Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Thomas Allmendinger is active.

Publication


Featured researches published by Thomas Allmendinger.


Medical Physics | 2009

Dual-source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: image reconstruction and assessment of image quality.

Thomas Flohr; Shuai Leng; Lifeng Yu; Thomas Allmendinger; Herbert Bruder; Martin Petersilka; Christian Eusemann; Karl Stierstorfer; Bernhard Schmidt; Cynthia H. McCollough

PURPOSE To present the theory for image reconstruction of a high-pitch, high-temporal-resolution spiral scan mode for dual-source CT (DSCT) and evaluate its image quality and dose. METHODS With the use of two x-ray sources and two data acquisition systems, spiral CT exams having a nominal temporal resolution per image of up to one-quarter of the gantry rotation time can be acquired using pitch values up to 3.2. The scan field of view (SFOV) for this mode, however, is limited to the SFOV of the second detector as a maximum, depending on the pitch. Spatial and low contrast resolution, image uniformity and noise, CT number accuracy and linearity, and radiation dose were assessed using the ACR CT accreditation phantom, a 30 cm diameter cylindrical water phantom or a 32 cm diameter cylindrical PMMA CTDI phantom. Slice sensitivity profiles (SSPs) were measured for different nominal slice thicknesses, and an anthropomorphic phantom was used to assess image artifacts. Results were compared between single-source scans atpitch=1.0 and dual-source scans at pitch=3.2. In addition, image quality and temporal resolution of an ECG-triggered version of the DSCT high-pitch spiral scan mode were evaluated with a moving coronary artery phantom, and radiation dose was assessed in comparison with other existing cardiac scan techniques. RESULTS No significant differences in quantitative measures of image quality were found between single-source scans atpitch=1.0 and dual-source scans at pitch=3.2 for spatial and low contrast resolution, CT number accuracy and linearity, SSPs, image uniformity, and noise. The pitch value (1.6≤pitch≤3.2) had only a minor impact on radiation dose and image noise when the effective tube current time product (mA s/pitch) was kept constant. However, while not severe, artifacts were found to be more prevalent for the dual-source pitch=3.2 scan mode when structures varied markedly along the z axis, particularly for head scans. Images of the moving coronary artery phantom acquired with the ECG-triggered high-pitch scan mode were visually free from motion artifacts at heart rates of 60 and 70 bpm. However, image quality started to deteriorate for higher heart rates. At equivalent image quality, the ECG-triggered high-pitch scan mode demonstrated lower radiation dose than other cardiac scan techniques on the same DSCT equipment (25% and 60% dose reduction compared to ECG-triggered sequential step-and-shoot and ECG-gated spiral with x-ray pulsing). CONCLUSIONS A high-pitch (up topitch=3.2), high-temporal-resolution (up to 75 ms) dual-source CT scan mode produced equivalent image quality relative to single-source scans using a more typical pitch value (pitch=1.0). The resultant reduction in the overall acquisition time may offer clinical advantage for cardiovascular, trauma, and pediatric CT applications. In addition, ECG-triggered high-pitch scanning may be useful as an alternative to ECG-triggered sequential scanning for patients with low to moderate heart rates up to 70 bpm, with the potential to scan the heart within one heart beat at reduced radiation dose.


Investigative Radiology | 2011

Dose reduction in abdominal computed tomography: intraindividual comparison of image quality of full-dose standard and half-dose iterative reconstructions with dual-source computed tomography.

M May; Wolfgang Wüst; Michael Brand; Christian Stahl; Thomas Allmendinger; Bernhard Schmidt; Michael Uder; Michael Lell

