Mary R. Olsovsky
University of Maryland, Baltimore
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by Mary R. Olsovsky.
Journal of the American College of Cardiology | 1998
Michael R. Gold; Mary R. Olsovsky; Michael Pelini; Robert W. Peters; Stephen R. Shorofsky
OBJECTIVES The purpose of this study was to compare defibrillation thresholds with lead systems consisting of an active left pectoral electrode and either single or dual transvenous coils. BACKGROUND Lead systems that include an active pectoral pulse generator reduce defibrillation thresholds and permit transvenous defibrillation in nearly all patients. A further improvement in defibrillation efficacy is desirable to allow for smaller pulse generators with a reduced maximal output. METHODS This prospective study was performed in 50 consecutive patients. Each patient was evaluated with two lead configurations with the order of testing randomized. Shocks were delivered between the right ventricular coil and either an active can alone (single coil) or an active can with the proximal atrial coil (dual coil). The right ventricular coil was the cathode for the first phase of the biphasic defibrillation waveform. RESULTS Delivered energy at the defibrillation threshold was 10.1+/-5.0 J for the single-coil configuration and 8.7+/-4.0 J for the dual-coil configuration (p < 0.02). Moreover, 98% of patients had low (<15 J) thresholds with the dual-coil lead system, compared with 88% of patients with the single-coil configuration (p=0.05). Leading edge voltage (p < 0.001) and shock impedance (p < 0.001) were also decreased with the dual-coil configuration, although peak current was increased (p < 0.001). CONCLUSIONS A dual-coil, active pectoral lead system reduces defibrillation energy requirements compared with a single-coil, unipolar configuration.
Pacing and Clinical Electrophysiology | 2002
Denice M. Hodgson; Mary R. Olsovsky; Stephen R. Shorofsky; Barry Daly; Michael R. Gold
HODGSON, D.M., et al.: Clinical Predictors of Defibrillation Thresholds with an Active Pectoral Pulse Generator Lead System. Active pectoral pulse generators are used routinely for initial ICD placement because they reduce DFTs and simplify the implantation procedure. Despite the common use of these systems, little is known regarding the clinical predictors of defibrillation efficacy with active pulse generator lead configurations. Such predictors would be helpful to identify patients likely to require higher output devices or more complicated implantations. This was a prospective evaluation of DFT using a uniform testing protocol in 102 consecutive patients with an active pectoral can and dual coil transvenous lead. For each patient, the DFT was measured with a step‐down protocol. In addition, 34 parameters were assessed including standard clinical echocardiographic and radiographic measures. Multivariate stepwise regression analysis was performed to identify independent predictors of the DFT. The mean DFT was 9.3 ± 4.6 J and 93% (95/102) of patients had a DFT ≤ 15 J. The QRS duration, interventricular septum thickness, left ventricular mass, and mass index were significant but weak (R < 0.3) univariate predictors of DFT. The left ventricular mass was the only independent predictor by multivariate analysis, but this parameter accounted for < 5% of the variability of DFT measured (adjusted R2= 0.047, P = 0.017). The authors concluded that an acceptable DFT (< 15 J) is observed in > 90% of patients with this dual coil and active pectoral can lead system. Clinical factors are of limited use for predicting DFTs and identifying those patients who will have high thresholds.
