Tsang Cheung
City University of Hong Kong
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by Tsang Cheung.
Physics in Medicine and Biology | 2005
Tsang Cheung; Martin J Butson; Peter K.N. Yu
Gafchromic EBT (International Specialty Products, NJ, USA), radiochromic film is one of the newest radiation-induced auto-developing x-ray analysis films available for therapeutic radiation dosimetry in radiotherapy applications. Part of any radiochromic film product which undergoes a polymerization reaction for automatic darkening is an associated post-irradiation colouration whereby the film continues to darken after irradiation has ceased. The Gafchromic EBT film has been shown to produce an approximate 6% to 9% increase in post-irradiation optical density within the first 12 h of irradiation within the 1 Gy to 5 Gy dose range. This is compared to approximately 13%, 15% and 19% for MD-55-2, XR type T and HS radiochromic film, respectively. It is also shown that the EBT films post-irradiation growth stabilizes to within 1% within the first 6 h. Thus EBT provides a reduced post-irradiation growth effect. However, to increase the accuracy of the film analysis, it is recommended that films be left for a significant period (at least 6 h) before the analysis is performed to provide a high level of accuracy. Also, calibration films must be read out with the same post-irradiation time to further enhance the accuracy of dosimetry.
Physics in Medicine and Biology | 2005
Martin J Butson; Tsang Cheung; Peter K.N. Yu
Gafchromic EBT radiochromic film is one of the newest radiation-induced auto-developing x-ray analysis films available for therapeutic radiation dosimetry in radiotherapy applications. The spectral absorption properties in the visible wavelengths have been investigated and results show two main peaks in absorption located at 636 nm and 585 nm. These absorption peaks are different to many other radiochromic film products such as Gafchromic MD-55 and HS film where two peaks were located at 676 nm and 617 nm respectively. The general shape of the absorption spectra is similar to older designs. A much higher sensitivity is found at high-energy x-rays with an average 0.6 OD per Gy variation in OD seen within the first Gy measured at 636 nm using 6 MV x-rays. This is compared to approximately 0.09 OD units for the first Gy at the 676 nm absorption peak for HS film at 6 MV x-ray energy. The films blue colour is visually different from older varieties of Gafchromic film with a higher intensity of mid-range blue within the film. The film provides adequate relative absorbed dose measurement for clinical radiotherapy x-ray assessment in the 1-2 Gy dose range which with further investigation may be useful for fractionated radiotherapy dose assessment.
Physics in Medicine and Biology | 2002
Tsang Cheung; Martin J Butson; Peter K.N. Yu
Assessment of skin dose delivered to patients from radiotherapy x-ray beams should be performed both inside and outside the prescribed treatment fields. A multilayer Gafchromic film detector which has high sensitivity for detection of radiation can be used to measure skin dose in a two-dimensional map over the skin surface if required. This is an advantage over other detectors, which only provide point dose estimates. A study of 25 patients undergoing breast irradiation was performed to analyse the ability of the multilayer detector to analyse skin dose and to assess both in-field and out-of-field radiation doses delivered during tangent field breast irradiation. Results show that the main contributor to total skin dose within the treatment field was delivered by exit dose. However, outside the field, most dose was delivered by entry beams. Patients with smaller breast separations where found, in general, to receive a higher total skin dose from entry and exiting beams at the central axis. Results also showed that a significant skin dose was delivered outside the treatment field and the main cause of this dose was from electron contamination from entry beams. The multilayer Gafchromic film detector provided adequate skin dose assessment within one fraction of treatment for in vivo results.
