Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Andrea Zang is active.

Publication


Featured researches published by Andrea Zang.


Physics in Medicine and Biology | 2013

On a dark-field signal generated by micrometer-sized calcifications in phase-contrast mammography

Thilo Michel; Jens Rieger; G. Anton; Florian Bayer; Matthias W. Beckmann; Jürgen Durst; Peter A. Fasching; Wilhelm Haas; Arndt Hartmann; Georg Pelzer; Marcus Radicke; Claudia Rauh; André Ritter; Peter Sievers; Rüdiger Schulz-Wendtland; Michael Uder; David L. Wachter; Thomas Weber; Evelyn Wenkel; Andrea Zang

We show that a distribution of micrometer-sized calcifications in the human breast which are not visible in clinical x-ray mammography at diagnostic dose levels can produce a significant dark-field signal in a grating-based x-ray phase-contrast imaging setup with a tungsten anode x-ray tube operated at 40 kVp. A breast specimen with invasive ductal carcinoma was investigated immediately after surgery by Talbot-Lau x-ray interferometry with a design energy of 25 keV. The sample contained two tumors which were visible in ultrasound and contrast-agent enhanced MRI but invisible in clinical x-ray mammography, in specimen radiography and in the attenuation images obtained with the Talbot-Lau interferometer. One of the tumors produced significant dark-field contrast with an exposure of 0.85 mGy air-kerma. Staining of histological slices revealed sparsely distributed grains of calcium phosphate with sizes varying between 1 and 40 μm in the region of this tumor. By combining the histological investigations with an x-ray wave-field simulation we demonstrate that a corresponding distribution of grains of calcium phosphate in the form of hydroxylapatite has the ability to produce a dark-field signal which would-to a substantial degree-explain the measured dark-field image. Thus we have found the appearance of new information (compared to attenuation and differential phase images) in the dark-field image. The second tumor in the same sample did not contain a significant fraction of these very fine calcification grains and was invisible in the dark-field image. We conclude that some tumors which are invisible in x-ray absorption mammography might be detected in the x-ray dark-field image at tolerable dose levels.


Optics Express | 2014

Simulation framework for coherent and incoherent X-ray imaging and its application in Talbot-Lau dark-field imaging.

André Ritter; Peter Bartl; Florian Bayer; Karl C. Gödel; Wilhelm Haas; Thilo Michel; Georg Pelzer; Jens Rieger; Thomas Weber; Andrea Zang; Gisela Anton

A simulation framework for coherent X-ray imaging, based on scalar diffraction theory, is presented. It contains a core C++ library and an additional Python interface. A workflow is presented to include contributions of inelastic scattering obtained with Monte-Carlo methods. X-ray Talbot-Lau interferometry is the primary focus of the framework. Simulations are in agreement with measurements obtained with such an interferometer. Especially, the dark-field signal of densely packed PMMA microspheres is predicted. A realistic modeling of the microsphere distribution, which is necessary for correct results, is presented. The framework can be used for both setup design and optimization but also to test and improve reconstruction methods.


Optics Express | 2013

Grating-based x-ray phase-contrast imaging with a multi energy-channel photon-counting pixel detector.

Georg Pelzer; Thomas Weber; G. Anton; Rafael Ballabriga; Florian Bayer; M. Campbell; Thomas Gabor; Wilhelm Haas; Florian Horn; X. Llopart; Norbert Michel; Uwe Mollenbauer; Jens Rieger; André Ritter; Ina Ritter; Peter Sievers; Stefan Wölfel; Winnie Wong; Andrea Zang; Thilo Michel

We have carried out grating-based x-ray differential phase-contrast measurements with a hybrid pixel detector in 16 energy channels simultaneously. A method for combining the energy resolved phase-contrast images based on energy weighting is presented. An improvement in contrast-to-noise ratio by 58.2% with respect to an emulated integrating detector could be observed in the final image. The same image quality could thus be achieved with this detector and with energy weighting at 60.0% reduced dose compared to an integrating detector. The benefit of the method depends on the object, spectrum, interferometer design and the detector efficiency.


