André Ritter
University of Erlangen-Nuremberg
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by André Ritter.
Physics in Medicine and Biology | 2013
Thilo Michel; Jens Rieger; G. Anton; Florian Bayer; Matthias W. Beckmann; Jürgen Durst; Peter A. Fasching; Wilhelm Haas; Arndt Hartmann; Georg Pelzer; Marcus Radicke; Claudia Rauh; André Ritter; Peter Sievers; Rüdiger Schulz-Wendtland; Michael Uder; David L. Wachter; Thomas Weber; Evelyn Wenkel; Andrea Zang
We show that a distribution of micrometer-sized calcifications in the human breast which are not visible in clinical x-ray mammography at diagnostic dose levels can produce a significant dark-field signal in a grating-based x-ray phase-contrast imaging setup with a tungsten anode x-ray tube operated at 40 kVp. A breast specimen with invasive ductal carcinoma was investigated immediately after surgery by Talbot-Lau x-ray interferometry with a design energy of 25 keV. The sample contained two tumors which were visible in ultrasound and contrast-agent enhanced MRI but invisible in clinical x-ray mammography, in specimen radiography and in the attenuation images obtained with the Talbot-Lau interferometer. One of the tumors produced significant dark-field contrast with an exposure of 0.85 mGy air-kerma. Staining of histological slices revealed sparsely distributed grains of calcium phosphate with sizes varying between 1 and 40 μm in the region of this tumor. By combining the histological investigations with an x-ray wave-field simulation we demonstrate that a corresponding distribution of grains of calcium phosphate in the form of hydroxylapatite has the ability to produce a dark-field signal which would-to a substantial degree-explain the measured dark-field image. Thus we have found the appearance of new information (compared to attenuation and differential phase images) in the dark-field image. The second tumor in the same sample did not contain a significant fraction of these very fine calcification grains and was invisible in the dark-field image. We conclude that some tumors which are invisible in x-ray absorption mammography might be detected in the x-ray dark-field image at tolerable dose levels.
Medical Physics | 2011
Thomas Weber; Peter Bartl; Florian Bayer; Jürgen Durst; Wilhelm Haas; Thilo Michel; André Ritter; G. Anton
PURPOSE Grating-based x-ray phase-contrast imaging is a fast developing new modality not only for medical imaging, but as well for other fields such as material sciences. While these many possible applications arise, the knowledge of the noise behavior is essential. METHODS In this work, the authors used a least squares fitting algorithm to calculate the noise behavior of the three quantities absorption, differential phase, and dark-field image. Further, the calculated error formula of the differential phase image was verified by measurements. Therefore, a Talbot interferometer was setup, using a microfocus x-ray tube as source and a Timepix detector for photon counting. Additionally, simulations regarding this topic were performed. RESULTS It turned out that the variance of the reconstructed phase is only dependent of the total number of photons used to generate the phase image and the visibility of the experimental setup. These results could be evaluated in measurements as well as in simulations. Furthermore, the correlation between absorption and dark-field image was calculated. CONCLUSIONS These results provide the understanding of the noise characteristics of grating-based phase-contrast imaging and will help to improve image quality.
Zeitschrift Fur Medizinische Physik | 2013
G. Anton; Florian Bayer; Matthias W. Beckmann; Jürgen Durst; Peter A. Fasching; Wilhelm Haas; Arndt Hartmann; Thilo Michel; Georg Pelzer; Marcus Radicke; Claudia Rauh; Jens Rieger; André Ritter; Rüdiger Schulz-Wendtland; Michael Uder; David L. Wachter; Thomas Weber; Evelyn Wenkel; Lukas Wucherer
Mastectomy specimens were investigated using a Talbot-Lau X-ray imaging set-up. Significant structures in the darkfield were observed, which revealed considerably higher contrast than those observed in digital mammography. Comparison with the histomorphometric image proofs that the darkfield signal correlates with a tumor region containing small calcification grains of 3 to 30μm size.
