Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Peter K.N. Yu is active.

Publication


Featured researches published by Peter K.N. Yu.


Physics in Medicine and Biology | 2005

Post-irradiation colouration of Gafchromic EBT radiochromic film

Tsang Cheung; Martin J Butson; Peter K.N. Yu

Gafchromic EBT (International Specialty Products, NJ, USA), radiochromic film is one of the newest radiation-induced auto-developing x-ray analysis films available for therapeutic radiation dosimetry in radiotherapy applications. Part of any radiochromic film product which undergoes a polymerization reaction for automatic darkening is an associated post-irradiation colouration whereby the film continues to darken after irradiation has ceased. The Gafchromic EBT film has been shown to produce an approximate 6% to 9% increase in post-irradiation optical density within the first 12 h of irradiation within the 1 Gy to 5 Gy dose range. This is compared to approximately 13%, 15% and 19% for MD-55-2, XR type T and HS radiochromic film, respectively. It is also shown that the EBT films post-irradiation growth stabilizes to within 1% within the first 6 h. Thus EBT provides a reduced post-irradiation growth effect. However, to increase the accuracy of the film analysis, it is recommended that films be left for a significant period (at least 6 h) before the analysis is performed to provide a high level of accuracy. Also, calibration films must be read out with the same post-irradiation time to further enhance the accuracy of dosimetry.


Physics in Medicine and Biology | 2005

Absorption spectra variations of EBT radiochromic film from radiation exposure.

Martin J Butson; Tsang Cheung; Peter K.N. Yu

Gafchromic EBT radiochromic film is one of the newest radiation-induced auto-developing x-ray analysis films available for therapeutic radiation dosimetry in radiotherapy applications. The spectral absorption properties in the visible wavelengths have been investigated and results show two main peaks in absorption located at 636 nm and 585 nm. These absorption peaks are different to many other radiochromic film products such as Gafchromic MD-55 and HS film where two peaks were located at 676 nm and 617 nm respectively. The general shape of the absorption spectra is similar to older designs. A much higher sensitivity is found at high-energy x-rays with an average 0.6 OD per Gy variation in OD seen within the first Gy measured at 636 nm using 6 MV x-rays. This is compared to approximately 0.09 OD units for the first Gy at the 676 nm absorption peak for HS film at 6 MV x-ray energy. The films blue colour is visually different from older varieties of Gafchromic film with a higher intensity of mid-range blue within the film. The film provides adequate relative absorbed dose measurement for clinical radiotherapy x-ray assessment in the 1-2 Gy dose range which with further investigation may be useful for fractionated radiotherapy dose assessment.


Physics in Medicine and Biology | 2002

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Tsang Cheung; Martin J Butson; Peter K.N. Yu

Assessment of skin dose delivered to patients from radiotherapy x-ray beams should be performed both inside and outside the prescribed treatment fields. A multilayer Gafchromic film detector which has high sensitivity for detection of radiation can be used to measure skin dose in a two-dimensional map over the skin surface if required. This is an advantage over other detectors, which only provide point dose estimates. A study of 25 patients undergoing breast irradiation was performed to analyse the ability of the multilayer detector to analyse skin dose and to assess both in-field and out-of-field radiation doses delivered during tangent field breast irradiation. Results show that the main contributor to total skin dose within the treatment field was delivered by exit dose. However, outside the field, most dose was delivered by entry beams. Patients with smaller breast separations where found, in general, to receive a higher total skin dose from entry and exiting beams at the central axis. Results also showed that a significant skin dose was delivered outside the treatment field and the main cause of this dose was from electron contamination from entry beams. The multilayer Gafchromic film detector provided adequate skin dose assessment within one fraction of treatment for in vivo results.


