Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Xing Gong is active.

Publication


Featured researches published by Xing Gong.


Medical Physics | 2006

A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone‐beam CT breast imaging

Xing Gong; Stephen J. Glick; Bob Liu; Aruna A. Vedula; Samta Thacker

Although conventional mammography is currently the best modality to detect early breast cancer, it is limited in that the recorded image represents the superposition of a three-dimensional (3D) object onto a 2D plane. Recently, two promising approaches for 3D volumetric breast imaging have been proposed, breast tomosynthesis (BT) and CT breast imaging (CTBI). To investigate possible improvements in lesion detection accuracy with either breast tomosynthesis or CT breast imaging as compared to digital mammography (DM), a computer simulation study was conducted using simulated lesions embedded into a structured 3D breast model. The computer simulation realistically modeled x-ray transport through a breast model, as well as the signal and noise propagation through a CsI based flat-panel imager. Polyenergetic x-ray spectra of Mo/Mo 28 kVp for digital mammography, Mo/Rh 28 kVp for BT, and W/Ce 50 kVp for CTBI were modeled. For the CTBI simulation, the intensity of the x-ray spectra for each projection view was determined so as to provide a total average glandular dose of 4 mGy, which is approximately equivalent to that given in conventional two-view screening mammography. The same total dose was modeled for both the DM and BT simulations. Irregular lesions were simulated by using a stochastic growth algorithm providing lesions with an effective diameter of 5 mm. Breast tissue was simulated by generating an ensemble of backgrounds with a power law spectrum, with the composition of 50% fibroglandular and 50% adipose tissue. To evaluate lesion detection accuracy, a receiver operating characteristic (ROC) study was performed with five observers reading an ensemble of images for each case. The average area under the ROC curves (Az) was 0.76 for DM, 0.93 for BT, and 0.94 for CTBI. Results indicated that for the same dose, a 5 mm lesion embedded in a structured breast phantom was detected by the two volumetric breast imaging systems, BT and CTBI, with statistically significant higher confidence than with planar digital mammography, while the difference in lesion detection between BT and CTBI was not statistically significant.


Physics in Medicine and Biology | 2004

Microcalcification detection using cone-beam CT mammography with a flat-panel imager

Xing Gong; Aruna A. Vedula; Stephen J. Glick

The purpose of this study was to investigate microcalcification detectability using CT mammography with a flat-panel imager. To achieve this, a computer simulation was developed to model an amorphous-silicon, CsI based flat-panel imager system using a linear cascaded model. The breast was modelled as a hemi-ellipsoid shape with composition of 50% adipose and 50% glandular tissue. Microcalcifications were modelled as small spheres having a composition of calcium carbonate. The results show that with a mean glandular dose equivalent to that typically used in two-view screening mammography, CT mammography with a flat-panel detector is capable of providing images where most microcalcifications are detectable. A receiver operating characteristic (ROC) study was conducted by five physicist observers viewing simulated CT mammography reconstructions. The results suggest that the microcalcification with its diameter equal to or greater than 0.175 mm can be detected with an average area under the ROC curve (AUC) greater than 0.95 using 0.1 or 0.2 mm pixelized detectors. The results also indicate that the optimal pixel size of the detector is around 0.2 mm for microcalcification detection, based on the trade-off between detectability of microcalcifications and the time required for data acquisition and reconstruction.


Medical Physics | 2006

Evaluating the impact of X-ray spectral shape on image quality in flat-panel CT breast imaging.

Stephen J. Glick; Samta Thacker; Xing Gong; Bob Liu

In recent years, there has been an increasing interest in exploring the feasibility of dedicated computed tomography (CT) breast imaging using a flat-panel digital detector in a truncated cone-beam imaging geometry. Preliminary results are promising and it appears as if three-dimensional tomographic imaging of the breast has great potential for reducing the masking effect of superimposed parenchymal structure typically observed with conventional mammography. In this study, a mathematical framework used for determining optimal design and acquisition parameters for such a CT breast imaging system is described. The ideal observer signal-to-noise ratio (SNR) is used as a figure of merit, under the assumptions that the imaging system is linear and shift invariant. Computation of the ideal observer SNR used a parallel-cascade model to predict signal and noise propagation through the detector, as well as a realistic model of the lesion detection task in breast imaging. For all evaluations, the total mean glandular dose for a CT breast imaging study was constrained to be approximately equivalent to that of a two-view conventional mammography study. The framework presented was used to explore the effect of x-ray spectral shape across an extensive range of kVp settings, filter material types, and filter thicknesses. The results give an indication of how spectral shape can affect image quality in flat-panel CT breast imaging.