American Journal of Cardiology | 2000
Michael R. Gold; Randall Brockman; Robert W. Peters; Mary R. Olsovsky; Stephen R. Shorofsky
The hemodynamic effects of pacing in patients with congestive heart failure (CHF) remain controversial. Early studies reported that pacing from the right ventricular (RV) apex improved acute hemodynamic parameters in patients with left ventricular systolic dysfunction, but these findings were not confirmed in subsequent controlled studies. More recently, it has been proposed that pacing from the RV side of the ventricular septum improves hemodynamic function compared with intrinsic conduction or apical pacing. Either dual-chamber or ventricular pacing have been evaluated, again with inconsistent findings. To assess the effects of pacing site and mode on acute hemodynamic function, we evaluated 21 subjects with CHF and intrinsic conduction disease. Hemodynamics were compared in AAI, VVI, and DDD modes with pacing from the RV apex or high septum. The pacing rate was constant in each patient and the order of testing was randomized. In the absence of ventricular pacing (AAI mode), the mean systemic arterial pressure was 85 +/- 11 mm Hg, the right atrial pressure was 11 +/- 4 mm Hg, the pulmonary capillary wedge pressure was 18 +/- 8 mm Hg and the cardiac index was 2.4 +/- 0.7 L/min/m(2). Compared with AAI pacing, there were no improvements in any hemodynamic parameter with DDD pacing from either RV site. Hemodynamic function worsened with VVI pacing from both RV sites. Subgroup analyses of patients with dilated cardiomyopathy, with prolonged PR interval, or with significant mitral regurgitation also failed to demonstrate an improvement with pacing. We conclude that pacing mode but not RV pacing site affects acute hemodynamic function. Pacing in the DDD mode prevents the deleterious effects of VVI pacing in this patient population.
Journal of Cardiovascular Electrophysiology | 1998
Mary R. Olsovsky; Stephen R. Shorofsky; Michael R. Gold
Polarity and Biphasic Defibrillation Thresholds. Introduction: The downsizing of implantable defibrillator pulse generators has made pectoral placement routine. A further reduction of defibrillation thresholds (DFTs) may simplify implantation defibrillation testing and allow for smaller, lower output pulse generators while maintaining an adequate defibrillation safety margin. One factor that may affect defibrillation efficacy is shock polarity.
Pacing and Clinical Electrophysiology | 1999
Michael R. Gold; William Hsu; Alan F. Marcovecchio; Mary R. Olsovsky; Douglas J. Lang; Stephen R. Shorofsky
Inappropriate therapies delivered by implantable cardioverter defibrillators (ICDs) for supraventricular arrhythmias remain a common problem, particularly in the event of rapidly conducted atrial fibrillation or marked sinus tachycardia. The ability to differentiate between ventricular tachycardia and supraventricular arrhythmias is the major goal of discrimination algorithms. Therefore, we developed a new algorithm, SimDis, utilizing morphological features of the shocking electrograms. This algorithm was developed from electrogram data obtained from 36 patients undergoing ICD implantation. An independent test set was evaluated in 25 patients. Recordings were made in sinus rhythm, sinus tachycardia, and following the induction of ventricular tachycardia and atrial fibrillation. The arrhythmia complex is defined as wide if the duration is at least 30% greater than the template in sinus rhythm. For narrow complexes, four maximum and minimum values were measured to form a 4‐element feature vector, which was compared with a representative feature vector during normal sinus rhythm. For each rhythm, any wide complex was classified as ventricular tachycardia. For narrow complexes, the second step of the algorithm compared the electrogram with the template, computing similarity and dissimilarity values. These values were then mapped to determine if they fell within a previously established discrimination boundary. On the independent test set, the SimDis algorithm correctly classified 100% of ventricular tachycardias (27/27), 98% of sinus tachycardias (54/55), and 100% of episodes of atrial fibrillation (37/37). We conclude that the SimDis algorithm yields high sensitivity (100%) and specificity (99%) for arrhythmia discrimination, using the computational capabilities of an ICD system.
American Journal of Cardiology | 1997
Mary R. Olsovsky; Denice M. Hodgson; Stephen R. Shorofsky; Neal G. Kavesh; Michael R. Gold
This study is a prospective, randomized comparison of monophasic and biphasic defibrillation thresholds in 19 patients with a single transvenous lead. Despite using reverse polarity and optimal tilts for the monophasic waveform, the defibrillation threshold was reduced with biphasic shocks from 15.8 +/- 11.3 to 11.5 +/- 6.1 (p <0.05) with comparable reductions of leading edge voltage and current.
Journal of Cardiovascular Electrophysiology | 1998
Mary R. Olsovsky; Michael A. Peltni; Stephen R. Shoroesky; Michael R. Gold
Stability of Defibrillation Thresholds. Introduction: Monophasic defibrillation thresholds rise over time with a variety of lead systems. These chronic changes are attenuated or eliminated by biphasic waveforms, although the effect appears dependent upon the lead system. With the downsizing of pulse generator size to allow for routine pectoral implantation, active can lead systems have now become standard. However, the temporal stability of such lead systems has not been evaluated previously.
