Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Jakob Helt-Hansen is active.

Publication


Featured researches published by Jakob Helt-Hansen.


Medical Physics | 2014

Detector to detector corrections: a comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams

Godfrey Azangwe; P. Grochowska; Dietmar Georg; Joanna Izewska; Johannes Hopfgartner; Wolfgang Lechner; Claus E. Andersen; Anders Ravnsborg Beierholm; Jakob Helt-Hansen; Hideyuki Mizuno; Akifumi Fukumura; Kaori Yajima; C. Gouldstone; Peter Sharpe; Ahmed Meghzifene; Hugo Palmans

PURPOSE The aim of the present study is to provide a comprehensive set of detector specific correction factors for beam output measurements for small beams, for a wide range of real time and passive detectors. The detector specific correction factors determined in this study may be potentially useful as a reference data set for small beam dosimetry measurements. METHODS Dose response of passive and real time detectors was investigated for small field sizes shaped with a micromultileaf collimator ranging from 0.6 × 0.6 cm(2) to 4.2 × 4.2 cm(2) and the measurements were extended to larger fields of up to 10 × 10 cm(2). Measurements were performed at 5 cm depth, in a 6 MV photon beam. Detectors used included alanine, thermoluminescent dosimeters (TLDs), stereotactic diode, electron diode, photon diode, radiophotoluminescent dosimeters (RPLDs), radioluminescence detector based on carbon-doped aluminium oxide (Al2O3:C), organic plastic scintillators, diamond detectors, liquid filled ion chamber, and a range of small volume air filled ionization chambers (volumes ranging from 0.002 cm(3) to 0.3 cm(3)). All detector measurements were corrected for volume averaging effect and compared with dose ratios determined from alanine to derive a detector correction factors that account for beam perturbation related to nonwater equivalence of the detector materials. RESULTS For the detectors used in this study, volume averaging corrections ranged from unity for the smallest detectors such as the diodes, 1.148 for the 0.14 cm(3) air filled ionization chamber and were as high as 1.924 for the 0.3 cm(3) ionization chamber. After applying volume averaging corrections, the detector readings were consistent among themselves and with alanine measurements for several small detectors but they differed for larger detectors, in particular for some small ionization chambers with volumes larger than 0.1 cm(3). CONCLUSIONS The results demonstrate how important it is for the appropriate corrections to be applied to give consistent and accurate measurements for a range of detectors in small beam geometry. The results further demonstrate that depending on the choice of detectors, there is a potential for large errors when effects such as volume averaging, perturbation and differences in material properties of detectors are not taken into account. As the commissioning of small fields for clinical treatment has to rely on accurate dose measurements, the authors recommend the use of detectors that require relatively little correction, such as unshielded diodes, diamond detectors or microchambers, and solid state detectors such as alanine, TLD, Al2O3:C, or scintillators.


Medical Physics | 2009

Characterization of a fiber-coupled Al2O3:C luminescence dosimetry system for online in vivo dose verification during 192Ir brachytherapy

Claus E. Andersen; S.K. Nielsen; Steffen Greilich; Jakob Helt-Hansen; Jacob Christian Lindegaard; Kari Tanderup

A prototype of a new dose-verification system has been developed to facilitate prevention and identification of dose delivery errors in remotely afterloaded brachytherapy. The system allows for automatic online in vivo dosimetry directly in the tumor region using small passive detector probes that fit into applicators such as standard needles or catheters. The system measures the absorbed dose rate (0.1 s time resolution) and total absorbed dose on the basis of radioluminescence (RL) and optically stimulated luminescence (OSL) from aluminum oxide crystals attached to optical fiber cables (1 mm outer diameter). The system was tested in the range from 0 to 4 Gy using a solid-water phantom, a Varian GammaMed Plus 192Ir PDR afterloader, and dosimetry probes inserted into stainless-steel brachytherapy needles. The calibrated system was found to be linear in the tested dose range. The reproducibility (one standard deviation) for RL and OSL measurements was 1.3%. The measured depth-dose profiles agreed well with the theoretical expectations computed with the EGSNRC Monte Carlo code, suggesting that the energy dependence for the dosimeter probes (relative to water) is less than 6% for source-to-probe distances in the range of 2-50 mm. Under certain conditions, the RL signal could be greatly disturbed by the so-called stem signal (i.e., unwanted light generated in the fiber cable upon irradiation). The OSL signal is not subject to this source of error. The tested system appears to be adequate for in vivo brachytherapy dosimetry.


Medical Physics | 2009

Characterization of a fiber-coupled Al{sub 2}O{sub 3}:C luminescence dosimetry system for online in vivo dose verification during {sup 192}Ir brachytherapy

Claus E. Andersen; Soeren Kynde Nielsen; Steffen Greilich; Jakob Helt-Hansen; Jacob Christian Lindegaard

A prototype of a new dose-verification system has been developed to facilitate prevention and identification of dose delivery errors in remotely afterloaded brachytherapy. The system allows for automatic online in vivo dosimetry directly in the tumor region using small passive detector probes that fit into applicators such as standard needles or catheters. The system measures the absorbed dose rate (0.1 s time resolution) and total absorbed dose on the basis of radioluminescence (RL) and optically stimulated luminescence (OSL) from aluminum oxide crystals attached to optical fiber cables (1 mm outer diameter). The system was tested in the range from 0 to 4 Gy using a solid-water phantom, a Varian GammaMed Plus 192Ir PDR afterloader, and dosimetry probes inserted into stainless-steel brachytherapy needles. The calibrated system was found to be linear in the tested dose range. The reproducibility (one standard deviation) for RL and OSL measurements was 1.3%. The measured depth-dose profiles agreed well with the theoretical expectations computed with the EGSNRC Monte Carlo code, suggesting that the energy dependence for the dosimeter probes (relative to water) is less than 6% for source-to-probe distances in the range of 2-50 mm. Under certain conditions, the RL signal could be greatly disturbed by the so-called stem signal (i.e., unwanted light generated in the fiber cable upon irradiation). The OSL signal is not subject to this source of error. The tested system appears to be adequate for in vivo brachytherapy dosimetry.


