Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Todsaporn Fuangrod is active.

Publication


Featured researches published by Todsaporn Fuangrod.


Medical Physics | 2013

Gantry-angle resolved VMAT pretreatment verification using EPID image prediction

Henry C. Woodruff; Todsaporn Fuangrod; Pejman Rowshanfarzad; Boyd McCurdy; Peter B. Greer

PURPOSE Pretreatment verification of volumetric modulated arc therapy (VMAT) dose delivery with electronic portal imaging device (EPID) uses images integrated over the entire delivery or over large subarcs. This work aims to develop a new method for gantry-angle-resolved verification of VMAT dose delivery using EPID. METHODS An EPID dose prediction model was used to calculate EPID images as a function of gantry angle for eight prostate patient deliveries. EPID image frames at 7.5 frames per second were acquired during delivery via a frame-grabber system. The gantry angle for each image was encoded in kV frames which were synchronized to the MV frames. Gamma analysis results as a function of gantry angle were assessed by integrating the frames over 2° subarcs with an angle-to-agreement tolerance of 0.5° about the measured image angle. RESULTS The model agreed with EPID images integrated over the entire delivery with average Gamma pass-rates at 2%, 2 mm of 99.7% (10% threshold). The accuracy of the kV derived gantry angle for each image was found to be 0.1° (1 SD) using a phantom test. For the gantry-resolved analysis all Gamma pass-rates were greater than 90% at 3%, 3 mm criteria (with only two exceptions), and more than 90% had a 95% pass-rate, with an average of 97.3%. The measured gantry angle lagged behind the predicted angle by a mean of 0.3°±0.3°, with a maximum lag of 1.3°. CONCLUSIONS The method provides a comprehensive and highly efficient pretreatment verification of VMAT delivery using EPID. Dose delivery accuracy is assessed as a function of gantry angle to ensure accurate treatment.


Medical Physics | 2013

A system for EPID-based real-time treatment delivery verification during dynamic IMRT treatment

Todsaporn Fuangrod; Henry C. Woodruff; Eric Van Uytven; Boyd McCurdy; Zdenka Kuncic; D.J. O'Connor; Peter B. Greer

PURPOSE To design and develop a real-time electronic portal imaging device (EPID)-based delivery verification system for dynamic intensity modulated radiation therapy (IMRT) which enables detection of gross treatment delivery errors before delivery of substantial radiation to the patient. METHODS The system utilizes a comprehensive physics-based model to generate a series of predicted transit EPID image frames as a reference dataset and compares these to measured EPID frames acquired during treatment. The two datasets are using MLC aperture comparison and cumulative signal checking techniques. The system operation in real-time was simulated offline using previously acquired images for 19 IMRT patient deliveries with both frame-by-frame comparison and cumulative frame comparison. Simulated error case studies were used to demonstrate the system sensitivity and performance. RESULTS The accuracy of the synchronization method was shown to agree within two control points which corresponds to approximately ∼1% of the total MU to be delivered for dynamic IMRT. The system achieved mean real-time gamma results for frame-by-frame analysis of 86.6% and 89.0% for 3%, 3 mm and 4%, 4 mm criteria, respectively, and 97.9% and 98.6% for cumulative gamma analysis. The system can detect a 10% MU error using 3%, 3 mm criteria within approximately 10 s. The EPID-based real-time delivery verification system successfully detected simulated gross errors introduced into patient plan deliveries in near real-time (within 0.1 s). CONCLUSIONS A real-time radiation delivery verification system for dynamic IMRT has been demonstrated that is designed to prevent major mistreatments in modern radiation therapy.