Optics Express | 2013
Florian Bayer; Simon Zabler; Christian Brendel; Georg Pelzer; Jens Rieger; André Ritter; Thomas Weber; Thilo Michel; G. Anton
Over the recent years X-ray differential phase-contrast imaging was developed for the hard X-ray regime as produced from laboratory X-ray sources. The technique uses a grating-based Talbot-Lau interferometer and was shown to yield image contrast gain, which makes it very interesting to the fields of medical imaging and non-destructive testing, respectively. In addition to X-ray attenuation contrast, the differential phase-contrast and dark-field images provide different structural information about a specimen. For the dark-field even at length scales much smaller than the spatial resolution of the imaging system. Physical interpretation of the dark-field information as present in radiographic and tomographic (CT) images requires a detailed look onto the geometric orientation between specimen and the setup. During phase-stepping the drop in intensity modulation, due to local scattering effects within the specimen is reproduced in the dark-field signal. This signal shows strong dependencies on micro-porosity and micro-fibers if these are numerous enough in the object. Since a grating-interferometer using a common unidirectional line grating is sensitive to X-ray scattering in one plane only, the dark-field image is influenced by the fiber orientations with respect to the grating bars, which can be exploited to obtain anisotropic structural information. With this contribution, we attempt to extend existing models for 2D projections to 3D data by analyzing dark-field contrast tomography of anisotropically structured materials such as carbon fiber reinforced carbon (CFRC).
Optics Express | 2014
André Ritter; Peter Bartl; Florian Bayer; Karl C. Gödel; Wilhelm Haas; Thilo Michel; Georg Pelzer; Jens Rieger; Thomas Weber; Andrea Zang; Gisela Anton
A simulation framework for coherent X-ray imaging, based on scalar diffraction theory, is presented. It contains a core C++ library and an additional Python interface. A workflow is presented to include contributions of inelastic scattering obtained with Monte-Carlo methods. X-ray Talbot-Lau interferometry is the primary focus of the framework. Simulations are in agreement with measurements obtained with such an interferometer. Especially, the dark-field signal of densely packed PMMA microspheres is predicted. A realistic modeling of the microsphere distribution, which is necessary for correct results, is presented. The framework can be used for both setup design and optimization but also to test and improve reconstruction methods.
Optics Express | 2013
Georg Pelzer; Thomas Weber; G. Anton; Rafael Ballabriga; Florian Bayer; M. Campbell; Thomas Gabor; Wilhelm Haas; Florian Horn; X. Llopart; Norbert Michel; Uwe Mollenbauer; Jens Rieger; André Ritter; Ina Ritter; Peter Sievers; Stefan Wölfel; Winnie Wong; Andrea Zang; Thilo Michel
We have carried out grating-based x-ray differential phase-contrast measurements with a hybrid pixel detector in 16 energy channels simultaneously. A method for combining the energy resolved phase-contrast images based on energy weighting is presented. An improvement in contrast-to-noise ratio by 58.2% with respect to an emulated integrating detector could be observed in the final image. The same image quality could thus be achieved with this detector and with energy weighting at 60.0% reduced dose compared to an integrating detector. The benefit of the method depends on the object, spectrum, interferometer design and the detector efficiency.
Proceedings of SPIE | 2011
Wilhelm Haas; Martin Bech; Peter Bartl; Florian Bayer; André Ritter; Thomas Weber; Georg Pelzer; Marian Willner; Klaus Achterhold; Jürgen Durst; Thilo Michel; Marcus Prümmer; Franz Pfeiffer; G. Anton; Joachim Hornegger
Phase-contrast imaging approaches suffer from a severe problem which is known in Magnetic Resonance Imaging (MRI) and Synthetic Aperture Radar (SAR) as phase-wrapping. This work focuses on an unwrapping solution for the grating based phase-contrast interferometer with X-rays. The approach delivers three types of information about the x-rayed object - the absorption, differential phase-contrast and dark-field information whereas the observed differential phase values are physically limited to the interval (-π, π]; values higher or lower than the interval borders are mapped (wrapped) back into it. In contrast to existing phase-unwrapping algorithms for MRI and SAR the presented algorithm uses the absorption image as additional information to identify and correct phase-wrapped values. The idea of the unwrapping algorithm is based on the observation that at locations with phase-wrapped values the contrast in the absorption image is high and the behavior of the gradient is similar to the real (unwrapped) phase values. This can be expressed as a cost function which has to be minimized by an integer optimizer. Applied on simulated and real datasets showed that 95.6% of phase-wraps were correctly unwrapped. Based on the results we conclude that it is possible to use the absorption information in order to identify and correct phase-wrapped values.