Optics Express | 2013

Increasing the darkfield contrast-to-noise ratio using a deconvolution-based information retrieval algorithm in X-ray grating-based phase-contrast imaging

Thomas Weber; Georg Pelzer; Florian Bayer; Florian Horn; Jens Rieger; André Ritter; Andrea Zang; Jürgen Durst; G. Anton; Thilo Michel

A novel information retrieval algorithm for X-ray grating-based phase-contrast imaging based on the deconvolution of the object and the reference phase stepping curve (PSC) as proposed by Modregger et al. was investigated in this paper. We applied the method for the first time on data obtained with a polychromatic spectrum and compared the results to those, received by applying the commonly used method, based on a Fourier analysis. We confirmed the expectation, that both methods deliver the same results for the absorption and the differential phase image. For the darkfield image, a mean contrast-to-noise ratio (CNR) increase by a factor of 1.17 using the new method was found. Furthermore, the dose saving potential was estimated for the deconvolution method experimentally. It is found, that for the conventional method a dose which is higher by a factor of 1.66 is needed to obtain a similar CNR value compared to the novel method. A further analysis of the data revealed, that the improvement in CNR and dose efficiency is due to the superior background noise properties of the deconvolution method, but at the cost of comparability between measurements at different applied dose values, as the mean value becomes dependent on the photon statistics used.


Journal of Instrumentation | 2015

The Dosepix detector—an energy-resolving photon-counting pixel detector for spectrometric measurements

Andrea Zang; G. Anton; Rafael Ballabriga; F. Bisello; M. Campbell; J.C. Celi; A. Fauler; M. Fiederle; M. Jensch; N. Kochanski; X. Llopart; N. Michel; U. Mollenhauer; Ina Ritter; F. Tennert; Stefan Wölfel; Winnie Wong; Thilo Michel

The Dosepix detector is a hybrid photon-counting pixel detector based on ideas of the Medipix and Timepix detector family. 1 mm thick cadmium telluride and 300 µm thick silicon were used as sensor material. The pixel matrix of the Dosepix consists of 16 x 16 square pixels with 12 rows of (200 µm) 2 and 4 rows of (55 µm) 2 sensitive area for the silicon sensor layer and 16 rows of pixels with 220 µm pixel pitch for CdTe. Besides digital energy integration and photon- counting mode, a novel concept of energy binning is included in the pixel electronics, allowing energy-resolved measurements in 16 energy bins within one acquisition. The possibilities of this detector concept range from applications in personal dosimetry and energy-resolved imaging to quality assurance of medical X-ray sources by analysis of the emitted


Proceedings of SPIE | 2013

Energy-resolved interferometric x-ray imaging

Georg Pelzer; Florian Bayer; Karl C. Gödel; Wilhelm Haas; Florian Horn; Jens Rieger; André Ritter; Peter Sievers; Thomas Weber; Andrea Zang; Jürgen Durst; Thilo Michel; G. Anton

Interferometric X-ray imaging becomes more and more attractive for applications such as medical imaging or non-destructive testing, where a compact setup is needed. Therefore a so-called Talbot-Lau interferometer in combination with a conventional X-ray tube is used. Thereby, three different kinds of images can be obtained. An attenuation image like in conventional X-ray imaging, an image of the differential phase-shifts caused by the object and the so-called dark-field image. The dark-field image shows information about the objects granularity even in sub-pixel dimensions what especially seems very promising for applications like mammography. With respect to optimizing the output of interferometric X-ray imaging in any application, it is inevitable to know the energy response of the interferometer as well as the energy dependence of the interactions of X- rays with matter. In this contribution, simulations and measurements using a Medipix 2 and a Timepix detector are presented.


Proceedings of SPIE | 2014

Characterization of a hybrid energy-resolving photon-counting detector

Andrea Zang; Georg Pelzer; G. Anton; R. Ballabriga Sune; F. Bisello; M. Campbell; A. Fauler; M. Fiederle; X. Llopart Cudie; Ina Ritter; F. Tennert; Stefan Wölfel; Winnie Wong; Thilo Michel

Photon-counting detectors in medical x-ray imaging provide a higher dose efficiency than integrating detectors. Even further possibilities for imaging applications arise, if the energy of each photon counted is measured, as for example K-edge-imaging or optimizing image quality by applying energy weighting factors. In this contribution, we show results of the characterization of the Dosepix detector. This hybrid photon- counting pixel detector allows energy resolved measurements with a novel concept of energy binning included in the pixel electronics. Based on ideas of the Medipix detector family, it provides three different modes of operation: An integration mode, a photon-counting mode, and an energy-binning mode. In energy-binning mode, it is possible to set 16 energy thresholds in each pixel individually to derive a binned energy spectrum in every pixel in one acquisition. The hybrid setup allows using different sensor materials. For the measurements 300 μm Si and 1 mm CdTe were used. The detector matrix consists of 16 x 16 square pixels for CdTe (16 x 12 for Si) with a pixel pitch of 220 μm. The Dosepix was originally intended for applications in the field of radiation measurement. Therefore it is not optimized towards medical imaging. The detector concept itself still promises potential as an imaging detector. We present spectra measured in one single pixel as well as in the whole pixel matrix in energy-binning mode with a conventional x-ray tube. In addition, results concerning the count rate linearity for the different sensor materials are shown as well as measurements regarding energy resolution.


