Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Walter W. Peppler is active.

Publication


Featured researches published by Walter W. Peppler.


Calcified Tissue International | 1981

Total body bone mineral and lean body mass by dual-photon absorptiometry. I. Theory and measurement procedure.

Walter W. Peppler; Richard B. Mazess

SummaryA method was developed for measuring total body bone mineral (TBBM) and lean body mass in vivo using dual-photon absorptiometry. The entire body was scanned in a rectilinear raster (transverse speed of 1 cm/s and longitudinal steps of 2.5 cm) with a modified nuclear medicine scanner and conventional nuclear counting electronics. The source was153Gd (1 Ci) with principal photopeaks at 44 and 100 keV. The scan time was about 70 min with an absorbed dose of under 1 mrem. The low dose allows measurements to be repeated at frequent intervals or used on children. Short-term (months) precision of TBBM was about 1.5% for isolated skeletons and about 2% on normal human subjects. Long-term (years) precision on skeletons was under 3%. The precision of percent fat was 0.9%, which would lead to an error of less than 1% in the TBBM. Geometry of measurements also had minimal (and correctable) influence on the accuracy of results. The accuracy (1 standard error of estimate) of TBBM on isolated skeletons (N=5) was 36 g (equivalent to about 13 g of Ca) with a correlation coefficient of 0.99; this error amounts to about 1–1.5% in normal adults, 2% in older women, and 2.5% in osteoporotic females. The dual-photon absorptiometry method could be implemented in many nuclear medicine departments to follow skeletal changes during growth and aging or to follow the course of a disease or treatment.


Calcified Tissue International | 1981

Total body bone mineral and lean body mass by dual-photon absorptiometry: II. Comparison with total body calcium by neutron activation analysis

Richard B. Mazess; Walter W. Peppler; Joan E. Harrison; McNeill Kg

SummaryTotal body bone mineral (TBBM) was measured in vivo using dual-photon absorptiometry (DPA) in 10 subjects. The total body calcium (TBCa) was measured in the same subjects using neutron activation analysis. The correlation between the two methods was very high (r>0.99) and the standard error of estimate was low. The TBCa relative to TBBM was about 39%. The two noninvasive methods provided nearly identical indications of skeletal mass, but the radiation exposure with DPA was 500 to 5000 times smaller (0.6 mrem vs 300 to 3000 mrem). The radius shaft bone mineral content was highly correlated with the TBBM (0.97) and the TBCa (0.98) and could be used to estimate the latter variables with errors (1 SEE) of 9% and 6%, respectively.


Calcified Tissue International | 1984

Total body and regional bone mineral by dual-photon absorptiometry in metabolic bone disease

Richard B. Mazess; Walter W. Peppler; Russell W. Chesney; Lange Ta; Urban Lindgren; Everett L. Smith

SummaryDual-photon absorptiometry (153Gd) was used to measure bone mineral of the total body and major anatomical areas. Patients with osteoporosis (♂=11, ♀=18) and with renal osteodystrophy (n=17) were significantly below (20%) normal females (n=72) and males (n=13) at most sites. In the osteoporotic patients, but not the renal patients, there was preferential osteopenia of the spine. Bone loss in all anatomical areas became evident after the menopause with an annual loss rate of about 0.7%.


Medical Physics | 1994

A regional convolution kernel algorithm for scatter correction in dual-energy images: comparison to single-kernel algorithms.

David G. Kruger; Frank E. Zink; Walter W. Peppler; David L. Ergun; Charles A. Mistretta

Single kernel scatter correction algorithms are based on the model that the scatter field can be predicted by convolution of the primary intensity (Iprim) with a spatially invariant scatter point-spread function (PSF). Practical limitations (Iprim unknown) suggest the substitution of the total detected intensity (Idet) for Iprim as the source image in the convolution. In regions of high scatter fraction (SF), Idet is a poor approximation of Iprim, thereby causing an overestimation of scatter originating in the region. This contributes to errors in estimating detected scatter in the mediastinum and neighboring regions. A technique using a regionally variable point-spread function that significantly reduces RMS error in estimation of the primary image as compared to the single PSF method is investigated. The regionally variable convolution method employs a larger PSF in the mediastinum and a smaller PSF in the lungs to reduce the error in estimating the scatter throughout the image. The method to allow for patient differences has also been expanded and various implementations of these methods have been compared. Results show that the dual-kernel algorithm is always more effective than an equivalent single-kernel algorithm. The dual-kernel algorithm using a predicted scatter fraction curve gives an overall RMS error in the primary of as low as 20.8% which is equivalent to 8.7% RMS error in the scatter. The dual-kernel method using a predicted scatter fraction curve approaches the accuracy of the single-kernel method using patient specific scatter measurements. Because using individual scatter measurements is a less desirable method for clinical use, we feel that the dual-kernel algorithm which uses two regions specific convolution kernels and a variable scatter fraction curve is the preferable method.