International Journal of Radiation Oncology Biology Physics | 2015

First Experience With Real-Time EPID-Based Delivery Verification During IMRT and VMAT Sessions

Henry C. Woodruff; Todsaporn Fuangrod; Eric Van Uytven; Boyd McCurdy; Timothy van Beek; Shashank Bhatia; Peter B. Greer

PURPOSE Gantry-mounted megavoltage electronic portal imaging devices (EPIDs) have become ubiquitous on linear accelerators. WatchDog is a novel application of EPIDs, in which the image frames acquired during treatment are used to monitor treatment delivery in real time. We report on the preliminary use of WatchDog in a prospective study of cancer patients undergoing intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) and identify the challenges of clinical adoption. METHODS AND MATERIALS At the time of submission, 28 cancer patients (head and neck, pelvis, and prostate) undergoing fractionated external beam radiation therapy (24 IMRT, 4 VMAT) had ≥1 treatment fraction verified in real time (131 fractions or 881 fields). EPID images acquired continuously during treatment were synchronized and compared with model-generated transit EPID images within a frame time (∼0.1 s). A χ comparison was performed to cumulative frames to gauge the overall delivery quality, and the resulting pass rates were reported graphically during treatment delivery. Every frame acquired (500-1500 per fraction) was saved for postprocessing and analysis. RESULTS The system reported the mean ± standard deviation in real time χ 91.1% ± 11.5% (83.6% ± 13.2%) for cumulative frame χ analysis with 4%, 4 mm (3%, 3 mm) criteria, global over the integrated image. CONCLUSIONS A real-time EPID-based radiation delivery verification system for IMRT and VMAT has been demonstrated that aims to prevent major mistreatments in radiation therapy.


Journal of Applied Clinical Medical Physics | 2016

An EPID-based system for gantry-resolved MLC quality assurance for VMAT.

Benjamin J. Zwan; Michael P. Barnes; Todsaporn Fuangrod; Cameron J. Stanton; D.J. O'Connor; P Keall; Peter B. Greer

Multileaf collimator (MLC) positions should be precisely and independently measured as a function of gantry angle as part of a comprehensive quality assurance (QA) program for volumetric‐modulated arc therapy (VMAT). It is also ideal that such a QA program has the ability to relate MLC positional accuracy to patient‐specific dosimetry in order to determine the clinical significance of any detected MLC errors. In this work we propose a method to verify individual MLC trajectories during VMAT deliveries for use as a routine linear accelerator QA tool. We also extend this method to reconstruct the 3D patient dose in the treatment planning system based on the measured MLC trajectories and the original DICOM plan file. The method relies on extracting MLC positions from EPID images acquired at 8.41 fps during clinical VMAT deliveries. A gantry angle is automatically tagged to each image in order to obtain the MLC trajectories as a function of gantry angle. This analysis was performed for six clinical VMAT plans acquired at monthly intervals for three months. The measured trajectories for each delivery were compared to the MLC positions from the DICOM plan file. The maximum mean error detected was 0.07 mm and a maximum root‐mean‐square error was 0.8 mm for any leaf of any delivery. The sensitivity of this system was characterized by introducing random and systematic MLC errors into the test plans. It was demonstrated that the system is capable of detecting random and systematic errors on the range of 1–2 mm and single leaf calibration errors of 0.5 mm. The methodology developed in the work has potential to be used for efficient routine linear accelerator MLC QA and pretreatment patient‐specific QA and has the ability to relate measured MLC positional errors to 3D dosimetric errors within a patient volume. PACS number(s): 87.55.QrMultileaf collimator (MLC) positions should be precisely and independently measured as a function of gantry angle as part of a comprehensive quality assurance (QA) program for volumetric-modulated arc therapy (VMAT). It is also ideal that such a QA program has the ability to relate MLC positional accuracy to patient-specific dosimetry in order to determine the clinical significance of any detected MLC errors. In this work we propose a method to verify individual MLC trajectories during VMAT deliveries for use as a routine linear accelerator QA tool. We also extend this method to reconstruct the 3D patient dose in the treatment planning system based on the measured MLC trajectories and the original DICOM plan file. The method relies on extracting MLC positions from EPID images acquired at 8.41 fps during clinical VMAT deliveries. A gantry angle is automatically tagged to each image in order to obtain the MLC trajectories as a function of gantry angle. This analysis was performed for six clinical VMAT plans acquired at monthly intervals for three months. The measured trajectories for each delivery were compared to the MLC positions from the DICOM plan file. The maximum mean error detected was 0.07 mm and a maximum root-mean-square error was 0.8 mm for any leaf of any delivery. The sensitivity of this system was characterized by introducing random and systematic MLC errors into the test plans. It was demonstrated that the system is capable of detecting random and systematic errors on the range of 1-2 mm and single leaf calibration errors of 0.5 mm. The methodology developed in the work has potential to be used for efficient routine linear accelerator MLC QA and pretreatment patient-specific QA and has the ability to relate measured MLC positional errors to 3D dosimetric errors within a patient volume. PACS number(s): 87.55.Qr.