Objectives:We sought to evaluate the image quality of iterative reconstruction in image space (IRIS) in half-dose (HD) datasets compared with full-dose (FD) and HD filtered back projection (FBP) reconstruction in abdominal computed tomography (CT). Materials and Methods:To acquire data with FD and HD simultaneously, contrast-enhanced abdominal CT was performed with a dual-source CT system, both tubes operating at 120 kV, 100 ref.mAs, and pitch 0.8. Three different image datasets were reconstructed from the raw data: Standard FD images applying FBP which served as reference, HD images applying FBP and HD images applying IRIS. For the HD data sets, only data from 1 tube detector-system was used. Quantitative image quality analysis was performed by measuring image noise in tissue and air. Qualitative image quality was evaluated according to the European Guidelines on Quality criteria for CT. Additional assessment of artifacts, lesion conspicuity, and edge sharpness was performed. Results:Image noise in soft tissue was substantially decreased in HD-IRIS (−3.4 HU, −22%) and increased in HD-FBP (+6.2 HU, +39%) images when compared with the reference (mean noise, 15.9 HU). No significant differences between the FD-FBP and HD-IRIS images were found for the visually sharp anatomic reproduction, overall diagnostic acceptability (P = 0.923), lesion conspicuity (P = 0.592), and edge sharpness (P = 0.589), while HD-FBP was rated inferior. Streak artifacts and beam hardening was significantly more prominent in HD-FBP while HD-IRIS images exhibited a slightly different noise pattern. Conclusions:Direct intrapatient comparison of standard FD body protocols and HD-IRIS reconstruction suggest that the latest iterative reconstruction algorithms allow for approximately 50% dose reduction without deterioration of the high image quality necessary for confident diagnosis.


Investigative Radiology | 2014

Ultralow-dose chest computed tomography for pulmonary nodule detection: first performance evaluation of single energy scanning with spectral shaping.

Sonja Gordic; Fabian Morsbach; Bernhard Schmidt; Thomas Allmendinger; Thomas Flohr; Daniela B. Husarik; Stephan Baumueller; Rainer Raupach; Paul Stolzmann; Sebastian Leschka; Thomas Frauenfelder; Hatem Alkadhi

PurposeThe purpose of this study was to evaluate the image quality and sensitivity of ultralow radiation dose single-energy computed tomography (CT) with tin filtration for spectral shaping and iterative reconstructions for the detection of pulmonary nodules in a phantom setting. MethodsSingle-energy CT was performed using third-generation dual-source CT (SOMATOM Force; 2 × 192 slices) at 70 kVp, 100 kVp with tin filtration (100Sn kVp), and 150Sn kV with tube current-time product adjustments resulting in standard dose (CT volume dose index, 3.1 mGy/effective dose, 1.3 mSv at a scan length of 30 cm), 1/10th dose level (0.3 mGy/0.13 mSv), and 1/20th dose level (0.15 mGy/0.06 mSv). An anthropomorphic chest phantom simulating an intermediate-sized adult with randomly distributed solid pulmonary nodules of various sizes (2–10 mm; attenuation, 75 HU at 120 kVp) was used. Images were reconstructed with advanced model-based iterative reconstruction (ADMIRE; strength levels 3 and 5) and were compared with those acquired with second-generation dual-source CT at 120 kVp (reconstructed with filtered back projection) and sinogram-affirmed iterative reconstruction (strength level 3) at the lowest possible dose at 120 kVp (CT volume dose index, 0.28 mGy). One blinded reader measured image noise, and 2 blinded, independent readers determined overall image quality on a 5-grade scale (1 = nondiagnostic to 5 = excellent) and marked nodule localization with confidence rates on a 5-grade scale (1 = unsure to 5 = high confidence). The constructional drawing of the phantom served as reference standard for calculation of sensitivity. Two patients were included, for proof of concept, who were scanned with the 100Sn kVp protocol at the 1/10th and 1/20th dose level. ResultsImage noise was highest in the images acquired with second-generation dual-source CT and reconstructed with filtered back projection. At both the 1/10th and 1/20th dose levels, image noise at a tube voltage of 100Sn kVp was significantly lower than in the 70 kVp and 150Sn kV data sets (ADMIRE 3, P < 0.01; ADMIRE 5, P < 0.05). Sensitivity of nodule detection was lowest in images acquired with second-generation dual-source CT at 120 kVp and the lowest possible dose. Protocols at 100Sn kVp and ADMIRE 5 showed highest sensitivity at the 1/10th and 1/20th dose levels. Highest numbers of false-positives occurred in second-generation dual-source CT images (range, 12–15), whereas lowest numbers occurred in the 1/10th and 1/20th dose data sets acquired with third-generation dual-source CT at 100Sn kVp and reconstructed with ADMIRE strength levels 3 and 5 (total of 1 and 0 false-positives, respectively). Diagnostic confidence at 100Sn kVp was significantly higher than at 70 kVp or 150Sn kV (ADMIRE 3, P < 0.05; ADMIRE 5, P < 0.01) at both the 1/10th and 1/20th dose levels. Images of the 2 patients scanned with 100Sn kVp at the 1/10th and 1/20th dose levels were of diagnostic quality. ConclusionsOur study suggests that chest CT for the detection of pulmonary nodules can be performed with third-generation dual-source CT producing high image quality, sensitivity, and diagnostic confidence at a very low effective radiation dose of 0.06 mSv when using a single-energy protocol at 100 kVp with spectral shaping and when using advanced iterative reconstruction techniques.