Medical Physics | 1989

A correlated noise reduction algorithm for dual-energy digital subtraction angiography.

Cynthia H. McCollough; Michael S. Van Lysel; Walter W. Peppler; Charles A. Mistretta

It has long been recognized that the problems of motion artifacts in conventional time subtraction digital subtraction angiography (DSA) may be overcome using energy subtraction techniques. Of the variety of energy subtraction techniques investigated, non-k-edge dual-energy subtraction offers the best signal-to-noise ratio (SNR). However, this technique achieves only 55% of the temporal DSA SNR. Noise reduction techniques that average the noisier high-energy image produce various degrees of noise improvement while minimally affecting iodine contrast and resolution. A more significant improvement in dual-energy DSA iodine SNR, however, results when the correlated noise that exists in material specific images is appropriately cancelled. The correlated noise reduction (CNR) algorithm presented here follows directly from the dual-energy computed tomography work of Kalender who made explicit use of noise correlations in material specific images to reduce noise. The results are identical to those achieved using a linear version of the two-stage filtering process described by Macovski in which the selective image is filtered to reduce high-frequency noise and added to a weighted, high SNR, nonselective image which has been processed with a high-frequency bandpass filter. The dual-energy DSA CNR algorithm presented here combines selective tissue and iodine images to produce a significant increase in the iodine SNR while fully preserving iodine spatial resolution. Theoretical calculations predict a factor of 2-4 improvement in SNR compared to conventional dual-energy images. The improvement factor achieved is dependent upon the x-ray beam spectra and the size of blurring kernel used in the algorithm.(ABSTRACT TRUNCATED AT 250 WORDS)


Medical Physics | 1996

Imaging characteristics of x‐ray capillary optics in digital mammography

David G. Kruger; Carmen C. Abreu; Eric G. Hendee; Armen Kocharian; Walter W. Peppler; Charles A. Mistretta; Carolyn A. MacDonald

Computed radiography (CR) has shown promise in digital mammographic screening due to its good low spatial frequency MTF and its relatively wide exposure latitude. The CR image format has not gained acceptance clinically because of reduced high spatial frequency resolution as compared to film-screen images. X-ray capillary optics, aligned between the breast and CR phosphor imaging plate, will capture primary x-ray photons almost exclusively. Due to the very small angle of acceptance, scattered photons angled more than about 1.6 x 10(-3) radians from primary trajectory will not be accepted at the capillary optic entrance. The virtual elimination of detected scatter means almost 100% of the possible primary contrast should be visible in the image. In addition, the image can be magnified without focal spot blurring. Effective resolution of CR images can be increased by a factor equal to that magnification. Clinical implementation of future capillary optics are expected to be either in the form of a large, stationary, post-patient optic that accepts primary from the entire breast or a fan-shaped optic that is scanned across the breast. Measurements of a test capillary optic showed a reduction of scatter fraction to 0.018. Images of a lucite contrast detail phantom revealed a corresponding increase in image contrast when compared to anti-scatter grid and no grid methods. Spectral transmission measurements using a high-purity germanium detector showed good primary transmission (45%-50%) in the mammographic energy range. The MTF measurements of both stationary and scanned capillary optics showed improvement at the 5% MTF level to 8.4 mm-1 for scanned optics and 9.2 mm-1 for stationary optics representing a 68% and 84% respective increase over the CR MTF without magnification or capillary optics.


Medical Physics | 1991

Coronary blood flow measurement using an angiographic first pass distribution technique: A feasibility study

Nicholas J. Hangiandreou; John D. Folts; Walter W. Peppler; Charles A. Mistretta

Due to the well-documented problems associated with visual interpretation of coronary angiograms, more physiologic means of assessing coronary artery stenosis are being investigated. One physiologic parameter that has been suggested is coronary flow reserve (CFR). A digital subtraction angiographic technique based on first pass distribution analysis (FPA) is proposed as a means of measuring CFR and absolute coronary flow. The theory of the FPA method is first outlined, and the implementation of a preliminary version of the FPA algorithm is described. Experiments verifying the utility of this algorithm for measuring absolute flow through a flow phantom, and through the canine circumflex artery are reported. It was determined that the preliminary FPA algorithm is capable of measuring canine coronary flow ratios (R) with accuracy and precision characteristics meeting or exceeding those reported for the parametric imaging technique (RFPA = 0.933.Rtrue, SEE = 0.16, r = 0.984). Accurate absolute flow (Q) measurements were obtained in all of the phantom experiments (QFPA = 1.054.Qtrue, r = 0.993), and in one of the three dogs that were studied (QFPA = 0.977.Qtrue, r = 0.935). The difficulty encountered in the other two dog experiments is attributed to the effects of system temporal lag, and would likely be corrected through the use of improved cameras. The feasibility of the general FPA method for measuring relative flow is established, and the potential for routine, absolute flow measurement is demonstrated.


