Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where P Summers is active.

Publication


Featured researches published by P Summers.


Medical Physics | 2013

Quality assurance of proton beams using a multilayer ionization chamber system

S Dhanesar; Narayan Sahoo; M Kerr; M. Brad Taylor; P Summers; X. Ronald Zhu; F Poenisch; M Gillin

PURPOSE The measurement of percentage depth-dose (PDD) distributions for the quality assurance of clinical proton beams is most commonly performed with a computerized water tank dosimetry system with ionization chamber, commonly referred to as water tank. Although the accuracy and reproducibility of this method is well established, it can be time-consuming if a large number of measurements are required. In this work the authors evaluate the linearity, reproducibility, sensitivity to field size, accuracy, and time-savings of another system: the Zebra, a multilayer ionization chamber system. METHODS The Zebra, consisting of 180 parallel-plate ionization chambers with 2 mm resolution, was used to measure depth-dose distributions. The measurements were performed for scattered and scanned proton pencil beams of multiple energies delivered by the Hitachi PROBEAT synchrotron-based delivery system. For scattered beams, the Zebra-measured depth-dose distributions were compared with those measured with the water tank. The principal descriptors extracted for comparisons were: range, the depth of the distal 90% dose; spread-out Bragg peak (SOBP) length, the region between the proximal 95% and distal 90% dose; and distal-dose fall off (DDF), the region between the distal 80% and 20% dose. For scanned beams, the Zebra-measured ranges were compared with those acquired using a Bragg peak chamber during commissioning. RESULTS The Zebra demonstrated better than 1% reproducibility and monitor unit linearity. The response of the Zebra was found to be sensitive to radiation field sizes greater than 12.5 × 12.5 cm; hence, the measurements used to determine accuracy were performed using a field size of 10 × 10 cm. For the scattered proton beams, PDD distributions showed 1.5% agreement within the SOBP, and 3.8% outside. Range values agreed within -0.1 ± 0.4 mm, with a maximum deviation of 1.2 mm. SOBP length values agreed within 0 ± 2 mm, with a maximum deviation of 6 mm. DDF values agreed within 0.3 ± 0.1 mm, with a maximum deviation of 0.6 mm. For the scanned proton pencil beams, Zebra and Bragg peak chamber range values demonstrated agreement of 0.0 ± 0.3 mm with a maximum deviation of 1.3 mm. The setup and measurement time for all Zebra measurements was 3 and 20 times less, respectively, compared to the water tank measurements. CONCLUSIONS Our investigation shows that the Zebra can be useful not only for fast but also for accurate measurements of the depth-dose distributions of both scattered and scanned proton beams. The analysis of a large set of measurements shows that the commonly assessed beam quality parameters obtained with the Zebra are within the acceptable variations specified by the manufacturer for our delivery system.


Journal of Applied Clinical Medical Physics | 2014

Relative stopping power measurements to aid in the design of anthropomorphic phantoms for proton radiotherapy

R Grant; P Summers; James L. Neihart; Anthony P. Blatnica; Narayan Sahoo; M Gillin; D Followill; Geoffrey S. Ibbott

The delivery of accurate proton dose for clinical trials requires that the appropriate conversion function from Hounsfield unit (HU) to relative linear stopping power (RLSP) be used in proton treatment planning systems (TPS). One way of verifying that the TPS is calculating the correct dose is an end‐to‐end test using an anthropomorphic phantom containing tissue‐equivalent materials and dosimeters. Many of the phantoms in use for such end‐to‐end tests were originally designed using tissue‐equivalent materials that had physical characteristics to match patient tissues when irradiated with megavoltage photon beams. The aim of this study was to measure the RLSP of materials used in the phantoms, as well as alternative materials to enable modifying phantoms for use at proton therapy centers. Samples of materials used and projected for use in the phantoms were measured and compared to the HU assigned by the treatment planning system. A percent difference in RLSP of 5% was used as the cutoff for materials deemed acceptable for use in proton therapy (i.e., proton equivalent). Until proper tissue‐substitute materials are identified and incorporated, institutions that conduct end‐to‐end tests with the phantoms are instructed to override the TPS with the measured stopping powers we provide. To date, the RLSPs of 18 materials have been measured using a water phantom and/or multilayer ion chamber (MLIC). Nine materials were identified as acceptable for use in anthropomorphic phantoms. Some of the failing tissue substitute materials are still used in the current phantoms. Further investigation for additional appropriate tissue substitute materials in proton beams is ongoing. Until all anthropomorphic phantoms are constructed of appropriate materials, a unique HU‐RLSP phantom has been developed to be used during site visits to verify the proton facilitys treatment planning HU‐RLSP calibration curve. PACS number: 87.53.Bn


Medical Physics | 2013

Independent dose per monitor unit review of eight U.S.A. proton treatment facilities

