Emelie Adolfsson
Linköping University
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by Emelie Adolfsson.
Medical Physics | 2010
Emelie Adolfsson; Gudrun Alm Carlsson; Jan-Erik Grindborg; Håkan Gustafsson; Eva Lund; Åsa Carlsson Tedgren
PURPOSE To investigate experimentally the energy dependence of the detector response of lithium formate EPR dosimeters for photon energies below 1 MeV relative to that atC60o energies. High energy photon beams are used in calibrating dosimeters for use in brachytherapy since the absorbed dose to water can be determined with high accuracy in such beams using calibrated ion chambers and standard dosimetry protocols. In addition to any differences in mass-energy absorption properties between water and detector, variations in radiation yield (detector response) with radiation quality, caused by differences in the density of ionization in the energy imparted (LET), may exist. Knowledge of an eventual deviation in detector response with photon energy is important for attaining high accuracy in measured brachytherapy dose distributions. METHODS Lithium formate EPR dosimeters were irradiated to known levels of air kerma in 25-250 kV x-ray beams and inC137s and C60o beams at the Swedish Secondary Standards Dosimetry Laboratory. Conversions from air kerma free in air into values of mean absorbed dose to the detectors were made using EGSnrc MC simulations and x-ray energy spectra measured or calculated for the actual beams. The signals from the detectors were measured using EPR spectrometry. Detector response (the EPR signal per mean absorbed dose to the detector) relative to that for C60o was determined for each beam quality. RESULTS Significant decreases in the relative response ranging from 5% to 6% were seen for x-ray beams at tube voltages≤180kV. No significant reduction in the relative response was seen for C137s and 250 kV x rays. CONCLUSIONS When calibrated inC60o or MV photon beams, corrections for the photon energy dependence of detector response are needed to achieve the highest accuracy when using lithium formate EPR dosimeters for measuring absorbed doses around brachytherapy sources emitting photons in the energy range of 20-150 keV such as Y169b and electronic sources.
Radiation Protection Dosimetry | 2014
Eva Lund; Emelie Adolfsson; Natallia Kolbun; Håkan Gustafsson
A one-dimensional electron paramagnetic resonance (EPR) imaging method for visualisation of dose distributions in photon fields has been developed. Pressed pellets of potassium dithionate were homogeneously irradiated in a (60)Co radiation field to 600 Gy. The EPR analysis was performed with an X-Band (9.6 GHz) Bruker E540 EPR and EPR imaging spectrometer equipped with an E540 GC2X two-axis X-band gradient coil set with gradients along the y axis (along the sample tube) and z axis (along B0) and an ER 4108TMHS resonator. Image reconstruction, including deconvolution, baseline corrections and corrections for the resonator sensitivity, was performed using an in-house-developed Matlab code for the purpose to have a transparent and complete algorithm for image reconstruction. With this method, it is possible to visualise a dose distribution with an accuracy of ∼5 % within ±5 mm from the centre of the resonator.
Radiotherapy and Oncology | 2014
Emelie Adolfsson; Håkan Gustafsson; Eva Lund; Gudrun Alm Carlsson; Sara Olsson; Åsa Carlsson Tedgren
The aim of this work was to develop and test a remote end-to-end audit system using lithium formate EPR dosimeters. Four clinics were included in a pilot study, absorbed doses determined in the PTV agreed with TPS calculated doses within ±5% for 3D-CRT and ±7% (k=1) for IMRT/VMAT dose plans.
World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany | 2009
Emelie Adolfsson; Gudrun Alm Carlsson; Jan-Erik Grindborg; Håkan Gustafsson; Eva Lund; Åsa Carlsson Tedgren
After development of sensitive dosimeter materials Electron Paramagnetic Resonance EPR dosimetry has been successfully used also in radiation therapy. The intensity of the EPR-signal is a measure o ...
