phantom was placed on a linear motion stage that moved sinusoidally. Fiducial motion was along the source-detector (z) axis of the SBDX system with ±10 mm amplitude and a programmed period of either 3.5 s or 5 s. The SBDX system was operated at 15 frames per second, 100 kVp, providing good apparent conspicuity of the fiducials. With the stage moving, detector data were acquired and subsequently reconstructed into 15 planes with a 12 mm plane-to-plane spacing using digital tomosynthesis. A tracking algorithm was applied to the image planes for each temporal frame to determine the position of each fiducial in (x,y,z)-space versus time. A 3D time-sinusoidal motion model was fit to the measured 3D coordinates and root mean square (RMS) deviations about the fitted trajectory were calculated. RESULTS Tracked motion was sinusoidal and primarily along the source-detector (z) axis. The RMS deviation of the tracked z-coordinate ranged from 0.53 to 0.71 mm. The motion amplitude derived from the model fit agreed with the programmed amplitude to within 0.28 mm for the steel sphere and within -0.77 mm for the gold seed. The model fit periods agreed with the programmed periods to within 7%. CONCLUSIONS Three dimensional fiducial tracking with approximately 1 mm or better accuracy and precision appears to be feasible with SBDX, supporting its use to guide radiotherapy.


Proceedings of SPIE | 2011

Real-time scanning beam digital x-ray image guidance system for transbronchial needle biopsy

Sungwon Yoon; Brian P. Wilfley; Keith Jasperson; Ganesh Krishna; Rebecca Fahrig

We investigate a real-time digital tomosynthesis (DTS) imaging modality, based on the scanning beam digital x-ray (SBDX) hardware, used in conjunction with an electromagnetic navigation bronchoscopy (ENB) system to provide improved image guidance for minimally invasive transbronchial needle biopsy (TBNbx). Because the SBDX system source uses electron beams, steered by electromagnets, to generate x-rays, and the ENB system generates an electromagnetic field to localize and track steerable navigation catheters, the two systems will affect each other when operated in proximity. We first investigate the compatibility of the systems by measuring the ENB system localization error as a function of distance between the two systems. The SBDX system reconstructs DTS images, which provide depth information, and so we investigate the improvement in lung nodule visualization using SBDX system DTS images and compare them to fluoroscopic images currently used for biopsy verification. Target localization error remains below 2mm (or virtually error free) if the volume-of-interest (VOI) is at least 50cm away from the SBDX system source and detector. Inside this region, tomographic angle ranges from 3° to 10° depending on the VOI location. Improved lung nodule (≤ 20mm diameter) contrast is achieved by imaging the VOI near the SBDX system detector, where the tomographic angle is maximized. The combination of the SBDX image guidance with an ENB system would provide real-time visualization during biopsy with improved localization of the target and needle/biopsy instruments, thereby increasing the average and lowering the variance of the yield for TBNbx.


Medical Physics | 2010

TU‐E‐204B‐02: Feasibility of Low‐Dose Single‐View Real‐Time 3D Tracking Concurrent with External Beam Delivery

Brian P. Wilfley; Michael A. Speidel; A Hsu; Dimitre Hristov

Purpose: Existing on‐board x‐ray imaging chains orthogonal to the delivery beam cannot recover 3D target trajectories from a single view in real‐time. This limits their utility for real‐time motion management concurrent with beam delivery. We investigate the feasibility of an alternative inverse‐geometry “single‐view” Scanning Beam Digital X‐ray (SBDX, NovaRay Medical, Inc.) system for real‐time 3D intra‐delivery tracking. Method and Materials: A chest phantom (Standard Imaging, WI) comprising a posterior wall, a central lung volume and an anterior wall was used for the investigations. The lung contained a mediastinal volume isodense with the chest. Along the mediastinum, two fiducials were placed: a 1 mm diameter steel sphere superiorly and a gold cylinder (2.6 × 0.9 mm) inferiorly. The phantom was placed on a linear motion stage (Standard Imaging, WI) which moved sinusoidally with peak‐to‐peak displacement of 2 cm, and a period of either 3.5 sec or 5 sec. The stage motion was substantially along the source‐detector (z) axis of the SBDX system. The system operated at 15 frames per second, 100 kVp, providing good apparent conspicuity of the fiducials. With the stage moving, detector data were acquired and subsequently reconstructed using digital tomosynthesis into 15 planes with spacing of 12 mm. Tracking was performed on the plane data for each (temporal) frame to yield the position of each fiducial in 3‐space versus time. Results: Tracking data for the z‐coordinate agreed with the trajectory calculated from the known amplitude and frequency of the motion and with an adjustable phase and z‐offset to within an RMS error of 1.05 mm for the gold cylinder and 0.61 mm for the steel sphere. Conclusion: Tracking with sub‐millimeter accuracy appears to be feasible with “single‐view” SBDX, supporting its use to guide radiotherapy.Conflict of Interest: Brian Wilfley has a financial interest in NovaRay Medical, Inc.

Collaboration


Dive into the Brian P. Wilfley's collaboration.

Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Kate Leeann Bechtel

Mansfield University of Pennsylvania

View shared research outputs
Top Co-Authors

Avatar

Tobias Funk

University of California

View shared research outputs
Top Co-Authors

Avatar

Joseph A. Heanue

University of Wisconsin-Madison

View shared research outputs
Top Co-Authors

Avatar

Michael A. Speidel

University of Wisconsin-Madison

View shared research outputs
Top Co-Authors

Avatar

Michael S. Van Lysel

University of Wisconsin-Madison

View shared research outputs
Top Co-Authors

Avatar

Augustus P. Lowell

Mansfield University of Pennsylvania

View shared research outputs
Top Co-Authors

Avatar

John F. Heanue

Mansfield University of Pennsylvania

View shared research outputs
Top Co-Authors

Avatar
Researchain Logo
Decentralizing Knowledge