R Sadagopan
University of Texas MD Anderson Cancer Center
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by R Sadagopan.
Medical Physics | 2010
Stanley H. Benedict; Kamil M. Yenice; D Followill; James M. Galvin; William H. Hinson; Brian D. Kavanagh; P Keall; Michael Lovelock; Sanford L. Meeks; Lech Papiez; Thomas G. Purdie; R Sadagopan; Michael C. Schell; Bill J. Salter; David Schlesinger; Almon S. Shiu; Timothy D. Solberg; Danny Y. Song; Volker W. Stieber; Robert D. Timmerman; Wolfgang A. Tomé; Dirk Verellen; Lu Wang; Fang-Fang Yin
Task Group 101 of the AAPM has prepared this report for medical physicists, clinicians, and therapists in order to outline the best practice guidelines for the external-beam radiation therapy technique referred to as stereotactic body radiation therapy (SBRT). The task group report includes a review of the literature to identify reported clinical findings and expected outcomes for this treatment modality. Information is provided for establishing a SBRT program, including protocols, equipment, resources, and QA procedures. Additionally, suggestions for developing consistent documentation for prescribing, reporting, and recording SBRT treatment delivery is provided.
Journal of Applied Clinical Medical Physics | 2009
R Sadagopan; J BenComo; Rafael Martin; Görgen Nilsson; Thomas Matzen; P Balter
Intensity‐modulated radiation therapy (IMRT) is a complex procedure that involves the delivery of complex intensity patterns from various gantry angles. Due to the complexity of the treatment plans, the standard care is to perform measurement‐based, patient‐specific quality assurance (QA). IMRT QA is traditionally done with film for relative dose in a plane and with an ion chamber for absolute dose. This is a laborious and time‐consuming process. In this work, we characterized, commissioned, and evaluated the QA capabilities of a novel commercial IMRT device, Delta4, (ScandiDos, Uppsala, Sweden). This device consists of diode matrices in two orthogonal planes inserted in a cylindrical acrylic phantom that is 22 cm in diameter. Although the system has detectors in only two planes, it provides a novel interpolation algorithm that is capable of estimating doses at points where no detectors are present. Each diode is sampled per beam pulse so that the dose distribution can be evaluated on segment‐by‐segment, beam‐by‐beam, or as a composite plan from a single set of measurements. The end user can calibrate the system to perform absolute dosimetry, eliminating the need for additional ion chamber measurements. The patients IMRT plan is imported into the device over the hospital LAN and the results of the measurements can be displayed as gamma profiles, distance‐to‐agreement maps, dose difference maps, or the measured dose distribution can be superimposed on the patients anatomy to display an as‐delivered plan. We evaluated the systems reproducibility, stability, pulse‐rate dependence, dose‐rate dependence, angular dependence, linearity of dose response, and energy response using carefully planned measurements. We also validated the systems interpolation algorithm by measuring a complex dose distribution from an IMRT treatment. Several simple and complex isodose distributions planned using a treatment planning system were delivered to the QA device; the planned and measured dose distributions were then compared and analyzed. In addition, the dose distributions measured by conventional IMRT QA, which uses an ion chamber and film, were compared. We found that the Delta4 device is accurate and reproducible and that its interpolation algorithm is valid. In addition, the supplied software and network interface allow a streamlined IMRT QA process. PACS number: 87.56Fc
International Journal of Radiation Oncology Biology Physics | 2011
Yong Peng; S. Vedam; Joe Y. Chang; Song Gao; R Sadagopan; M Bues; P Balter