Radiology | 2014

Closing in on the K Edge: Coronary CT Angiography at 100, 80, and 70 kV—Initial Comparison of a Second- versus a Third-Generation Dual-Source CT System

Mathias Meyer; Holger Haubenreisser; U. Joseph Schoepf; Rozemarijn Vliegenthart; Christianne Leidecker; Thomas Allmendinger; Ralf Lehmann; Sonja Sudarski; Martin Borggrefe; Stefan O. Schoenberg; Thomas Henzler

PURPOSE To prospectively evaluate radiation and contrast medium requirements for performing high-pitch coronary computed tomographic (CT) angiography at 70 kV using a third-generation dual-source CT system in comparison to a second-generation dual-source CT system. MATERIALS AND METHODS All patients gave informed consent for this institutional review board-approved study. Forty-five patients (median age, 52 years; 27 men) were imaged in high-pitch mode with a third-generation dual-source CT system at 70 kV (n = 15) or with a second-generation dual-source CT system at 80 or 100 kV (n = 15 for each). Tube voltage was based on body mass index: 80 or 70 kV for less than 26 kg/m(2) versus 100 kV for 26-30 kg/m(2). For the 80- and 100-kV protocols, 80 mL of contrast material was injected, versus 45 mL for the 70-kV protocol. Data were reconstructed by using a second-generation iterative reconstruction algorithm for second-generation dual-source CT and a recently introduced third-generation iterative reconstruction algorithm for third-generation dual-source CT. Objective image quality was measured for various regions of interest, and subjective image quality was evaluated with a five-point Likert scale. RESULTS The signal-to-noise ratio of the coronary CT angiography studies acquired with 70 kV was significantly higher (70 kV: 14.3-17.6 vs 80 kV: 7.1-12.9 vs 100 kV: 9.8-12.9; P < .0497) than those acquired with the other two protocols for all coronary arteries. Qualitative image quality analyses revealed no significant differences between the three CT angiography protocols (median score, 5; P > .05). The mean effective dose was 75% and 108% higher (0.92 mSv ± 0.3 [standard deviation] and 0.78 mSv ± 0.2 vs 0.44 mSv ± 0.1; P < .0001), respectively, for the 80- and 100-kV CT angiography protocols than for the 70-kV CT angiography protocol. CONCLUSION In nonobese patients, third-generation high-pitch coronary dual-source CT angiography at 70 kV results in robust image quality for studying the coronary arteries, at significantly reduced radiation dose (0.44 mSv) and contrast medium volume (45 mL), thus enabling substantial radiation dose and contrast medium savings as compared with second-generation dual-source CT.