Optics Express | 2013
Thomas Weber; Georg Pelzer; Florian Bayer; Florian Horn; Jens Rieger; André Ritter; Andrea Zang; Jürgen Durst; G. Anton; Thilo Michel
A novel information retrieval algorithm for X-ray grating-based phase-contrast imaging based on the deconvolution of the object and the reference phase stepping curve (PSC) as proposed by Modregger et al. was investigated in this paper. We applied the method for the first time on data obtained with a polychromatic spectrum and compared the results to those, received by applying the commonly used method, based on a Fourier analysis. We confirmed the expectation, that both methods deliver the same results for the absorption and the differential phase image. For the darkfield image, a mean contrast-to-noise ratio (CNR) increase by a factor of 1.17 using the new method was found. Furthermore, the dose saving potential was estimated for the deconvolution method experimentally. It is found, that for the conventional method a dose which is higher by a factor of 1.66 is needed to obtain a similar CNR value compared to the novel method. A further analysis of the data revealed, that the improvement in CNR and dose efficiency is due to the superior background noise properties of the deconvolution method, but at the cost of comparability between measurements at different applied dose values, as the mean value becomes dependent on the photon statistics used.
Journal of Instrumentation | 2012
Thomas Weber; Florian Bayer; Wilhelm Haas; Georg Pelzer; Jens Rieger; André Ritter; Lukas Wucherer; Jürgen Durst; Thilo Michel; G. Anton
The visibility is a crucial quality parameter in grating-based X-ray phase-contrast imaging as it directly influences the noise of the obtained phase signal. In order to evaluate its behaviour with respect to energy, we used the photon counting semiconductor detector Timepix to obtain a series of measurements at different energy thresholds. These energy-dependent measurements of the visibility response of a Talbot-Lau Interferometer for X-ray phase-contrast imaging are presented in this paper. Furthermore, we report on the results of our effort to model this behaviour in our simulation, where a very good agreement between measurement and simulation could be achieved for two available setups. With this tested workflow we are now able to predict the expected visibility response of a specific setup. This capability is an important prerequisite for the optimisation of an X-ray Talbot-Lau Interferometer. It is predicted that the visibility of our current laboratory setup could be doubled by choosing an appropriate X-ray spectrum or filter material.
Proceedings of SPIE | 2010
Peter Bartl; Juergen Durst; Wilhelm Haas; Eckhard Hempel; Thilo Michel; André Ritter; Thomas Weber; G. Anton
We present a simulation framework for X-ray phase-contrast computed tomography imaging (PCTI) inheriting the wave- as well as the particle-behavior of photons. The developed tool includes the modeling of a partially coherent X-ray source, the propagation of the X-ray photons through samples, and the interfering properties of photons. Hence, the simulation is capable of physically modeling a grating-based interferometric imaging system reported in e.g. Pfeiffer et al.5 The information gained comprises the three potentially measurable images, which are the absorption image, the phase image, and the darkfield image. Results on such a setup concerning spatial and temporal coherence will be shown. Samples consisting of elements and structures similar to biological tissue were implemented to demonstrate the applicability on medical imaging. For the purpose of CT-imaging a head-like phantom was simulated and the results show the advantage of PCTI for thick biological objects. The simulation was developed with a modular concept so that the influences of each imaging component can be considered seperately. Thus the grating based interferometry for X-ray phase-contrast imaging can be optimized towards dedicated medical applications using this simulation-tool.