Michael F. Moyers; Geoffrey S. Ibbott; R Grant; P Summers; D Followill

PURPOSE Compare the dose per monitor unit at different proton treatment facilities using three different dosimetry methods. METHODS Measurements of dose per monitor unit were performed by a single group at eight facilities using 11 test beams and up to six different clinical portal treatment sites. These measurements were compared to the facility reported dose per monitor unit values. RESULTS Agreement between the measured and reported doses was similar using any of the three dosimetry methods. Use of the ICRU 59 ND,w based method gave results approximately 3% higher than both the ICRU 59 NX and ICRU 78 (TRS-398) ND,w based methods. CONCLUSIONS Any single dosimetry method could be used for multi-institution trials with similar conformity between facilities. A multi-institutional trial could support facilities using both the ICRU 59 NX based and ICRU 78 (TRS-398) ND,w based methods but use of the ICRU 59 ND,w based method should not be allowed simultaneously with the other two until the difference is resolved.


Medical Physics | 2013

TU‐A‐108‐07: Design and Verification of a Heterogeneous Proton Equivalent Thorax Phantom for Use in End‐To‐End Assessment of Pencil Beam Proton Therapy

J Neihart; Narayan Sahoo; P Balter; P Summers; Matthew R. Palmer; M Kerr; D Followill

PURPOSE Design and commission a dynamic proton-equivalent lung phantom for implementation as an end-to-end audit tool for credentialing proton therapy centers participating in proton lung clinical trials. METHODS Phantom materials were tested and chosen with relative stopping powers (RSP) similar to those found on the Eclipse tissue calibration curve. One material was determined to be in good agreement with the Eclipse HU vs. RSP calibration curve to simulate bone. A phantom was designed with materials that could be imaged, planned, and treated without need for manual adjustment of HU. The lung target was designed to simulate 2cm respiratory motion. The phantom was simulated on a 4DCT and planned on the resulting average CT images. A proton pencil beam plan was generated to cover the ITV and spare other structures, and was verified by 2-D ion chamber array QA measurements. Image-guided spot scanned therapy was delivered to the phantom. The film and TLD within the moving dosimetry insert were read and registered to the planned 3D dose distribution. Using TLD for absolute dose and film for dose distribution, a gamma analysis was performed in the sagittal, coronal, and axial planes. RESULTS A bone equivalent material was found with an agreement of 0.9% to Eclipse calculation. For the two initial spot scanned proton beam treatment deliveries, the Gamma index ranged from 86% to 73% at ±5%/5mm about an RPC/Institution ratio of 0.97 for all film planes. The regions of highest failure were found to be in the distal region beyond the target volume. The measured vs. calculated dose ratio ranged from 0.96-1.01 with an average of 0.98. CONCLUSION A proton-equivalent lung phantom, with a dynamic thoracic target, was designed and tested as an audit tool for credentialing proton therapy centers. Spot scanned therapy without motion management was determined inadequate for 2cm motion. This work was supported by Public Health Service grants CA010953, CA081647, and CA21661 awarded by the National Cancer Institute, United States Department of Health and Human Services.


Medical Physics | 2014

SU-E-CAMPUS-T-03: Development and Implementation of An Anthropomorphic Pediatric Spine Phantom for the Assessment of Craniospinal Irradiation Procedures in Proton Therapy

D Lewis; P Summers; D Followill; Narayan Sahoo; Anita Mahajan; Francesco C. Stingo; Stephen F. Kry

PURPOSE To design an anthropomorphic pediatric spine phantom for use in the evaluation of proton therapy facilities for clinical trial participation by the Imaging and Radiation Oncology Core (IROC) Houston QA Center (formerly RPC). METHODS This phantom was designed to perform an end-to-end audit of the proton spine treatment process, including simulation, dose calculation by the treatment planning system (TPS), and proton treatment delivery. The design incorporated materials simulating the thoracic spinal column of a pediatric patient, along with two thermoluminescent dosimeter (TLD)-100 capsules and radiochromic film embedded in the phantom for dose evaluation. Fourteen potential materials were tested to determine relative proton stopping power (RSP) and Hounsfield unit (HU) values. Each material was CT scanned at 120kVp, and the RSP was obtained from depth ionization scans using the Zebra multilayer ion chamber (MLIC) at two energies: 160 MeV and 250 MeV. To determine tissue equivalency, the measured RSP for each material was compared to the RSP calculated by the Eclipse TPS for a given HU. RESULTS The materials selected as bone, tissue, and cartilage substitutes were Techron HPV Bearing Grade (Boedeker Plastics, Inc.), solid water, and blue water, respectively. The RSP values did not differ by more than 1.8% between the two energies. The measured RSP for each selected material agreed with the RSP calculated by the Eclipse TPS within 1.2%. CONCLUSION An anthropomorphic pediatric proton spine phantom was designed to evaluate proton therapy delivery. The inclusion of multiple tissue substitutes increases heterogeneity and the level of difficulty for institutions to successfully treat the phantom. The following attributes will be evaluated: absolute dose agreement, distal range, field width, junction match and right/left dose profile alignment. The phantom will be tested at several institutions using a 5% dose agreement criterion, and a 5%/3mm gamma analysis criterion for the film planes. Work supported by grants CA10953, CA059267, and CA81647 (NCI, DHHS).