Proceedings of SPIE | 2013

Grating-based dark-field breast imaging

Jens Rieger; Florian Bayer; Jürgen Durst; Karl C. Gödel; Wilhelm Haas; Florian Horn; Thilo Michel; Georg Pelzer; André Ritter; Thomas Weber; Andrea Zang; G. Anton

Grating-based X-ray phase-contrast imaging (XPCI) is a promising modality to increase soft-tissue contrast in medical imaging and especially in the case of mammography. Several groups worldwide have performed investigations on grating-based Talbot-Lau X-ray imaging of breast tissue, but in most cases focussed on the soft tissue contrast enhancement of the differential phase image. In this contribution, we present promising measurements with a Talbot-Lau interferometer of several mastectomy breast tissue samples especially focussed on the sensitivity of the dark-field signal of microcalcifications and with a comparable dose value to conventional mammography. We can present a contrast improvement for calcifications in surrounding breast tissue for the dark-field image by a factor of 10 related to the attenuation image.


Journal of Instrumentation | 2015

Improvement of the energy resolution of pixelated CdTe detectors for applications in 0νββ searches

T. Gleixner; G. Anton; M. Filipenko; P. Seller; Matthew C. Veale; Matthew D. Wilson; Andrea Zang; Thilo Michel

Experiments trying to detect 0νββ are very challenging. Their requirements include a good energy resolution and a good detection efficiency. With current fine pixelated CdTe detectors there is a trade off between the energy resolution and the detection efficiency, which limits their performance. It will be shown with simulations that this problem can be mostly negated by analysing the cathode signal which increases the optimal sensor thickness. We will compare different types of fine pixelated CdTe detectors (Timepix, Dosepix, HEXITEC) from this point of view.


Proceedings of SPIE | 2014

Energy weighting in grating-based X-ray phase-contrast imaging

Georg Pelzer; Thomas Weber; G. Anton; Rafael Ballabriga Sune; Florian Bayer; M. Campbell; Wilhelm Haas; Florian Horn; Xavi Llopart Cudie; Norbert Michel; Uwe Mollenbauer; Jens Rieger; André Ritter; Ina Ritter; Stefan Wölfel; Winnie Wong; Andrea Zang; Thilo Michel

With energy-resolving photon-counting detectors in grating-based x-ray phase-contrast imaging it is possible to reduce the dose needed and optimize the imaging chain towards better performance. The advantage of photon- counting detector’s linear energy response and absence of electronic noise in attenuation based imaging is known. The access to the energy information of the photons counted provides even further potential for optimization by applying energy weighting factors. We have evaluated energy weighting for grating-based phase-contrast imaging. Measurements with the hybrid photon-counting detector Dosepix were performed. The concept of energy binning implemented in the pixel electronics allows individual storing of the energy information of the incoming photons in 16 energy bins for each pixel. With this technique the full spectral information can be obtained pixel wise from one single acquisition. On the differential phase-contrast data taken, we applied different types of energy weighting factors. The results presented in this contribution demonstrate the advantages of energy-resolved photon-counting in differential phase-contrast imaging. Using a x-ray spectrum centred significantly above the interferometers design energy leads to poor image quality. But with the proposed method and detector the quality was enhanced by 2.8 times in signal-to-noise ratio squared. As this is proportional to dose, energy- resolved photon-counting might be valuable especially for medical applications.

Collaboration


Dive into the Andrea Zang's collaboration.

Top Co-Authors

Avatar

Thilo Michel

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

G. Anton

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Georg Pelzer

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

André Ritter

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Florian Bayer

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Jens Rieger

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Thomas Weber

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Florian Horn

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Wilhelm Haas

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Jürgen Durst

University of Erlangen-Nuremberg

View shared research outputs
Researchain Logo
Decentralizing Knowledge