Investigative Radiology | 2010

Quantitative Whole Heart Stress Perfusion CT Imaging as Noninvasive Assessment of Hemodynamics in Coronary Artery Stenosis: Preliminary Animal Experience

Andreas H. Mahnken; Ernst Klotz; Hubertus Pietsch; Bernhard Schmidt; Thomas Allmendinger; Ulrike Haberland; Willi A. Kalender; Thomas Flohr

Purpose:To quantify differences in regional myocardial perfusion in coronary artery stenosis by the use of dual source computed tomography (DSCT) in an animal model. Material and Methods:In 5 pigs, an 80% stenosis of the left anterior descending artery was successfully induced by partial balloon occlusion (ischemia group). Five animals served as control group. All animals underwent contrast enhanced whole heart DSCT (Definition Flash, Siemens, Germany) perfusion imaging using a prototype electrocardiogram -triggered dynamic scan mode. Imaging was performed at rest as well as under stress conditions during continuous infusion of adenosine (240 mg/kg/min). For contrast enhancement 60 mL Iopromide 300 (Ultravist 300, Bayer-Schering Pharma, Berlin, Germany) were injected at a rate of 6 mL/s. Myocardial blood flow (MBF), first pass distribution volume, and intravascular blood volume were volumetrically quantified. Results:In the control group MBF increased significantly from 98.2 mL/100 mL/min to 134.0 mL/100 mL/min if adenosine was administered (P = 0.0153). There were no significant differences in the perfusion parameters comparing the control and ischemia group at rest. In the ischemia group MBF under stress was 74.0 ± 21.9 mL/100 mL/min in the poststenotic myocardium and 117.4 ± 18.6 mL/100 mL/min in the remaining normal myocardium (P = 0.0024). Conclusion:DSCT permits quantitative whole heart perfusion imaging. As this technique is able to show the hemodynamic effect of high grade coronary artery stenosis, it exceeds the present key limitation of cardiac computed tomography, which currently only allows a morphologic assessment of coronary artery stenosis.


Journal of Cardiovascular Computed Tomography | 2009

Feasibility of dual-source cardiac CT angiography with high-pitch scan protocols

Jörg Hausleiter; Bernhard Bischoff; Franziska Hein; Tanja Meyer; Martin Hadamitzky; Carsten Thierfelder; Thomas Allmendinger; Thomas Flohr; Albert Schömig; Stefan Martinoff

BACKGROUND Cardiac CT angiography (CCTA) has become a frequently used diagnostic tool in clinical practice, but concern remains about the radiation exposure. Because of the second x-ray acquisition system, dual-source CT systems might allow for high-pitch CT data acquisition and thus for examination of the whole heart during a single heart beat, with the potential for radiation dose reduction. OBJECTIVE We assessed the feasibility of a high-pitch scan mode with a dual-source CT system. METHODS High-pitch modes were used in patients undergoing CCTA with a dual-source CT system. Diagnostic image quality for cardiac structures and coronary arteries was assessed. Radiation dose was estimated from the scanner-generated dose-length product (DLP). RESULTS CCTA was performed in 14 patients during a single heart beat applying a pitch value of 3.4. Mean heart rate during examination was 56.4+/-8.1 beats/min. Diagnostic image quality for the assessment of larger cardiac structures was obtained in all patients, whereas diagnostic image quality could be achieved in 82% of all coronary segments. With a mean DLP of 145+/-47 mGy x cm, the resulting estimated radiation dose was 2.0+/-0.7 mSv. CONCLUSIONS This proof-of-concept study shows the ability of dual-source CT scanners to scan the whole heart during one single heart beat at low radiation dose.


Investigative Radiology | 2015

Very low-dose (0.15 mGy) chest CT protocols using the COPDGene 2 test object and a third-generation dual-source CT scanner with corresponding third-generation iterative reconstruction software.