Medical Imaging 2003: Physics of Medical Imaging | 2003

Computer simulation of CT mammography using a flat-panel imager

Aruna A. Vedula; Stephen J. Glick; Xing Gong

Software has been developed to simulate a cone-beam CT mammography imaging system that consists of an x-ray tube and a flat-panel detector that rotate simultaneously around the pendant breast. The simulation uses an analytical expression or ray-tracing to generate projection sets of breast phantoms at 1 keV intervals dictated by the input x-ray energy spectra. The x-ray focal spot was modeled as having a Gaussian distribution. The detector was modeled as an amorphous silicon (aSi:H) flat-panel imager that uses a structured CsI scintillator. Noise propagation through the detector was simulated by modeling statistical variations of the projection images at each energy interval as a scaled Poisson process. Scintillator blurring was simulated by using an empirically determined modulation transfer function. After introducing noise and detector blur, projection sets simulated at each energy were then combined and reconstructed using Feldkamps cone-beam reconstruction algorithm. Using this framework, the effects of a number of acquisition and reconstruction parameters can be investigated. Some examples are shown including the impact of the kVp setting and the number of projection angles on the reconstructed image.


Medical Imaging 2007: Physics of Medical Imaging | 2007

Characterization of a prototype tabletop x-ray CT breast imaging system

J. Michael O'Connor; Stephen J. Glick; Xing Gong; Clay Didier; Mufeed Mahd

Planar X-ray mammography is the standard medical imaging modality for the early detection of breast cancer. Based on advancements in digital flat-panel detector technology, dedicated x-ray computed tomography (CT) mammography is a modality under investigation that offers the potential for improved breast tumor imaging. We have implemented a prototype half cone-beam CT breast imaging system that utilizes an indirect flat-panel detector. This prototype can be used to explore and evaluate the effect of varying acquisition and reconstruction parameters on image quality. This report describes our system and characterizes the performance of the system through the analysis of Modulation Transfer Function (MTF) and Noise Power Spectrum (NPS). All CT reconstructions were made using Feldkamps filtered backprojection algorithm. The 3D MTF was determined by the analysis of the plane spread function (PlSF) derived from the surface spread function (SSF) of reconstructed 6.3mm spheres. 3D NPS characterization was performed through the analysis of a 3D volume extracted from zero-mean CT noise of air reconstructions. The effect of varying locations on MTF and the effect of different Butterworth filter cutoff frequencies on NPS are reported. Finally, we present CT images of mastectomy excised breast tissue. Breast specimen images were acquired on our CTMS using an x-ray technique similar to the one used during performance characterization. Specimen images demonstrate the inherent CT capability to reduce the masking effect of anatomical noise. Both the quantitative system characterization and the breast specimen images continue to reinforce the hope that dedicated flat-panel detector, x-ray cone-beam CT will eventually provide enhanced breast cancer detection capability.