Acta Oncologica | 2009

Medical reference dosimetry using EPR measurements of alanine: development of an improved method for clinical dose levels.

Jakob Helt-Hansen; Flemming Rosendal; Inger Matilde Kofoed; Claus E. Andersen

Background. Electron spin resonance (EPR) is used to determine the absorbed dose of alanine dosimeters exposed to clinical photon beams in a solid-water phantom. Alanine is potentially suitable for medical reference dosimetry, because of its near water equivalence over a wide energy spectrum, low signal fading, non-destructive measurement and small dosimeter size. Material and Methods. A Bruker EMX-micro EPR spectrometer with a rectangular cavity and a measurement time of two minutes per dosimeter was used for reading of irradiated alanine dosimeters. Under these conditions a new algorithm based on scaling of known spectra was developed to extract the alanine signal. Results. The dose accuracy, including calibration uncertainty, is less than 2% (k=1) above 4 Gy (n=4). The measurement uncertainty is fairly constant in absolute terms (∼30 mGy) and the relative uncertainty therefore rises for dose measurements below 4 Gy. Typical reproducibility is <1% (k=1) above 10 Gy and <2% between 4 and 10 Gy. Below 4 Gy the uncertainty is higher. A depth dose curve measurement was performed in a solid-water phantom irradiated to a dose of 20 Gy at the maximum dose point (dmax) in 6 and 18 MV photon beams. The typical difference between the dose measured with alanine in solid water and the dose measured with an ion chamber in a water tank was about 1%. A difference of 2% between 6 and 18 MV was found, possibly due to non-water equivalence of the applied phantom. Discussion. Compared to previously published methods the proposed algorithm can be applied without normalisation of phase shifts caused by changes in the g-value of the cavity. The study shows that alanine dosimetry is a suitable candidate for medical reference dosimetry especially for quality control applications.


Proceedings of SPIE | 2014

Optical surface scanning for respiratory motion monitoring in radiotherapy: a feasibility study

Susanne Lise Bekke; Faisal Mahmood; Jakob Helt-Hansen; Claus F. Behrens

Purpose. We evaluated the feasibility of a surface scanning system (Catalyst) for respiratory motion monitoring of breast cancer patients treated with radiotherapy in deep inspiration breath-hold (DIBH). DIBH is used to reduce the radiation dose to the heart and lung. In contrast to RPM, a competing marker-based system, Catalyst does not require an objectmarker on the patient’s skin. Materials and Methods. Experiment 1: a manikin was used to simulate sinusoidal breathing. The amplitude, period and baseline (signal value at end-expiration) were estimated with RPM and Catalyst. Experiment 2 and 3: the Quasar phantom was used to study if the angle of the monitored surface affects the amplitude of the recorded signal. Results. Experiment 1: we observed comparable period estimates for both systems. The amplitudes were 8 ± 0.1 mm (Catalyst) and 4.9 ± 0.1 mm (RPM). Independent check with in-room lasers showed an amplitude of approximately 8 mm, supporting Catalyst measurements. Large baseline errors were seen with RPM. Experiment 2: RPM underestimated the amplitude if the object-marker was angled during vertical motion. This result explains the amplitude underestimation by RPM seen in Experiment 1. Experiment 3: an increased (fixed) surface angle during breathing motion resulted in an overestimated amplitude with RPM, while the amplitude estimated by Catalyst was unaffected. Conclusion. Our study showed that Catalyst can be used as a better alternative to the RPM. With Catalyst, the amplitude estimates are more accurate and do not depend on the angle of the tracked surface, and the baseline errors are smaller.


Radiation Physics and Chemistry | 2004

RisøScan—a new dosimetry software

Jakob Helt-Hansen; Arne Miller


Radiation Physics and Chemistry | 2005

Dose response of thin-film dosimeters irradiated with 80–120 keV electrons

Jakob Helt-Hansen; Arne Miller; Peter Sharpe


Radiation Physics and Chemistry | 2010

A comparison of the microbicidal effectiveness of gamma rays and high and low energy electron radiations

A. Tallentire; Arne Miller; Jakob Helt-Hansen


Radiation Physics and Chemistry | 2010

Dμ—A new concept in industrial low-energy electron dosimetry

Jakob Helt-Hansen; Arne Miller; Peter Sharpe; Bengt Laurell; Doug Weiss; Gary Pageau


Radiation Physics and Chemistry | 2004

Calibration of thin-film dosimeters irradiated with 80-120 kev electrons

Jakob Helt-Hansen; Arne Miller; M McEwen; Peter Sharpe; Simon Duane

Collaboration


Dive into the Jakob Helt-Hansen's collaboration.

Top Co-Authors

Avatar

Arne Miller

Technical University of Denmark

View shared research outputs
Top Co-Authors

Avatar

Claus E. Andersen

Technical University of Denmark

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Peter Sharpe

National Physical Laboratory

View shared research outputs
Top Co-Authors

Avatar

C.F. Behrens

Copenhagen University Hospital

View shared research outputs
Top Co-Authors

Avatar

S.J. Zimmermann

Odense University Hospital

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Patrik Sibolt

Technical University of Denmark

View shared research outputs
Top Co-Authors

Avatar

Rickard O. Cronholm

Copenhagen University Hospital

View shared research outputs
Researchain Logo
Decentralizing Knowledge