Matthew K. Fuld; Thomas Allmendinger; Jered Sieren; Kung-Sik Chan; Junfeng Guo; Eric A. Hoffman

ObjectivesThe purpose of this study was to evaluate the impact of ultralow radiation dose single-energy computed tomographic (CT) acquisitions with Sn prefiltration and third-generation iterative reconstruction on density-based quantitative measures of growing interest in phenotyping pulmonary disease. Materials and MethodsThe effects of both decreasing dose and different body habitus on the accuracy of the mean CT attenuation measurements and the level of image noise (SD) were evaluated using the COPDGene 2 test object, containing 8 different materials of interest ranging from air to acrylic and including various density foams. A third-generation dual-source multidetector CT scanner (Siemens SOMATOM FORCE; Siemens Healthcare AG, Erlangen, Germany) running advanced modeled iterative reconstruction (ADMIRE) software (Siemens Healthcare AG) was used.We used normal and very large body habitus rings at dose levels varying from 1.5 to 0.15 mGy using a spectral-shaped (0.6-mm Sn) tube output of 100 kV(p). Three CT scans were obtained at each dose level using both rings. Regions of interest for each material in the test object scans were automatically extracted. The Hounsfield unit values of each material using weighted filtered back projection (WFBP) at 1.5 mGy was used as the reference value to evaluate shifts in CT attenuation at lower dose levels using either WFBP or ADMIRE. Statistical analysis included basic statistics, Welch t tests, multivariable covariant model using the F test to assess the significance of the explanatory (independent) variables on the response (dependent) variable, and CT mean attenuation, in the multivariable covariant model including reconstruction method. ResultsMultivariable regression analysis of the mean CT attenuation values showed a significant difference with decreasing dose between ADMIRE and WFBP. The ADMIRE has reduced noise and more stable CT attenuation compared with WFBP. There was a strong effect on the mean CT attenuation values of the scanned materials for ring size (P < 0.0001) and dose level (P < 0.0001). The number of voxels in the region of interest for the particular material studied did not demonstrate a significant effect (P > 0.05). The SD was lower with ADMIRE compared with WFBP at all dose levels and ring sizes (P < 0.05). ConclusionsThe third-generation dual-source CT scanners using third-generation iterative reconstruction methods can acquire accurate quantitative CT images with acceptable image noise at very low-dose levels (0.15 mGy). This opens up new diagnostic and research opportunities in CT phenotyping of the lung for developing new treatments and increased understanding of pulmonary disease.


Journal of Computer Assisted Tomography | 2012

Validation of dual-source single-tube reconstruction as a method to obtain half-dose images to evaluate radiation dose and noise reduction: phantom and human assessment using CT colonography and sinogram-affirmed iterative reconstruction (SAFIRE).

Joel G. Fletcher; Jeff L. Fidler; Maria Shiung; Lifeng Yu; Jia Wang; Bernhard Schmidt; Thomas Allmendinger; Cynthia H. McCollough

Objective To evaluate a method for obtaining half-dose CT images for observer studies evaluating lower-dose CT. Methods Phantoms of varying sizes were scanned at multiple tube potentials using dose-matched dual-source (DS) and single-source (SS) protocols. Images from single-tube reconstruction of DS data were compared with SS images acquired at half-original CTDIvol. Thirty patients underwent supine SS and dose-matched prone DS CT colonography (CTC). Half-dose prone images were reconstructed with sinogram-affirmed iterative reconstruction (SAFIRE). Two radiologists scored image quality on 2-dimensional (2D) and 3D images. Results Image noise was similar between half-dose SS images and DS images reconstructed from one tube only with tube potential of 120 kV or more for phantoms 40 cm or smaller (P < 0.05). For both readers, the patients’ CTC image quality scores were more than 84% concordant between SS or DS CTC images, and half-dose–prone CTC images with SAFIRE had 84% or more concordance with routine-dose CTC except for 3D image noise. Conclusions In appropriately sized patients, DS acquisition with single-tube reconstruction can create half-dose images, permitting comparison to full-dose images. For CTC, there is comparable image quality for colonic evaluation between full-dose and half-dose images reconstructed with SAFIRE.