Medical Imaging 2006: Physics of Medical Imaging | 2006

Optimal spectra for indirect detector breast tomosynthesis

Stephen J. Glick; Xing Gong

The detection of lesions in conventional mammography is a difficult task, predominantly due to the masking effect of superimposed parenchymal breast patterns. Breast tomosynthesis is a technique that has been proposed to reduce this masking effect, by providing the radiologist with tomographic image slices through the breast. The goal of this research was to investigate the impact of varying x-ray spectra on image quality of breast tomosynthesis using an indirect CsI based detector. The ideal observer SNR was used as a figure-of-merit, under the assumption that the imaging system is linear and shift-invariant. Computations of the ideal observer SNR used a serial cascade model to predict signal and noise propagation through the detector, as well as a model of the lesion detection task in breast imaging. An indirect detector breast tomosynthesis prototype system was modeled which acquires 11 projection views by rotating the x-ray tube over a 50° angular range, with the breast and detector remaining stationary. Specific attention was focused on the impact of electronic noise for indirect detector breast tomosynthesis. Three different target/filters were studied including Mo/Mo, Mo/Rh, and W/Rh. Spectra were scaled to give a total of 2.4 mGy average glandular dose to the breast. It was observed that theW/Rh target/filter exhibited the best performance. In addition, electronic noise was observed to have a moderate effect on the SNR with more impact for thicker breasts and lower kVp settings.


Medical Physics | 2005

TU‐EE‐A3‐05: Scatter Radiation in Digital Tomosynthesis

Bob Liu; Stephen J. Glick; Xing Gong

Purpose: To investigate the characteristics of scattered radiation and its effects on image quality in digital breast tomosynthesis. Method and Materials: A GEANT 4 based Monte Carlo package was used to simulate a rotating target/detector tomosynthesis system. The compressed breast was modeled as a cubic block imbedded with 24 small cylinders of different radii and heights in the central layer. An 11cm air gap between the breast and detector was modeled. The incident photon energy was set to 20 keV to avoid the complexity of beam hardening effects. A primary image and a scatter image were generated for each projection. The gantry was rotated around the breast from −25 degree to 25 degree with a 5 degree increment. Reconstructions of the 3D breast were computed from the 11 projection images using primary x-rays only, and primary plus scattered x-rays. Results: The magnitude of scatter radiation does not change very much from one projection to another because the scatter volume does not change. However, the primary radiation detected can be significantly different in different projections due to different path length. As a result, scatter to primary ratio is very different for different projection. Even with the 11 cm air gap, scatter to primary ratio was observed to be as high as 0.4 for a 5 cm thick breast. 3D breast images reconstructed from projections with only primary x-rays showed higher contrast than those reconstructed from projections with both primary and scatter. Further evaluation is needed to determine if this reduced contrast can affect tumor detectability. Conclusion: Scatter to primary ratio changed significantly from one projection to another and was observed to be as high as 0.4 for a 5 cm thick breast. Tomosynthesis slices showed a moderate decrease in contrast due to scatter.


Medical Imaging 2005: Physics of Medical Imaging | 2005

A comparison of lesion detection accuracy using digital mammography and flat-panel CT breast imaging (Honorable Mention Poster Award)

Xing Gong; Aruna A. Vedula; Samta Thacker; Stephen J. Glick

Although conventional mammography is currently the best modality to detect early breast cancer, it is limited in that the recorded image represents the superposition of a 3D object onto a 2D plane. As an alternative, cone-beam CT breast imaging with a CsI based flat-panel imager (CTBI) has been proposed with the ability to provide 3D visualization of breast tissue. To investigate possible improvements in lesion detection accuracy using CTBI over digital mammography (DM), a computer simulation study was conducted using simulated lesions embedded into a structured 3D breast model. The computer simulation realistically modeled x-ray transport through a breast model, as well as the signal and noise propagation through the flat-panel imager. Polyenergetic x-ray spectra of W/Al 50 kVp for CTBI and Mo/Mo 28 kVp for DM were modeled. For the CTBI simulation, the intensity of the x-ray spectra for each projection view was determined so as to provide a total mean glandular dose (MGD) of 4 mGy, which is approximately equivalent to that given in a conventional two-view screening mammography study. Since only one DM view was investigated here, the intensity of the DM x-ray spectra was defined to give 2 mGy MGD. Irregular lesions were simulated by using a stochastic growth algorithm providing lesions with an effective diameter of 5 mm. Breast tissue was simulated by generating an ensemble of backgrounds with a power law spectrum. To evaluate lesion detection accuracy, a receiver operating characteristic (ROC) study was performed with 4 observers reading an ensemble of images for each case. The average area under the ROC curves (Az) was 0.94 for CTBI, and 0.81 for DM. Results indicate that a 5 mm lesion embedded in a structured breast phantom can be detected by CT breast imaging with statistically significant higher confidence than with digital mammography.