PURPOSE To analyze tumor position reproducibility of feedback-guided voluntary deep inspiration breath-hold (FGBH) gating for cone beam computed tomography (CBCT)-based stereotactic body radiotherapy (SBRT). METHODS AND MATERIALS Thirteen early-stage lung cancer patients eligible for SBRT with tumor motion of >1cm were evaluated for FGBH-gated treatment. Multiple FGBH CTs were acquired at simulation, and single FGBH CBCTs were also acquired prior to each treatment. Simulation CTs and treatment CBCTs were analyzed to quantify reproducibility of tumor positions during FGBH. Benefits of FGBH gating compared to treatment during free breathing, as well treatment with gating at exhalation, were examined for lung sparing, motion margins, and reproducibility of gross tumor volume (GTV) position relative to nonmoving anatomy. RESULTS FGBH increased total lung volumes by 1.5 times compared to free breathing, resulting in a proportional drop in total lung volume receiving 10 Gy or more. Intra- and inter-FGBH reproducibility of GTV centroid positions at simulation were 1.0 ± 0.5 mm, 1.3 ± 1.0 mm, and 0.6 ± 0.4 mm in the anterior-posterior (AP), superior-inferior (SI), and left-right lateral (LR) directions, respectively, compared to more than 1 cm of tumor motion at free breathing. During treatment, inter-FGBH reproducibility of the GTV centroid with respect to bony anatomy was 1.2 ± 0.7 mm, 1.5 ± 0.8 mm, and 1.0 ± 0.4 mm in the AP, SI, and LR directions. In addition, the quality of CBCTs was improved due to elimination of motion artifacts, making this technique attractive for poorly visualized tumors, even with small motion. CONCLUSIONS The extent of tumor motion at normal respiration does not influence the reproducibility of the tumor position under breath hold conditions. FGBH-gated SBRT with CBCT can improve the reproducibility of GTV centroids, reduce required margins, and minimize dose to normal tissues in the treatment of mobile tumors.
British Journal of Radiology | 2017
Julianne M. Pollard; Z Wen; R Sadagopan; Jihong Wang; Geoffrey S. Ibbott
Advances in image-guided radiotherapy (RT) have allowed for dose escalation and more precise radiation treatment delivery. Each decade brings new imaging technologies to help improve RT patient setup. Currently, the most frequently used method of three-dimensional pre-treatment image verification is performed with cone beam CT. However, more recent developments have provided RT with the ability to have on-board MRI coupled to the teleradiotherapy unit. This latest tool for treating cancer is known as MR-guided RT. Several varieties of these units have been designed and installed in centres across the globe. Their prevalence, history, advantages and disadvantages are discussed in this review article. In preparation for the next generation of image-guided RT, this review also covers where MR-guided RT might be heading in the near future.
Medical Physics | 2007
R Sadagopan; J BenComo; Rafael Martin; P Balter; S. Vedam; G. Nilsson
Purpose: Characterize, commission and evaluate a dual plane diode matrix IMRT QA device. Method and Materials: A novel device consisting of diode matrices in two orthogonal planes inserted in a cylindrical acrylic phantom of 22cm diameter is characterized, commissioned and evaluated for radiotherapyquality assurance. The system interfaces readily with a networked computer making the whole IMRT QA process very efficient in multi accelerator and multi physicist department. . It detects charge per accelerator pulse, computes and displays measured dose distribution in 3D space. The temperature dependence of the diode is corrected. The precision, stability, pulse rate dependence, dose rate dependence, angular dependence, linear response, energy response of the system and the calculation accuracy at non detector locations are evaluated in addition to comparing multiple simple and complex iso‐dose distributions from TPS to measured distributions. The software readily analyses dose profiles in any orientations, %dose, DTA and gamma index of the entire 3D distributions. Results: The precision and the day‐to‐day reproducibility of measured data of a single field are excellent, making additional ion chamber measurement unnecessary. The measured data indicated excellent dose linearity and pulse rate independence. Comparison of simple and complex treatment plans with delivered treatment showed good agreement considering the error bars. Conclusions: The Delta 4 system is highly efficient, accurate and reproducible. The instantaneous and automatic data acquisition combined with the error analysis, report and database capability built into the system make it easy, convenient and efficient to use in a busy clinic. Conflict of Interest Statement: One of the co‐author is President and CEO of ScandiDos AB Company, which supplied Delta4 at no cost for evaluation.