Australasian Physical & Engineering Sciences in Medicine | 2006
Martin J Butson; Tsang Cheung; Pkn Yu
Gafchromic EBT film, a new high sensitivity radiochromic film has been tested for variations in optical properties due to scanning orientation. Gafchromic EBT film has been shown to produce a scanning orientation effect whereby variations in measured relative optical density are found due to the films orientation relative to the scanner direction. This relative optical density change was found to be relatively consistent for different films exposed to varying dose levels ranging from 0 Gy to 3 Gy. A maximum variation of 0.0157±0.0035 in optical density (OD) was found. This relates to an approximate 15 % variation in net OD for a 50 cGy irradiated film and 4 % variation for a 3 Gy irradiated film. No noticeable effects or variations were seen with changing scanning resolution or with the film placed “up or down” during scanning. Other Gafchromic film types were tested and compared to EBT for unirradiated film to assess the magnitude of this orientation effect on the scanner used and results showed that EBT produced a significantly higher effect that MD-55-2, HS, XR type T and XR type R film by up to 3 times. As such, providing the same orientation of EBT film when scanning for dosimetric analysis becomes an essential part of EBT film dosimetry.
Physics in Medicine and Biology | 2004
Tsang Cheung; Martin J Butson; Peter K.N. Yu
This note investigates temperature effects on dosimetry using a metal oxide semiconductor field effect transistor (MOSFET) for radiotherapy x-ray treatment. This was performed by analysing the dose response and threshold voltage outputs for MOSFET dosimeters as a function of ambient temperature. Results have shown that the clinical semiconductor dosimetry system (CSDS) MOSFET provides stable dose measurements with temperatures varying from 15 degrees C up to 40 degrees C. Thus standard irradiations performed at room temperature can be directly compared to in vivo dose assessments performed at near body temperature without a temperature correction function. The MOSFET dosimeter threshold voltage varies with temperature and this level is dependent on the dose history of the MOSFET dosimeter. However, the variation can be accounted for in the measurement method. For accurate dosimetry, the detector should be placed for approximately 60 s on a patient to allow thermal equilibrium before measurements are taken with the final reading performed whilst still attached to the patient or conversely left for approximately 120 s after removal from the patient if initial readout was measured at room temperature to allow temperature equilibrium to be established.
Medical Physics | 2006
Tsang Cheung; Martin J Butson; Peter K.N. Yu
High energy x-ray beam penumbra are measured using Gafchromic™ EBT film. Gafchromic™ EBT, due to its limited energy dependence and high spatial resolution provide a high level of accuracy for dose assessment in penumbral regions. The spatial resolution of film detector systems is normally limited by the scanning resolution of the densitometer. Penumbral widths (80%/20%) measured at Dmax were found to be 2.8, 3.0, 3.2, and 3.4mm(±0.2mm) using 5, 10, 20, and 30cm square field sizes, respectively, for a 6MV linear accelerator produced x-ray beam. This is compared to 3.2mm±0.2mm (Kodak EDR2) and 3.6mm±0.2mm (Kodak X-Omat V) at 10cm×10cm measured using radiographic film. Using a zero volume extrapolation technique for ionization chamber measurements, the 10cm×10cm field penumbra at Dmax was measured to be 3.1mm, a close match to Gafchromic™ EBT results. Penumbral measurements can also be made at other depths, including the surface, as the film does not suffer significantly from dosimetric variations caused by changing x-ray energy spectra. Gafchromic™ EBT film provides an adequate measure of penumbral dose for high energy x-ray beams.
Physics in Medicine and Biology | 2002
Martin J Butson; Peter K.N. Yu; Tsang Cheung; Peter E Metcalfe
This short note investigates the dose characteristics of a relatively new high sensitivity radiochromic film (Gafchromic HS) and compares dose and energy response to various Gafchromic film types and radiographic (EDR-2) film. The original MD-55-1 and two improved sensitivity films, MD-55-2 and HS film, were investigated for energy and dose response. Results show that the energy response of the new HS film is relatively the same as the original MD-55-1 and MD-55-2 films with a decrease in sensitivity at lower x-ray energies, with response decreasing down to approximately 0.64 (normalized to 1 for a 6 MV beam) for a 28 keV effective energy beam. This is compared to an over response of 9.2 at the same energy for EDR-2 film. The dose response at the maximum absorption peak was found to be approximately 3.8 and 1.9 times more sensitive than MD-55-1 and MD-55-2 films, respectively. At the absorption peak yielding the maximum optical density change, HS was found to be approximately 0.2 to 0.25 times the sensitivity of EDR-2.