Physics in Medicine and Biology | 2017

Quantifying the accuracy and precision of a novel real-time 6 degree-of-freedom kilovoltage intrafraction monitoring (KIM) target tracking system

J Kim; Doan Trang Nguyen; C-Y Huang; Todsaporn Fuangrod; Vincent Caillet; Ricky O’Brien; P.R. Poulsen; Jeremy T. Booth; P Keall

Target rotation can considerably impact the delivered radiotherapy dose depending on the tumour shape. More accurate tumour pose during radiotherapy treatment can be acquired through tracking in 6 degrees-of-freedom (6 DoF) rather than in translation only. A novel real-time 6 DoF kilovoltage intrafraction monitoring (KIM) target tracking system has recently been developed. In this study, we experimentally evaluated the accuracy and precision of the 6 DoF KIM implementation. Real-time 6 DoF KIM motion measurements were compared against the ground truth motion retrospectively derived from kV/MV triangulation for a range of lung and prostate tumour motion trajectories as well as for various static poses using a phantom. The accuracy and precision of 6 DoF KIM were calculated as the mean and standard deviation of the differences between KIM and kV/MV triangulation for each DoF, respectively. We found that KIM is able to provide 6 DoF motion with sub-degree and sub-millimetre accuracy and precision for a range of realistic tumour motion.


Medical Physics | 2017

Commissioning and quality assurance for VMAT delivery systems: An efficient time‐resolved system using real‐time EPID imaging

Benjamin J. Zwan; Michael P. Barnes; Jonathan Hindmarsh; Seng B. Lim; D Lovelock; Todsaporn Fuangrod; D.J. O'Connor; P Keall; Peter B. Greer

Purpose: An ideal commissioning and quality assurance (QA) program for Volumetric Modulated Arc Therapy (VMAT) delivery systems should assess the performance of each individual dynamic component as a function of gantry angle. Procedures within such a program should also be time‐efficient, independent of the delivery system and be sensitive to all types of errors. The purpose of this work is to develop a system for automated time‐resolved commissioning and QA of VMAT control systems which meets these criteria. Methods: The procedures developed within this work rely solely on images obtained, using an electronic portal imaging device (EPID) without the presence of a phantom. During the delivery of specially designed VMAT test plans, EPID frames were acquired at 9.5 Hz, using a frame grabber. The set of test plans was developed to individually assess the performance of the dose delivery and multileaf collimator (MLC) control systems under varying levels of delivery complexities. An in‐house software tool was developed to automatically extract features from the EPID images and evaluate the following characteristics as a function of gantry angle: dose delivery accuracy, dose rate constancy, beam profile constancy, gantry speed constancy, dynamic MLC positioning accuracy, MLC speed and acceleration constancy, and synchronization between gantry angle, MLC positioning and dose rate. Machine log files were also acquired during each delivery and subsequently compared to information extracted from EPID image frames. Results: The largest difference between measured and planned dose at any gantry angle was 0.8% which correlated with rapid changes in dose rate and gantry speed. For all other test plans, the dose delivered was within 0.25% of the planned dose for all gantry angles. Profile constancy was not found to vary with gantry angle for tests where gantry speed and dose rate were constant, however, for tests with varying dose rate and gantry speed, segments with lower dose rate and higher gantry speed exhibited less profile stability. MLC positional accuracy was not observed to be dependent on the degree of interdigitation. MLC speed was measured for each individual leaf and slower leaf speeds were shown to be compensated for by lower dose rates. The test procedures were found to be sensitive to 1 mm systematic MLC errors, 1 mm random MLC errors, 0.4 mm MLC gap errors and synchronization errors between the MLC, dose rate and gantry angle controls systems of 1° In general, parameters measured by both EPID and log files agreed with the plan, however, a greater average departure from the plan was evidenced by the EPID measurements. Conclusion: QA test plans and analysis methods have been developed to assess the performance of each dynamic component of VMAT deliveries individually and as a function of gantry angle. This methodology relies solely on time‐resolved EPID imaging without the presence of a phantom and has been shown to be sensitive to a range of delivery errors. The procedures developed in this work are both comprehensive and time‐efficient and can be used for streamlined commissioning and QA of VMAT delivery systems.