ACS Nano | 2015

Self-Monitoring and Self-Delivery of Photosensitizer-Doped Nanoparticles for Highly Effective Combination Cancer Therapy in Vitro and in Vivo

Jinfeng Zhang; Yu-Chuan Liang; Xudong Lin; Xiaoyue Zhu; L. Yan; Shengliang Li; Xia Yang; Guangyu Zhu; Andrey L. Rogach; Peter K.N. Yu; Peng Shi; Lung-Chen Tu; Chia-Ching Chang; Xiaohong Zhang; Xianfeng Chen; Wenjun Zhang; Chun-Sing Lee

Theranostic nanomedicine is capable of diagnosis, therapy, and monitoring the delivery and distribution of drug molecules and has received growing interest. Herein, a self-monitored and self-delivered photosensitizer-doped FRET nanoparticle (NP) drug delivery system (DDS) is designed for this purpose. During preparation, a donor/acceptor pair of perylene and 5,10,15,20-tetro (4-pyridyl) porphyrin (H2TPyP) is co-doped into a chemotherapeutic anticancer drug curcumin (Cur) matrix. In the system, Cur works as a chemotherapeutic agent. In the meantime, the green fluorescence of Cur molecules is quenched (OFF) in the form of NPs and can be subsequently recovered (ON) upon release in tumor cells, which enables additional imaging and real-time self-monitoring capabilities. H2TPyP is employed as a photodynamic therapeutic drug, but it also emits efficient NIR fluorescence for diagnosis via FRET from perylene. By exploiting the emission characteristics of these two emitters, the combinatorial drugs provide a real-time dual-fluorescent imaging/tracking system in vitro and in vivo, and this has not been reported before in self-delivered DDS which simultaneously shows a high drug loading capacity (77.6%Cur). Overall, our carrier-free DDS is able to achieve chemotherapy (Cur), photodynamic therapy (H2TPyP), and real-time self-monitoring of the release and distribution of the nanomedicine (Cur and H2TPyP). More importantly, the as-prepared NPs show high cancer therapeutic efficiency both in vitro and in vivo. We expect that the present real-time self-monitored and self-delivered DDS with multiple-therapeutic and multiple-fluorescent ability will have broad applications in future cancer therapy.


Physics in Medicine and Biology | 1998

Effects of read-out light sources and ambient light on radiochromic film.

Martin J Butson; Peter K.N. Yu; Peter E Metcalfe

Both read-out light sources and ambient light sources can produce a marked effect on coloration of radiochromic film. Fluorescent, helium neon laser, light emitting diode (LED) and incandescent read-out light sources produce an equivalent dose coloration of 660 cGy h(-1), 4.3 cGy h(-1), 1.7 cGy h(-1) and 2.6 cGy h(-1) respectively. Direct sunlight, fluorescent light and incandescent ambient light produce an equivalent dose coloration of 30 cGy h(-1), 18 cGy h(-1) and 0 cGy h(-1) respectively. Continuously on, fluorescent light sources should not be used for film optical density evaluation and minimal exposure to any light source will increase the accuracy of results.


Medical Physics | 1999

Extrapolated surface dose measurements with radiochromic film

Martin J Butson; Peter K.N. Yu; Peter E Metcalfe

A radiochromic film extrapolation method is described for the measurement of surface dose from high energy photon beams. Extrapolated central axis entrance surface dose using Gafchromic film for a 10×10 cm 2 field size is 15%±2% and 13%±2% of D max for 6 and 10 MV x rays, respectively. Extrapolated surface dose for a 30×30 cm 2 field with a 10 mm perspex block tray is 49%±2% and 48%±2% of D max for 6 and 10 MV beams, respectively. All results agree with uncorrected Attix parallel plate ionization chambersurface ionization within 4% for the same beam energies and configurations.


Physics in Medicine and Biology | 2004

Effects of temperature variation on MOSFET dosimetry

Tsang Cheung; Martin J Butson; Peter K.N. Yu

This note investigates temperature effects on dosimetry using a metal oxide semiconductor field effect transistor (MOSFET) for radiotherapy x-ray treatment. This was performed by analysing the dose response and threshold voltage outputs for MOSFET dosimeters as a function of ambient temperature. Results have shown that the clinical semiconductor dosimetry system (CSDS) MOSFET provides stable dose measurements with temperatures varying from 15 degrees C up to 40 degrees C. Thus standard irradiations performed at room temperature can be directly compared to in vivo dose assessments performed at near body temperature without a temperature correction function. The MOSFET dosimeter threshold voltage varies with temperature and this level is dependent on the dose history of the MOSFET dosimeter. However, the variation can be accounted for in the measurement method. For accurate dosimetry, the detector should be placed for approximately 60 s on a patient to allow thermal equilibrium before measurements are taken with the final reading performed whilst still attached to the patient or conversely left for approximately 120 s after removal from the patient if initial readout was measured at room temperature to allow temperature equilibrium to be established.