American Journal of Roentgenology | 2013

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

Xinhui Duan; Jia Wang; Shuai Leng; Bernhard Schmidt; Thomas Allmendinger; Thomas Flohr; Cynthia H. McCollough

OBJECTIVE The objective of our study was to evaluate in phantoms the differences in CT image noise and artifact level between two types of commercial CT detectors: one with distributed electronics (conventional) and one with integrated electronics intended to decrease system electronic noise. MATERIALS AND METHODS Cylindric water phantoms of 20, 30, and 40 cm in diameter were scanned using two CT scanners, one equipped with integrated detector electronics and one with distributed detector electronics. All other scanning parameters were identical. Scans were acquired at four tube potentials and 10 tube currents. Semianthropomorphic phantoms were scanned to mimic the shoulder and abdominal regions. Images of two patients were also selected to show the clinical values of the integrated detector. RESULTS Reduction of image noise with the integrated detector depended on phantom size, tube potential, and tube current. Scans that had low detected signal had the greatest reductions in noise, up to 40% for a 30-cm phantom scanned using 80 kV. This noise reduction translated into up to 50% in dose reduction to achieve equivalent image noise. Streak artifacts through regions of high attenuation were reduced by up to 45% on scans obtained using the integrated detector. Patient images also showed superior image quality for the integrated detector. CONCLUSION For the same applied radiation level, the use of integrated electronics in a CT detector showed a substantially reduced level of electronic noise, resulting in reductions in image noise and artifacts, compared with detectors having distributed electronics.


Radiology | 2014

Iterative Image Reconstruction Techniques for CT Coronary Artery Calcium Quantification: Comparison with Traditional Filtered Back Projection in Vitro and in Vivo

Andreas Schindler; Rozemarijn Vliegenthart; U. Joseph Schoepf; Philipp Blanke; Ullrich Ebersberger; Young Jun Cho; Thomas Allmendinger; Sebastian Vogt; Rainer Raupach; Christian Fink; Tobias Saam; Fabian Bamberg; Konstantin Nikolaou; Paul Apfaltrer

PURPOSE To investigate in vitro and in vivo the use of image-based and raw data-based iterative reconstruction algorithms for quantification of coronary artery calcium by using the Agatston score and subsequent cardiac risk stratification. MATERIALS AND METHODS In vitro data were obtained by using a moving anthropomorphic cardiac phantom containing calcium inserts of different concentrations and sizes. With institutional review board approval and HIPAA compliance, coronary calcium imaging data of 110 consecutive patients (mean age ± standard deviation, 58.2 years ± 9.8; 48 men) were reconstructed with filtered back projection (FBP), iterative reconstruction in image space (IRIS), and sinogram-affirmed iterative reconstruction (SAFIRE). Image noise was measured and the Agatston score was obtained for all reconstructions. Assignment to Agatston scores and percentile-based cardiac risk categories was compared. Statistical analysis included the Cohen κ coefficient and Friedman and Wilcoxon testing. RESULTS In vitro, mean Agatston scores ± standard deviation for FBP (638.9 ± 9.6), IRIS (622.7 ± 15.2), and SAFIRE (631.4 ± 17.6) were comparable (P = .30). The smallest phantom calcifications were more frequently detected when iterative reconstruction techniques were used. The Agatston scores in the patient cohort were not significantly different among FBP, IRIS, and SAFIRE in paired comparisons (median Agatston score [25th and 75th percentiles]: 76.0 [20.6, 243.9], 76.4 [22, 249.3], and 75.7 [21.5, 49.1], respectively; P = .20 each). Comparison of categorization based on Agatston score percentiles showed excellent agreement for both IRIS and SAFIRE with FBP (κ = 0.975 [0.942-1.00] and κ = 0.963 [0.922-1.00], respectively). The mean effective dose was 1.02 mSv ± 0.51. Mean image noise was significantly (P < .001) higher with FBP than that with iterative reconstructions. CONCLUSION In comparison with FBP, iterative reconstruction techniques do not have a profound effect on the reproducible quantification of coronary calcium according to Agatston score and subsequent cardiac risk classification, although risk reclassification may occur in a small subset of subjects.

Collaboration


Dive into the Thomas Allmendinger's collaboration.

Top Co-Authors

Avatar

Michael Uder

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

U. Joseph Schoepf

Medical University of South Carolina

View shared research outputs
Top Co-Authors

Avatar
Researchain Logo
Decentralizing Knowledge