Medical Physics | 2014
R Sadagopan; J Yang; Heng Li
PURPOSE To develop a tool enabling soft tissue based image guidance using dual energy radiographs for cases when on-board CT is not available. METHOD Dual energy planar radiographs can be applied to image guidance for targeting lung lesions because the bone based alignment only may not be sufficient as the lesions move. We acquired images of an anthropomorphic thorax phantom at 120 and 60 KVp respectively. Using a weighted logarithmic subtraction of these dual energy images, a soft tissue enhanced and a bone enhanced image were generated and they could be used for the image guidance purpose. Similar processing was also applied to a dual energy image set acquired for a patient undergoing a proton therapy. RESULTS The soft tissue enhanced images suppressed bones (ribs and scapula) overlying on lung, thus enabling a better visualization of soft tissue and lesion, while the bone enhanced image suppressed the soft tissue. These enhanced effects were visually apparent without further processing for display enhancements, such as using histogram or edge enhancement technique. CONCLUSIONS The phantom image processing was encouraging. The initial test on the patient image set showed that other post processing might still be able to add value in visualizing soft tissues in addition to the dual energy soft tissue enhancement. More evaluations are needed to determine the potential benefit of this technique in the clinic.
Medical Physics | 2006
Narayan Sahoo; A Kazi; S Cho; R Sadagopan; C. Wang; Almon S. Shiu
Purpose: Very small fields and segments of area less than one square centimeter are routinely being used for IMRT and stereotactic radiosurgery. Accurate measurement of beam profile is essential for treatment planning.Ion chambers with very small cavity radius, specialized diodes and films are commonly used for these beam data measurement. The purpose of this investigation is both to study the effect of the detector properties on the measured beam profiles of the small field MV x‐ray beams and to apply the necessary correction to determine the real profiles. Method and Materials: Two ionization chambers with cavity radius of 2 mm and 1 mm, a stereotactic diode and XV film were used to measure the beam profiles of circular fields of stereotactic cones and small square fields defined by collimator jaws and MLC. The penumbra widths of the profiles were compared to study the effect of the physical properties of the detectors, such as, size, energy dependence and dose rate dependence on the measured beam profiles. The profiles measured by the larger ionization chamber were corrected for the detector size effect by using a semi‐empirical procedure [1] and was used as the reference profile to derive the detector response function of other detectors with smaller size and better spatial resolution. The detector response functions were then used to correct the measured profiles of small fields. Results: The differences in the profiles measured by different detectors were significantly reduced after the profiles were corrected with detector response functions. Conclusion: The accuracy of the profile measurement of small therapy beams can be significantly improved when appropriate corrections are applied to take into account the variation of detector response in different regions of the beam.
Medical Physics | 2016
J Pollard; S Prajapati; S Gao; P Nitsch; R Sadagopan; X Wang; P Balter
PURPOSE To evaluate correlation between the reproducibility of tumor position under feedback guided voluntary deep inspiration breath hold gating at simulation and at treatment. METHODS All patients treated with breath hold (BH) have 3-6 BH CTs taken at simulation (sim). In addition, if the relationship between the tumor and nearby bony anatomy on treatment BH CT(or CBCT) is found to be greater than 5 mm different at treatment than it was at sim, a repeat BH CT is taken before treatment. We retrospectively analyzed the sim CTs for 19 patients who received BH SBRT lung treatments and had repeat BH CT on treatment. We evaluated the reproducibility of the tumor position during the simulation CTs and compared this to the reproducibility of the tumor position on the repeat treatment CT with our in-house CT alignment software (CT-Assisted Targeting for Radiotherapy). RESULTS Comparing the tumor position for multiple simulation BH CTs, we calculated: maximum difference (max) = 0.69cm; average difference (x) = 0.28cm; standard deviation (σ) = 0.18cm. Comparing the repeat BH CBCTs on treatment days we calculated: max = 0.44cm; x = 0.16cm; σ = 0.22cm. We also found that for 95% of our BH cases, the absolute variation in tumor position within the same imaging day was within 5mm of the range at the time of simulation and treatment. We found that 75% of the BH cases had less residual tumor motion on treatment days than at simulation. CONCLUSION This suggests that a GTV contour based upon the residual tumor motion in multiple BH datasets plus 2 mm margin should be sufficient to cover the full range of residual tumor motion on treatment days.