Medical Physics | 2006

Measurement of high energy x-ray beam penumbra with Gafchromic EBT radiochromic film.

Tsang Cheung; Martin J Butson; Peter K.N. Yu

High energy x-ray beam penumbra are measured using Gafchromic™ EBT film. Gafchromic™ EBT, due to its limited energy dependence and high spatial resolution provide a high level of accuracy for dose assessment in penumbral regions. The spatial resolution of film detector systems is normally limited by the scanning resolution of the densitometer. Penumbral widths (80%/20%) measured at Dmax were found to be 2.8, 3.0, 3.2, and 3.4mm(±0.2mm) using 5, 10, 20, and 30cm square field sizes, respectively, for a 6MV linear accelerator produced x-ray beam. This is compared to 3.2mm±0.2mm (Kodak EDR2) and 3.6mm±0.2mm (Kodak X-Omat V) at 10cm×10cm measured using radiographic film. Using a zero volume extrapolation technique for ionization chamber measurements, the 10cm×10cm field penumbra at Dmax was measured to be 3.1mm, a close match to Gafchromic™ EBT results. Penumbral measurements can also be made at other depths, including the surface, as the film does not suffer significantly from dosimetric variations caused by changing x-ray energy spectra. Gafchromic™ EBT film provides an adequate measure of penumbral dose for high energy x-ray beams.


Physics in Medicine and Biology | 2002

High sensitivity radiochromic film dose comparisons

Martin J Butson; Peter K.N. Yu; Tsang Cheung; Peter E Metcalfe

This short note investigates the dose characteristics of a relatively new high sensitivity radiochromic film (Gafchromic HS) and compares dose and energy response to various Gafchromic film types and radiographic (EDR-2) film. The original MD-55-1 and two improved sensitivity films, MD-55-2 and HS film, were investigated for energy and dose response. Results show that the energy response of the new HS film is relatively the same as the original MD-55-1 and MD-55-2 films with a decrease in sensitivity at lower x-ray energies, with response decreasing down to approximately 0.64 (normalized to 1 for a 6 MV beam) for a 28 keV effective energy beam. This is compared to an over response of 9.2 at the same energy for EDR-2 film. The dose response at the maximum absorption peak was found to be approximately 3.8 and 1.9 times more sensitive than MD-55-1 and MD-55-2 films, respectively. At the absorption peak yielding the maximum optical density change, HS was found to be approximately 0.2 to 0.25 times the sensitivity of EDR-2.


Physics in Medicine and Biology | 2000

Verification of lung dose in an anthropomorphic phantom calculated by the collapsed cone convolution method

Martin J Butson; Rebecca Elferink; Tsang Cheung; Peter K.N. Yu; Michael Stokes; Kim You Quach; Peter E Metcalfe

Verification of calculated lung dose in an anthropomorphic phantom is performed using two dosimetry media. Dosimetry is complicated by factors such as variations in density at slice interfaces and appropriate position on CT scanning slice to accommodate these factors. Dose in lung for a 6 MV and 10 MV anterior-posterior field was calculated with a collapsed cone convolution method using an ADAC Pinnacle, 3D planning system. Up to 5% variations between doses calculated at the centre and near the edge of the 2 cm phantom slice positioned at the beam central axis were seen, due to the composition of each phantom slice. Validation of dose was performed with LiF thermoluminescent dosimeters (TLDs) and X-Omat V radiographic film. Both dosimetry media produced dose results which agreed closely with calculated results nearest their physical positioning in the phantom. The collapsed cone convolution method accurately calculates dose within inhomogeneous lung regions at 6 MV and 10 MV x-ray energy.

Collaboration


Dive into the Peter K.N. Yu's collaboration.

Top Co-Authors

Avatar

Martin J Butson

City University of Hong Kong

View shared research outputs
Top Co-Authors

Avatar

Tsang Cheung

City University of Hong Kong

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Hani Alnawaf

University of Wollongong

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Monica W.K. Kan

City University of Hong Kong

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Paul K.S. Lam

City University of Hong Kong

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Guangyu Zhu

City University of Hong Kong

View shared research outputs
Researchain Logo
Decentralizing Knowledge