Journal of the American College of Cardiology | 1999
Michael Cooklin; Mary R. Olsovsky; Randall G Brockman; Stephen R. Shorofsky; Michael R. Gold
OBJECTIVES The purpose of this study was to compare transvenous atrial defibrillation thresholds with lead configurations consisting of an active left pectoral electrode and either single or dual transvenous coils. BACKGROUND Low atrial defibrillation thresholds are achieved using complex lead systems including coils in the coronary sinus. However, the efficacy of more simple ventricular defibrillation leads with active pectoral pulse generators to defibrillate atrial fibrillation (AF) is unknown. METHODS This study was a prospective, randomized assessment of shock configuration on atrial defibrillation thresholds in 32 patients. The lead system was a dual coil Endotak DSP lead with a left pectoral pulse generator emulator. Shocks were delivered either between the right ventricular coil and an active can in common with the proximal atrial coil (triad) or between the atrial coil and active can (transatrial). RESULTS Delivered energy at defibrillation threshold was 7.1 +/- 6.0 J in the transatrial configuration and 4.0 +/- 4.2 J in the triad configuration (p < 0.005). Moreover, a low threshold (< or = 3 J) was observed in 69% of subjects in the triad configuration but only 47% in the transatrial configuration. Peak voltage and shock impedance were also lowered significantly in the triad configuration. Left atrial size was the only clinical predictor of the defibrillation threshold (r = 0.57, p < 0.002). CONCLUSIONS These results indicate that low atrial defibrillation thresholds can be achieved using a single-pass transvenous ventricular defibrillation lead with a conventional ventricular defibrillation pathway. These data support the development of the combined atrial and ventricular defibrillator system.
Journal of the American College of Cardiology | 2001
Eric J. Rashba; Mary R. Olsovsky; Stephen R. Shorofsky; Malcolm Kirk; Robert W. Peters; Michael R. Gold
OBJECTIVES The objective of this study was to characterize temporal changes in defibrillation thresholds (DFTs) after implantation with an active pectoral, dual-coil transvenous lead system. BACKGROUND Ventricular DFTs rise over time when monophasic waveforms are used with non-thoracotomy lead systems. This effect is attenuated when biphasic waveforms are used with transvenous lead systems; however, significant increases in DFT still occur in a minority of patients. The long-term stability of DFTs with contemporary active pectoral lead systems is unknown. METHODS This study was a prospective assessment of temporal changes in DFT using a uniform testing algorithm, shock polarity and dual-coil active pectoral lead system. Thresholds were measured at implantation, before discharge and at long-term follow-up (70 +/- 40 weeks) in 50 patients. RESULTS The DFTs were 9.2 +/- 5.4 J at implantation, 8.3 +/- 5.8 J before discharge and 6.9 +/- 3.6 J at long-term follow-up (p < 0.01 by analysis of variance; p < 0.05 for long-term follow-up vs. at implantation or before discharge). The effect was most marked in a prespecified subgroup with high implant DFTs (> or =15 J). No patient developed an inadequate safety margin (< 9 J) during follow-up. CONCLUSIONS The DFTs declined significantly after implantation with an active pectoral, dual-coil transvenous lead system, and no clinically significant increases in DFT were observed. Therefore, routine defibrillation testing may not be required during the first two years after implantation with this lead system, in the absence of a change in the cardiac substrate or treatment with antiarrhythmic drugs.
Pacing and Clinical Electrophysiology | 1999
Neal G. Kavesh; Mary R. Olsovsky; Ronald S. Freudenberger; John V. Conte; Robert W. Peters; Stephen R. Shorofsky; Michael R. Gold
We report the case of an electrical storm in a cardiac arrest survivor with an ICD, in whom chronic oral amiodarone failed to suppress ventricular arrhythmias, and in whom intravenous amiodarone resulted in stability for 6 weeks prior to successful cardiac transplantation. Intravenous amiodarone can be successful in suppressing life‐threatening ventricular arrhythmias, even when chronic oral amiodarone is unsuccessful.