Medical Physics | 1999

Use of a slit camera for MTF measurements

Carla D. Bradford; Walter W. Peppler; Jacob M. Waidelich

The slit camera was analyzed in order to establish its utility and limitations as an MTF measurement tool for characterizing radiographic imaging systems. Commercial slit cameras are attractive for MTF measurements because the beveled edges significantly reduce their alignment sensitivity as compared to the conventional parallel jaw slit. Radiation passing through the beveled edges increases the effective width of the slit camera so that a correction based on the nominal slit width would leave residual error in the MTF measurement. Experimental and Monte Carlo simulated MTF measurements were made on a slit camera (10 microm nominal slit width) in order to estimate its sensitivity in alignment, quantify the error in MTF due to transmission through the beveled jaws, and provide a correction factor. The alignment tolerances of the slit camera were found to be about 12 times larger than for the parallel jaw slit at small HVLs (approximately 1.3 mm Al) of the incident beam and 9 times larger at higher HVLs (approximately 7 mm Al). The magnitude of the residual error in MTF was dependent on the quality of the incident spectrum. For incident spectra with high kVp and HVL (> or = 120 kVp, > or =5 mm Al HVL), transmission through the beveled edges produced errors in MTF up to 15% at 5 cycles/mm and 30% at 10 cycles/mm. By assuming a rectangular slit profile with an effective width based on the kVp, HVL, and filtration material of the incident beam, an MTF correction factor was determined. Application of this correction factor reduced the errors to less than 4% up to 10 cycles/mm. At low beam energies and spatial frequencies, the correction is less critical. Ease of alignment and greater availability make a commercial slit camera useful for MTF measurements. Accurate MTF measurements can be made if appropriate correction factors are applied.


Calcified Tissue International | 1975

Vertebral and Total Body Bone Mineral Content by Dual Photon Absorptiometry

M. Madsen; Walter W. Peppler; Richard B. Mazess

ConclusionsTechniques have been developed using dual-photon absorptiometry for precise and accurate measurement of bone mineral content in the lower spine and in the total body. A spinal measurement requires only 20 min and a total body scan about 60 min. The dose from these measurements is very low allowing them to be repeated at frequent intervals. We feel these improved techniques will prove of great value both for examining the skeleton in normal populations and for evaluation of osteoporosis and other skeletal abnormalities.


Medical Physics | 1995

Measurements of capillary x-ray optics with potential for use in mammographic imaging.

Carmen C. Abreu; David G. Kruger; Carolyn A. MacDonald; Charles A. Mistretta; Walter W. Peppler; Qi-Fan Xiao

Capillary optic arrays are bundles of hollow glass capillaries which guide x rays in a manner similar to the way fiber optics guide light. Focused postpatient capillary optic arrays have the potential to significantly improve both contrast and resolution of mammographic images compared to conventional antiscatter grids. Contrast can be improved by the nearly total scatter rejection of the optic. Effective resolution can be improved by geometric magnification without increased focal spot blurring. The best results were found for borosilicate glasses, with transmissions in excess of 60% for 22-cm-long fibers. To evaluate the scatter rejection properties, the transmission of off-axis radiation was measured. Transmission drops to < 1% at an angular displacement of 2.7 mrad. Transmission of a bulk capillary array dropped to near zero if the source was at an angle of 2.5 mrad. This implies excellent scatter rejection capabilities. To evaluate whether unchanneled photons might still reach the detector, absorption measurements were also performed on fibers and arrays. Absorption was found to be adequate for scatter rejection. All of the data agreed well with numerical simulations. Performance calculations for two potential optics geometries gave promising results.

Collaboration


Dive into the Walter W. Peppler's collaboration.

Top Co-Authors

Avatar

Charles A. Mistretta

University of Wisconsin-Madison

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Michael S. Van Lysel

University of Wisconsin-Madison

View shared research outputs
Top Co-Authors

Avatar

David L. Ergun

University of Wisconsin-Madison

View shared research outputs
Top Co-Authors

Avatar

M. Van Lysel

University of Wisconsin-Madison

View shared research outputs
Top Co-Authors

Avatar

Shaikh Naimuddin

University of Wisconsin-Madison

View shared research outputs
Top Co-Authors

Avatar

Ching-Shan Lee

University of Wisconsin-Madison

View shared research outputs
Top Co-Authors

Avatar

Gary Wendt

University of Wisconsin-Madison

View shared research outputs
Top Co-Authors

Avatar

Jack T. Cusma

University of Wisconsin-Madison

View shared research outputs
Researchain Logo
Decentralizing Knowledge