Physics in Medicine and Biology | 2015
Emelie Adolfsson; Shane White; Guillaume Landry; Eva Lund; Håkan Gustafsson; Frank Verhaegen; Brigitte Reniers; Åsa Carlsson Tedgren; Gudrun Alm Carlsson
Interest in high dose rate (HDR) electronic brachytherapy operating at 50 kV is increasing. For quality assurance it is important to identify dosimetry systems that can measure the absorbed doses in absolute terms which is difficult in this energy region. In this work a comparison is made between two dosimetry systems, EPR lithium formate dosimeters and radiochromic EBT2 film. Both types of dosimeters were irradiated simultaneously in a PMMA phantom using the Axxent EBS. Absorbed dose to water was determined at distances of 10 mm, 30 mm and 50 mm from the EBS. Results were traceable to different primary standards as regards to absorbed dose to water (EPR) and air kerma (EBT2). Monte Carlo simulations were used in absolute terms as a third estimate of absorbed dose to water. Agreement within the estimated expanded (k = 2) uncertainties (5% (EPR), 7% (EBT2)) was found between the results at 30 mm and 50 mm from the x-ray source. The same result was obtained in 4 repetitions of irradiation, indicating high precision in the measurements with both systems. At all distances, agreement between EPR and Monte Carlo simulations was shown as was also the case for the film measurements at 30mm and 50mm. At 10mm the geometry for the film measurements caused too large uncertainty in measured values depending on the exact position (within sub-mm distances) of the EBS and the 10 mm film results were exculded from comparison. This work has demonstrated good performance of the lithium formate EPR dosimetry system in accordance with earlier experiments at higher photon energies ((192)Ir HDR brachytherapy). It was also highlighted that there might be issues regarding the energy dependence and intrinsic efficiency of the EBT2 film that need to be considered for measurements using low energy sources.
Radiation Protection Dosimetry | 2014
Natallia Kolbun; Emelie Adolfsson; Håkan Gustafsson; Eva Lund
Electron paramagnetic resonance imaging (EPRI) was performed to visualise 2D dose distributions of homogenously irradiated potassium dithionate tablets and to demonstrate determination of 1D dose profiles along the height of the tablets. Mathematical correction was applied for each relative dose profile in order to take into account the inhomogeneous response of the resonator using X-band EPRI. The dose profiles are presented with the spatial resolution of 0.6 mm from the acquired 2D images; this value is limited by pixel size, and 1D dose profiles from 1D imaging with spatial resolution of 0.3 mm limited by the intrinsic line-width of potassium dithionate. In this paper, dose profiles from 2D reconstructed electron paramagnetic resonance (EPR) images using the Xepr software package by Bruker are focussed. The conclusion is that using potassium dithionate, the resolution 0.3 mm is sufficient for mapping steep dose gradients if the dosemeters are covering only ±2 mm around the centre of the resonator.
Radiation Measurements | 2011
P. Fattibene; A. Wieser; Emelie Adolfsson; L.A. Benevides; Maria Brai; Freddy Callens; V. Chumak; B. Ciesielski; S. Della Monaca; K. Emerich; Håkan Gustafsson; Y. Hirai; Masaharu Hoshi; Axel Israelsson; Alexander Ivannikov; D. Ivanov; J. Kaminska; Wu Ke; Eva Lund; Maurizio Marrale; L. Martens; Chuzou Miyazawa; N. Nakamura; W. Panzer; S. Pivovarov; Ricardo A. Reyes; Mohd Rodzi; A. Romanyukha; A. Rukhin; S. Sholom
Physics in Medicine and Biology | 2012
Emelie Adolfsson; Mattias Karlsson; Gudrun Alm Carlsson; Åsa Carlsson Tedgren; Eva Lund; Sara Olsson; Håkan Gustafsson
Radiation Measurements | 2014
Emelie Adolfsson; Åsa Carlsson Tedgren; Gudrun Alm Carlsson; Håkan Gustafsson; Eva Lund
Radiotherapy and Oncology | 2016
Emelie Adolfsson; P. Wesolowska; J. Izewska; Eva Lund; M. Olsson; A. Carlsson Tedgren