Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Christina E. Agnew is active.

Publication


Featured researches published by Christina E. Agnew.


Medical Physics | 2013

Octavius 4D characterization for flattened and flattening filter free rotational deliveries

Conor K. McGarry; Barry F. O'Connell; Mark W. D. Grattan; Christina E. Agnew; Denise M. Irvine; A.R. Hounsell

PURPOSE In this study the Octavius detector 729 ionization chamber (IC) array with the Octavius 4D phantom was characterized for flattening filter (FF) and flattening filter free (FFF) static and rotational beams. The device was assessed for verification with FF and FFF RapidArc treatment plans. METHODS The response of the detectors to field size, dose linearity, and dose rate were assessed for 6 MV FF beams and also 6 and 10 MV FFF beams. Dosimetric and mechanical accuracy of the detector array within the Octavius 4D rotational phantom was evaluated against measurements made using semiflex and pinpoint ionization chambers, and radiochromic film. Verification FF and FFF RapidArc plans were assessed using a gamma function with 3%∕3 mm tolerances and 2%∕2 mm tolerances and further analysis of these plans was undertaken using film and a second detector array with higher spatial resolution. RESULTS A warm-up dose of >6 Gy was required for detector stability. Dose-rate measurements were stable across a range from 0.26 to 15 Gy∕min and dose response was linear, although the device overestimated small doses compared with pinpoint ionization chamber measurements. Output factors agreed with ionization chamber measurements to within 0.6% for square fields of side between 3 and 25 cm and within 1.2% for 2 × 2 cm(2) fields. The Octavius 4D phantom was found to be consistent with measurements made with radiochromic film, where the gantry angle was found to be within 0.4° of that expected during rotational deliveries. RapidArc FF and FFF beams were found to have an accuracy of >97.9% and >90% of pixels passing 3%∕3 mm and 2%∕2 mm, respectively. Detector spatial resolution was observed to be a factor in determining the accurate delivery of each plan, particularly at steep dose gradients. This was confirmed using data from a second detector array with higher spatial resolution and with radiochromic film. CONCLUSIONS The Octavius 4D phantom with associated Octavius detector 729 ionization chamber array is a dosimetrically and mechanically stable device for pretreatment verification of FF and FFF RapidArc treatments. Further improvements may be possible through use of a detector array with higher spatial resolution (detector size and∕or detector spacing).


Journal of Applied Clinical Medical Physics | 2014

Correlation of phantom-based and log file patient-specific QA with complexity scores for VMAT

Christina E. Agnew; Denise M. Irvine; Conor K. McGarry

The motivation for this study was to reduce physics workload relating to patient‐specific quality assurance (QA). VMAT plan delivery accuracy was determined from analysis of pre‐ and on‐treatment trajectory log files and phantom‐based ionization chamber array measurements. The correlation in this combination of measurements for patient‐specific QA was investigated. The relationship between delivery errors and plan complexity was investigated as a potential method to further reduce patient‐specific QA workload. Thirty VMAT plans from three treatment sites — prostate only, prostate and pelvic node (PPN), and head and neck (H&N) — were retrospectively analyzed in this work. The 2D fluence delivery reconstructed from pretreatment and on‐treatment trajectory log files was compared with the planned fluence using gamma analysis. Pretreatment dose delivery verification was also carried out using gamma analysis of ionization chamber array measurements compared with calculated doses. Pearson correlations were used to explore any relationship between trajectory log file (pretreatment and on‐treatment) and ionization chamber array gamma results (pretreatment). Plan complexity was assessed using the MU/ arc and the modulation complexity score (MCS), with Pearson correlations used to examine any relationships between complexity metrics and plan delivery accuracy. Trajectory log files were also used to further explore the accuracy of MLC and gantry positions. Pretreatment 1%/1 mm gamma passing rates for trajectory log file analysis were 99.1% (98.7%–99.2%), 99.3% (99.1%–99.5%), and 98.4% (97.3%–98.8%) (median (IQR)) for prostate, PPN, and H&N, respectively, and were significantly correlated to on‐treatment trajectory log file gamma results (R=0.989,p<0.001). Pretreatment ionization chamber array (2%/2 mm) gamma results were also significantly correlated with on‐treatment trajectory log file gamma results (R=0.623,p<0.001). Furthermore, all gamma results displayed a significant correlation with MCS (R>0.57,p<0.001), but not with MU/arc. Average MLC position and gantry angle errors were 0.001±0.002mm and 0.025°±0.008° over all treatment sites and were not found to affect delivery accuracy. However, variability in MLC speed was found to be directly related to MLC position accuracy. The accuracy of VMAT plan delivery assessed using pretreatment trajectory log file fluence delivery and ionization chamber array measurements were strongly correlated with on‐treatment trajectory log file fluence delivery. The strong correlation between trajectory log file and phantom‐based gamma results demonstrates potential to reduce our current patient‐specific QA. Additionally, insight into MLC and gantry position accuracy through trajectory log file analysis and the strong correlation between gamma analysis results and the MCS could also provide further methodologies to both optimize the VMAT planning and QA process. PACS number: 87.53.Bn, 87.55.Kh, 87.55.Qr