Physics in Medicine and Biology | 2000
Martin J Butson; Rebecca Elferink; Tsang Cheung; Peter K.N. Yu; Michael Stokes; Kim You Quach; Peter E Metcalfe
Verification of calculated lung dose in an anthropomorphic phantom is performed using two dosimetry media. Dosimetry is complicated by factors such as variations in density at slice interfaces and appropriate position on CT scanning slice to accommodate these factors. Dose in lung for a 6 MV and 10 MV anterior-posterior field was calculated with a collapsed cone convolution method using an ADAC Pinnacle, 3D planning system. Up to 5% variations between doses calculated at the centre and near the edge of the 2 cm phantom slice positioned at the beam central axis were seen, due to the composition of each phantom slice. Validation of dose was performed with LiF thermoluminescent dosimeters (TLDs) and X-Omat V radiographic film. Both dosimetry media produced dose results which agreed closely with calculated results nearest their physical positioning in the phantom. The collapsed cone convolution method accurately calculates dose within inhomogeneous lung regions at 6 MV and 10 MV x-ray energy.
Australasian Physical & Engineering Sciences in Medicine | 2009
Martin J Butson; Tsang Cheung; Peter K.N. Yu; Hani Alnawaf
With new advancements in radiochromic film designs and sensitivity to suit different niche applications, EBT2 is the latest offering for the megavoltage radiotherapy market. New construction specifications including different physical construction and the use of a yellow coloured dye has provided the next generation radiochromic film for therapy applications. The film utilises the same active chemical for radiation measurement as its predecessor, EBT Gafchromic. Measurements have been performed using photo spectrometers to analyse the absorption spectra properties of this new EBT2 Gafchromic, radiochromic film. Results have shown that whilst the physical coloration or absorption spectra of the film, which turns yellow to green as compared to EBT film, (clear to blue) is significantly different due to the added yellow dye, the net change in absorption spectra properties for EBT2 are similar to the original EBT film. Absorption peaks are still located at 636nm and 585nm positions. A net optical density change of 0.590 ± 0.020 (2SD) for a 1 Gy radiation absorbed dose using 6 MV x-rays when measured at the 636nm absorption peak was found. This is compared to 0.602 ± 0.025 (2SD) for the original EBT film (2005 Batch) and 0.557 ± 0.027 (2009 Batch) at the same absorption peak. The yellow dye and the new coating material produce a significantly different visible absorption spectra results for the EBT2 film compared to EBT at wavelengths especially below approximately 550nm. At wavelengths above 550nm differences in absolute OD are seen however, when dose analysis is performed at wavelengths above 550nm using net optical density changes, no significant variations are seen. If comparing results of the late production EBT to new production EBT2 film, net optical density variations of approximately 10 % to 15 % are seen. As all new film batches should be calibrated for sensitivity upon arrival this should not be of concern.
Physics in Medicine and Biology | 2001
Martin J Butson; Tsang Cheung; Peter K.N. Yu
Radiochromic film is investigated for use in dosimetry in water phantoms as opposed to solid phantoms. Investigations are performed to measure the penetration rates of water into radiochromic film and to assess the effects on optical density that this penetration causes. The effects of film orientation during irradiation in water are also tested. Results show that only a small penetration rate is seen from water into the film which only affects the outer areas of the film, with penetration being less than 0.5 mm per hour. The optical density measurements of the film at 660 nm remain unchanged in the unaffected regions of the radiochromic film. Minimal effects are seen due to beam orientation in a water phantom as opposed to solid water phantoms in which an overestimation in dose is normally seen for parallel irradiation. Radiochromic film seems to be an adequate detector for dosimetry in a water phantom where high spatial resolution is needed and angle of beam incidence at the point of interest is important.