Medical Physics | 2012

MO‐G‐213AB‐03: Simulations of Real‐Time Geometric and Dosimetic Verification System Using EPID

Todsaporn Fuangrod; Henry C. Woodruff; E. VanUytven; B McCurdy; D.J. O'Connor; Peter B. Greer

Purpose: To demonstrate a new method for real‐time geometric and dosimetric verification of IMRT and VMAT using synchronization between predicted and measured EPIDimages.Methods: Predicted EPIDimages were calculated using a comprehensive physics‐based model. Each predicted image represents the integrated signal expected from the delivery between control points. The measured images are acquired in cine mode and compared to the set of predicted images in real‐time. The system performs geometric verification prior to dosimetric verification. When the measured image is acquired, the algorithm automatically detects the MLC leaf positions. A comparison between the leaf positions of the measured image and control points in the MLC file is made using the cosine similarity technique. The similarity index(SI) provides geometric MLC verification and synchronization between the measured and predicted images, as a uniform dose‐rate cannot be assumed for IMRT or VMAT deliveries. The SI threshold was based on a series of experiments including 21 dynamic‐IMRT fields defining pass/fail boundary(5 brain, 8 H&N, and 8 prostate cases).If geometric verification is successful, dosimetic verification is performed with the Gamma comparison(3%,3mm).The system reports the verification Result in real‐time. Results: The system was simulated by MATLAB/SIMULINK and detected geometric and dosimetric errors during delivery. Both artificially introduced errors and clinical data were used for testing and analysis of the system performance. For a tested prostate field, the cumulative dose comparisons showed the minimum and maximum number of points with Gamma index<1 as 93.5% and 98.5%, respectively. For individual dose comparisons on the same field, the values were 87% and 97%, respectively. Conclusion: This method includes automatic MLC leaf positioning, synchronization, and dosimetric verification. The pass/fail boundary of geometry was calculated based on the experiments. This system is a useful approach to detect unexpected possible errors occurring in the clinical setting and to prevent patient overdoses during radiotherapy especially in complex deliveries such as arc‐IMRT.