Practical radiation oncology | 2014

Improvement in clinical step and shoot intensity modulated radiation therapy delivery accuracy on an integrated linear accelerator control system.

Christina E. Agnew; Denise M. Irvine; A.R. Hounsell; Conor K. McGarry

PURPOSE The dose delivery accuracy of 30 clinical step and shoot intensity modulated radiation therapy plans was investigated using the single integrated multileaf collimator controller of the Varian Truebeam linear accelerator (linac) (Varian Medical Systems, Palo Alto, CA) and compared with the dose delivery accuracy on a previous generation Varian 2100CD C-Series linac. METHODS AND MATERIALS Ten prostate, 10 prostate and pelvic node, and 10 head-and-neck cases were investigated in this study. Dose delivery accuracy on each linac was assessed using Farmer ionization chamber point dose measurements, 2-dimensional planar ionization chamber array measurements, and the corresponding Varian dynamic log files. Absolute point dose measurements, fluence delivery accuracy, leaf position accuracy, and the overshoot effect were assessed for each plan. RESULTS Absolute point dose delivery accuracy increased by 1.5% on the Truebeam compared with the 2100CD linac. No improvement in fluence delivery accuracy between the linacs, at a gamma criterion of 3%/3 mm was measured using the 2-dimensional ionization chamber array, with median (interquartile range) gamma passing rates of 98.99% (97.70%-99.72%) and 99.28% (98.26%-99.75%) for the Truebeam and 2100CD linacs, respectively. Varian log files also showed no improvement in fluence delivery between the linacs at 3%/3 mm, with median gamma passing rates of 99.97% (99.93%-99.99%) and 99.98% (99.94%-100%) for the Truebeam and 2100CD linacs, respectively. However, log files revealed improved leaf position accuracy and fluence delivery at 1%/1 mm criterion on the Truebeam (99.87%; 99.78%-99.94%) compared with the 2100CD linac (97.87%; 91.93%-99.49%). The overshoot effect, characterized on the 2100CD linac, was not observed on the Truebeam. CONCLUSIONS The integrated multileaf collimator controller on the Varian Truebeam improves clinical treatment delivery accuracy of step and shoot intensity modulated radiation therapy fields compared with delivery on a Varian C-series linac.


IEEE Transactions on Biomedical Engineering | 2011

Comparison of RootMUSIC and Discrete Wavelet Transform Analysis of Doppler Ultrasound Blood Flow Waveforms to Detect Microvascular Abnormalities in Type I Diabetes

Christina E. Agnew; Aaron McCann; Christopher J. Lockhart; Paul K. Hamilton; Gary E. McVeigh; R.C. McGivern

The earliest signs of cardiovascular disease occur in microcirculations. Changes to mechanical and structural properties of these small resistive vessels alter the impedance to flow, subsequent reflected waves, and consequently, flow waveform morphology. In this paper, we compare two frequency analysis techniques: 1) rootMUSIC and 2) the discrete wavelet transform (DWT) to extract features of flow velocity waveform morphology captured using Doppler ultrasound from the ophthalmic artery (OA) in 30 controls and 38 age and sex matched Type I diabetics. Conventional techniques for characterizing Doppler velocity waveforms, such as mean velocity, resistive index, and pulsatility index, revealed no significant differences between the groups. However, rootMUSIC and the DWT provided highly correlated results with the spectral con tent in bands 2-7 (30-0.8 Hz) significantly elevated in the diabetic group (p <; 0.05). The spectral distinction between the groups may be attributable to manifestations of underlying pathophysiological processes in vascular impedance and consequent wave reflections, with bands 5 and 7 related to age. Spectral descriptors of OA blood velocity waveforms are better indicators of preclinical microvascular abnormalities in Type I diabetes than conventional measures. Although highly correlated DWT proved slightly more discriminatory than rootMUSIC and has the advantage of extending to subheart rate frequencies, which may be of interest.