Medical Imaging 2005: Physics of Medical Imaging | 2005

The importance of modeling normal mammographic structure in optimizing flat-panel CT breast imaging systems

Stephen J. Glick; Samta Thacker; Xing Gong

In recent years, there has been interest in exploring the feasibility of CT breast imaging using flat-panel digital detectors in a truncated cone-beam geometry. Preliminary results are promising and it appears as if 3D tomographic imaging of the breast has great potential for reducing the masking effect of superimposed parenchymal structure typically observed with conventional mammography. In this study, a mathematical framework used for determining optimal design and acquisition parameters for such a CT breast imaging system is described. The ideal observer SNR is used as a figure-of-merit, under the assumptions that the imaging system is linear and shift-invariant. Computation of the ideal observer SNR used a parallel-cascade model to predict signal and noise propagation through the detector, as well as a realistic model of the lesion detection task in breast imaging. For all optimizations discussed here, the total mean glandular dose for a CT breast imaging study is constrained to be approximately equivalent to that of a two-view conventional mammography study. The framework presented is used to explore the affect of the specific task on the optimal exposure technique of flat-panel CT breast imaging. In particular, it is observed that modeling the normal mammographic structure in the projection images can sometimes impact the optimal kVp settings.


Medical Physics | 2006

SU-FF-I-29: Investigation of the Perspecta Display for 4D Visualization

Mark J. Rivard; Christopher S. Melhus; J Napoli; Xing Gong; J Chu

Purpose: Recent advancements in radiologic imaging(IGRT) have acquired 4D anatomic data permitting characterization of organ motion towards improving radiotherapy delivery. For radiationoncology patients, images illustrating temporal migration or tumor motion as a result of innate biological function can provide significant benefit towards improving target accuracy and minimizing healthy tissue dose. This study examines the utility of the Perspecta Spatial 3D system (Actuality Systems Inc) to display dynamic 3D data in comparison to flat panel 2D displays. Method and Materials: The AqSim (Philips Medical Systems) CT scanner was used to obtain scans of a patient with lungcancer, and entered into the Pinnacle3treatment planning system (Philips Medical Systems). A clearly delineated lungtumor was contoured in each pertinent CT slice. Ten scans (64 slices each) were obtained during the breathing cycle. Data were viewed side‐by‐side on a flat panel display and the Perspecta 3D system for comparison. Results: The Perspecta display permitted simultaneous visualization of ten CT scans at ∼ 1 Hz per dataset which was similar to the natural breathing rate during image acquisition. Optimal static beam orientation for dynamic target coverage and OAR avoidance was more easily accomplished on the Perspecta than on the 2D display. Conclusion: The 3D Perspecta display successfully depicted anatomic motion, clearly indicating tumor and OAR motion. In comparison to the 2D flat panel display, the Perspecta display permitted the radiationoncology team to readily visualize the temporal nature of lungtumor location for consideration during treatment planning. This application could play an important role in defining and displaying 4D patient data, which was previously relegated to predominantly 2D RTP systems. Furthermore, breath‐hold and coached breathing techniques may be quantitatively evaluated using this method. Conflict of Interest Statement: Actuality Systems Inc. provided the 3D display used in this study.

Collaboration


Dive into the Xing Gong's collaboration.

Top Co-Authors

Avatar

Stephen J. Glick

Food and Drug Administration

View shared research outputs
Top Co-Authors

Avatar

Aruna A. Vedula

University of Massachusetts Medical School

View shared research outputs
Top Co-Authors

Avatar

Samta Thacker

University of Massachusetts Medical School

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Clay Didier

University of Massachusetts Lowell

View shared research outputs
Top Co-Authors

Avatar

J Chu

Rush University Medical Center

View shared research outputs
Top Co-Authors

Avatar

J. Michael O'Connor

University of Massachusetts Medical School

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Mufeed Mahd

University of Massachusetts Lowell

View shared research outputs
Researchain Logo
Decentralizing Knowledge