Radiotherapy and Oncology | 2018

The accuracy and precision of Kilovoltage Intrafraction Monitoring (KIM) six degree-of-freedom prostate motion measurements during patient treatments

Jung-Ha Kim; Doan Trang Nguyen; Jeremy T. Booth; Chen-Yu Huang; Todsaporn Fuangrod; P.R. Poulsen; R. O'Brien; Vincent Caillet; Thomas Eade; Andrew Kneebone; P Keall

BACKGROUND AND PURPOSE To perform a quantitative analysis of the accuracy and precision of Kilovoltage Intrafraction Monitoring (KIM) six degree-of-freedom (6DoF) prostate motion measurements during treatments. MATERIAL AND METHODS Real-time 6DoF prostate motion was acquired using KIM for 14 prostate cancer patients (377 fractions). KIM outputs the 6DoF prostate motion, combining 3D translation and 3D rotational motion information relative to its planning position. The corresponding groundtruth target motion was obtained post-treatment based on kV/MV triangulation. The accuracy and precision of the 6DoF KIM motion estimates were calculated as the mean and standard deviation differences compared with the ground-truth. RESULTS The accuracy ± precision of real-time 6DoF KIM-measured prostate motion were 0.2 ± 1.3° for rotations and 0.1 ± 0.5 mm for translations, respectively. The magnitude of KIM-measured motion was well-correlated with the magnitude of ground-truth motion resulting in Pearson correlation coefficients of  ≥0.88 in all DoF. CONCLUSIONS The results demonstrate that KIM is capable of providing the real-time 6DoF prostate target motion during patient treatments with an accuracy ± precision of within 0.2 ± 1.3° and 0.1 ± 0.5 mm for rotation and translation, respectively. As KIM only requires a single X-ray imager, which is available on most modern cancer radiotherapy devices, there is potential for widespread adoption of this technology.


Journal of Applied Clinical Medical Physics | 2017

A novel and independent method for time-resolved gantry angle quality assurance for VMAT

Todsaporn Fuangrod; Peter B. Greer; Benjamin J. Zwan; Michael P. Barnes; Joerg Lehmann

Abstract Volumetric‐modulated arc therapy (VMAT) treatment delivery requires three key dynamic components; gantry rotation, dose rate modulation, and multi‐leaf collimator motion, which are all simultaneously varied during the delivery. Misalignment of the gantry angle can potentially affect clinical outcome due to the steep dose gradients and complex MLC shapes involved. It is essential to develop independent gantry angle quality assurance (QA) appropriate to VMAT that can be performed simultaneously with other key VMAT QA testing. In this work, a simple and inexpensive fully independent gantry angle measurement methodology was developed that allows quantitation of the gantry angle accuracy as a function of time. This method is based on the analysis of video footage of a “Double dot” pattern attached to the front cover of the linear accelerator that consists of red and green circles printed on A4 paper sheet. A standard mobile phone is placed on the couch to record the video footage during gantry rotation. The video file is subsequently analyzed and used to determine the gantry angle from each video frame using the relative position of the two dots. There were two types of validation tests performed including the static mode with manual gantry angle rotation and dynamic mode with three complex test plans. The accuracy was 0.26° ± 0.04° and 0.46° ± 0.31° (mean ± 1 SD) for the static and dynamic modes, respectively. This method is user friendly, cost effective, easy to setup, has high temporal resolution, and can be combined with existing time‐resolved method for QA of MLC and dose rate to form a comprehensive set of procedures for time‐resolved QA of VMAT delivery system.


Radiotherapy and Oncology | 2016

OC-0361: Simulation of clinical relevance errors detected by real-time EPID-based patient verification system

Todsaporn Fuangrod; J. Simpson; R. Middleton; Peter B. Greer

Conclusion: The σ results of this study should be an indicator of the overall positioning uncertainty in our IGRT process for these treatments, i.e. kVCT-MVCT image registration, patient movement, and respiratory motion. Even if only one projection of the treatment data was used in the estimation, our results compare very well with similar studies’ (von Tienhoven et al (1991), Smith et al (2005), Wang et al (2013)) findings on breast displacement due to respiratory motion. Furthermore, the novelty of this study is its evaluation of the breast position was performed on exit detector fluence of intensity-modulated fields, which we believe to be a first.

Collaboration


Dive into the Todsaporn Fuangrod's collaboration.

Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

P Keall

University of Sydney

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Researchain Logo
Decentralizing Knowledge