QJM: An International Journal of Medicine | 2011

Flow-mediated dilatation of the brachial artery is a poorly reproducible indicator of microvascular function in Type I diabetes mellitus

P. K. Hamilton; Christopher J. Lockhart; Aaron McCann; Christina E. Agnew; Mark Harbinson; Vivienne McClenaghan; C. Bleakley; R.C. McGivern; Gary E. McVeigh

BACKGROUND Flow-mediated dilatation (FMD) of the brachial artery is commonly measured as a surrogate marker of endothelial function. Its measurement is, however, technically demanding and reports regarding its reproducibility have not always been favourable. AIM Two Type I diabetes and control group comparator studies were conducted to assess the reproducibility of FMD and to analyse blood flow data normally discarded during FMD measurement. DESIGN The studies were sequential and differed only with regard to operator and ultrasound machine. Seventy-two subjects with diabetes and 71 controls were studied in total. METHODS Subjects had FMD measured conventionally. Blood velocity waveforms were averaged over 10 pulses post forearm ischaemia and their component frequencies analysed using the wavelet transform, a mathematical tool for waveform analysis. The component frequencies were grouped into 11 bands to facilitate analysis. RESULTS Subjects were well-matched between studies. In Study 1, FMD was significantly impaired in subjects with Type I diabetes vs. controls (median 4.35%, interquartile range 3.10-4.80 vs. 6.50, 4.79-9.42, P < 0.001). No differences were detected between groups in Study 2, however. However, analysis of blood velocity waveforms yielded significant differences between groups in two frequency bands in each study. CONCLUSION This report highlights concerns over the reproducibility of FMD measures. Further work is required to fully elucidate the role of analysing velocity waveforms after forearm ischaemia.


Radiotherapy and Oncology | 2016

A tool to include gamma analysis software into a quality assurance program

Christina E. Agnew; Conor K. McGarry

PURPOSE To provide a tool to enable gamma analysis software algorithms to be included in a quality assurance (QA) program. METHODS Four image sets were created comprising two geometric images to independently test the distance to agreement (DTA) and dose difference (DD) elements of the gamma algorithm, a clinical step and shoot IMRT field and a clinical VMAT arc. The images were analysed using global and local gamma analysis with 2 in-house and 8 commercially available software encompassing 15 software versions. The effect of image resolution on gamma pass rates was also investigated. RESULTS All but one software accurately calculated the gamma passing rate for the geometric images. Variation in global gamma passing rates of 1% at 3%/3mm and over 2% at 1%/1mm was measured between software and software versions with analysis of appropriately sampled images. CONCLUSION This study provides a suite of test images and the gamma pass rates achieved for a selection of commercially available software. This image suite will enable validation of gamma analysis software within a QA program and provide a frame of reference by which to compare results reported in the literature from various manufacturers and software versions.


Medical Engineering & Physics | 2009

Root-MUSIC analysis of nitric oxide-mediated changes in ophthalmic artery blood flow velocity waveforms

Christina E. Agnew; Derrick J. Rea; Aaron McCann; Christopher J. Lockhart; Paul K. Hamilton; Cathy E. Quinn; Gary E. McVeigh; R.C. McGivern

Clinical and experimental studies indicate that structural and functional changes in the microvasculature can predate or accompany risk factors for cardiovascular disease at the earliest stages in the disease process. In the current work, both simulated and actual Doppler ultrasound maximum blood velocity waveform envelopes recorded from the ophthalmic artery were analysed using a root-MUSIC and least squares fitting approach to determine amplitude frequency spectra. Both amplitude and frequency components of noise contaminated simulated waveforms were reliably determined indicating the robustness of the technique. The technique was then used to compare the spectral content of the ophthalmic artery blood velocity waveforms of normal controls in three test states: at baseline, following administration of GTN, a nitric oxide donor, and following administration of L-Name, a nitric oxide inhibitor. Principal components derived from root-MUSIC analysis discriminated between waveforms in baseline and non-baseline test states (p<0.00001) and between GTN and non-GTN test states (p=0.0002).