Medical Physics | 2015
D Lewis; X Wang; S Gao; P Nitsch; J Pollard; R Sadagopan; P Balter
Purpose: To assess the feasibility of treating lung SBRT patients with the ipsilateral arm adducted beside the body instead of elevated above the head. Methods: Patients receiving lung SBRT at our institution are typically treated with both arms raised above their head. However, several patients had difficulty maintaining their arms in an elevated position. In this study, lung SBRT patients who underwent PET-CT imaging with an adducted ipsilateral arm were selected to investigate the dosimetric effects of this treatment setup. PET-CT datasets were fused with treatment planning CT images to simulate the adducted arm position. One VMAT treatment plan was created per patient using the Pinnacle treatment planning system. Plans were optimized to achieve minimal dose to the ipsilateral arm while keeping the target coverage and critical structure doses within clinical limits. The target dose coverage, conformity index (CI) for the target, and DVHs of critical structures for the adducted arm plan were calculated. These parameters were compared with the clinical plan and reported along with the mean and maximum doses of the ipsilateral arm. Results: The target coverage, CI and DVHs for the adducted arm plans of two patients (one with peripheral lesion and one with central lesion) were comparable with the clinical plans. Dose constraints for the chest wall limited further reduction of ipsilateral arm doses for the peripheral lesion plan. The mean ipsilateral arm doses for the central and peripheral lesions were 0.33 Gy and 2.4 Gy, respectively. The maximum ipsilateral arm doses for the central and peripheral lesions were 1.0 Gy and 6.2 Gy, respectively. Conclusion: The results suggested patients with central lung SBRT tumors were more suitable for treatment with the adducted arm position. More patients with various lung tumor locations will be studied to find optimal tumor locations for treatment with this arm position.
Medical Physics | 2015
S Gao; P Balter; X Wang; R Sadagopan; J Pollard
Purpose: To determine the best dosimetric metric measured by our routine QA devices for diagnosing photon target failure on a Varian C-series linac. Methods: We have retrospectively reviewed and analyzed the dosimetry data from a Varian linac with a target degradation that was undiagnosed for one year. A failure in the daily QA symmetry tests was the first indication of an issue. The beam was steered back to a symmetric shape and water scans indicated the beam energy had changed but stayed within the manufacturers specifications and agreed reasonably with our treatment planning system data. After the problem was identified and the target was replaced, we retrospectively analyzed our QA data including diagonals normalized flatness (F_DN) from the daily device (DQA3), profiles from an ionization chamber array (IC Profiler), as well as profiles and PDDs from a 3D water Scanner (3DS). These metrics were cross-compared to determine which was the best early indicator of target degradation. Results: A 3% change in FDN measured by the DQA3 was found to be an early indicator of target degradation. It is more sensitive than changes in output, symmetry, flatness or PDD. All beam shape metrics (flatness at dmax and 10 cm depth, and F_DN) indicated an energy increase while the PDD indicated an energy decrease. This disagreement between the beam-shape based energy metrics (F_DN and flatness) and PDD based energy metric may indicate target failure as opposed to an energy change resulting from changes in the incident electron energy. Conclusion: Photon target degradation has been identified as a failure mode for linacs. The manufacturers test for this condition is highly invasive and requires machine down time. We have demonstrated that the condition could be caught early based upon data acquired during routine QA activities, such as the F_DN.