British Journal of Radiology | 2016

The use of log file analysis within VMAT audits

Conor K. McGarry; Christina E. Agnew; M. Hussein; Yatman Tsang; A.R. Hounsell; Catharine H. Clark

OBJECTIVE This work investigated the delivery accuracy of different Varian linear accelerator models using log file-derived multileaf collimator (MLC) root mean square (RMS) values. METHODS Seven centres independently created a plan on the same virtual phantom using their own planning system, and the log files were analyzed following delivery of the plan in each centre to assess MLC positioning accuracy. A single standard plan was also delivered by the seven centres to remove variations in complexity, and the log files were analyzed for Varian TrueBeams and Clinacs (2300IX or 2100CD models). RESULTS Varian TrueBeam accelerators had better MLC positioning accuracy (<1.0 mm) than the 2300IX (<2.5 mm) following delivery of the plans created by each centre and also the standard plan. In one case, log files provided evidence that reduced delivery accuracy was not associated with the linear accelerator model but was due to planning issues. CONCLUSION Log files are useful in identifying differences between linear accelerator models and isolate errors during end-to-end testing in volumetric-modulated arc therapy (VMAT) audits. Log file analysis can rapidly eliminate the machine delivery from the process and divert attention with confidence to other aspects. ADVANCES IN KNOWLEDGE Log file evaluation was shown to be an effective method to rapidly verify satisfactory treatment delivery when a dosimetric evaluation fails during end-to-end dosimetry audits. MLC RMS values for Varian TrueBeams were shown to be much smaller than those for Varian Clinacs for VMAT deliveries.


Microcirculation | 2011

Ocular Blood Flow Analysis Detects Microvascular Abnormalities in Impaired Glucose Tolerance

Catherine E. Quinn; Paul K. Hamilton; Aaron McCann; Christina E. Agnew; Auleen Millar; Christopher J. Lockhart; Mark Harbinson; Gary E. McVeigh

Please cite this paper as: Quinn, Hamilton, McCann, Agnew, Millar, Lockhart, Harbinson and McVeigh (2011). Ocular Blood Flow Analysis Detects Microvascular Abnormalities in Impaired Glucose Tolerance. Microcirculation 18(7), 532–540.


Physics in Medicine and Biology | 2016

Time-resolved dosimetric verification of respiratory-gated radiotherapy exposures using a high-resolution 2D ionisation chamber array

Raymond King; Christina E. Agnew; B.F. O'Connell; Kevin Prise; A.R. Hounsell; Conor K. McGarry

The aim of this work was to track and verify the delivery of respiratory-gated irradiations, performed with three versions of TrueBeam linac, using a novel phantom arrangement that combined the OCTAVIUS(®) SRS 1000 array with a moving platform. The platform was programmed to generate sinusoidal motion of the array. This motion was tracked using the real-time position management (RPM) system and four amplitude gating options were employed to interrupt MV beam delivery when the platform was not located within set limits. Time-resolved spatial information extracted from analysis of x-ray fluences measured by the array was compared to the programmed motion of the platform and to the trace recorded by the RPM system during the delivery of the x-ray field. Temporal data recorded by the phantom and the RPM system were validated against trajectory log files, recorded by the linac during the irradiation, as well as oscilloscope waveforms recorded from the linac target signal. Gamma analysis was employed to compare time-integrated 2D x-ray dose fluences with theoretical fluences derived from the probability density function for each of the gating settings applied, where gamma criteria of 2%/2 mm, 1%/1 mm and 0.5%/0.5 mm were used to evaluate the limitations of the RPM system. Excellent agreement was observed in the analysis of spatial information extracted from the SRS 1000 array measurements. Comparisons of the average platform position with the expected position indicated absolute deviations of  <0.5 mm for all four gating settings. Differences were observed when comparing time-resolved beam-on data stored in the RPM files and trajectory logs to the true target signal waveforms. Trajectory log files underestimated the cycle time between consecutive beam-on windows by 10.0  ±  0.8 ms. All measured fluences achieved 100% pass-rates using gamma criteria of 2%/2 mm and 50% of the fluences achieved pass-rates  >90% when criteria of 0.5%/0.5 mm were used. Results using this novel phantom arrangement indicate that the RPM system is capable of accurately gating x-ray exposure during the delivery of a fixed-field treatment beam.

Collaboration


Dive into the Christina E. Agnew's collaboration.

Top Co-Authors

Avatar

Conor K. McGarry

Belfast Health and Social Care Trust

View shared research outputs
Top Co-Authors

Avatar

A.R. Hounsell

Belfast Health and Social Care Trust

View shared research outputs
Top Co-Authors

Avatar

Aaron McCann

Belfast Health and Social Care Trust

View shared research outputs
Top Co-Authors

Avatar

Gary E. McVeigh

Queen's University Belfast

View shared research outputs
Top Co-Authors

Avatar

Paul K. Hamilton

Queen's University Belfast

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Denise M. Irvine

Belfast Health and Social Care Trust

View shared research outputs
Top Co-Authors

Avatar

Mark Harbinson

Queen's University Belfast

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

R.C. McGivern

Belfast Health and Social Care Trust

View shared research outputs
Researchain Logo
Decentralizing Knowledge