Medical Physics | 2014

SU-E-T-509: Validation of the Use of OSLD for Carbon Beam Remote Dosimetry

P Summers; J Lowenstein; Oliver Jäkel; H Prokesch; P Alvarez; D Followill

PURPOSE To describe the commissioning of Aluminum Oxide Optically Stimulated Luminescent Dosimeters (OSLD) for the use in Carbon beam remote dosimetry for centers participating in NCI-funded cooperative group clinical trials. METHODS As Carbon therapy centers express interest in participating in cooperative group clinical trials, the Imaging and Radiation Oncology Core Group (IROC) Houston QA Center (formerly RPC) is developing a way to remotely monitor the machine output of these Carbon facilities. OSLD have been commissioned for photon, electron and proton dosimetry, so an experiment was designed to commission the same dosimeters for Carbon. OSLD were irradiated in a Carbon therapy beam produced by the Siemens synchrotron at the Heidelberg Ion Therapy facility in Heidelberg, Germany. The OSLD were placed in acrylic phantoms, imaged with a CT scanner, and plans were developed using the Siemens treatment planning system. The OSLD were irradiated in uniform fields with maximum energies of 216, 301, and 402 MeV and at dose levels of 50, 100, 200 and 300 cGy. RESULTS The response of the OSLD in the Carbon beam, as compared to the Cobalt-60 reference condition, required an energy correction of 1.85 to account for the particle correction. OSLD dose calculations typically have a linearity correction to account for the change in response relative to the change in dose delivered. However, the response of the OSLD in the Carbon beam was found to be independent of the dose level; thus the linearity correction is 1.00. CONCLUSION IROC Houston has commissioned OSLD for the use of remote output checks for Carbon therapy facilities to help ensure consistency across clinical trial participants. Work supported by grant CA10953 (NCI, DHHS).


Medical Physics | 2014

SU‐E‐T‐133: Dosimetric Impact of Scan Orientation Relative to Target Motion During Spot Scanning Proton Therapy

J Stoker; P Summers; Xiaoqiang Li; Daniel R. Gomez; Narayan Sahoo; X Zhu; M Gillin

PURPOSE This study seeks to evaluate the dosimetric effects of intra-fraction motion during spot scanning proton beam therapy as a function of beam-scan orientation and target motion amplitude. METHOD Multiple 4DCT scans were collected of a dynamic anthropomorphic phantom mimicking respiration amplitudes of 0 (static), 0.5, 1.0, and 1.5 cm. A spot-scanning treatment plan was developed on the maximum intensity projection image set, using an inverse-planning approach. Dynamic phantom motion was continuous throughout treatment plan delivery. The target nodule was designed to accommodate film and thermoluminescent dosimeters (TLD). Film and TLDs were uniquely labeled by location within the target. The phantom was localized on the treatment table using the clinically available orthogonal kV on-board imaging device. Film inserts provided data for dose uniformity; TLDs provided a 3% precision estimate of absolute dose. An inhouse script was developed to modify the delivery order of the beam spots, to orient the scanning direction parallel or perpendicular to target motion. TLD detector characterization and analysis was performed by the Imaging and Radiation Oncology Core group (IROC)-Houston. Film inserts, exhibiting a spatial resolution of 1mm, were analyzed to determine dose homogeneity within the radiation target. RESULTS Parallel scanning and target motions exhibited reduced target dose heterogeneity, relative to perpendicular scanning orientation. The average percent deviation in absolute dose for the motion deliveries relative to the static delivery was 4.9±1.1% for parallel scanning, and 11.7±3.5% (p<<0.05) for perpendicularly oriented scanning. Individual delivery dose deviations were not necessarily correlated to amplitude of motion for either scan orientation. CONCLUSIONS Results demonstrate a quantifiable difference in dose heterogeneity as a function of scan orientation, more so than target amplitude. Comparison to the analyzed planar dose of a single field hint that multiple-field delivery alters intra-fraction beam-target motion synchronization and may mitigate heterogeneity, though further study is warranted.


Medical Physics | 2014

SU-C-BRD-07: The Radiological Physics Center (RPC): 45 Years of Improving Radiotherapy Dosimetry

D Followill; J Lowenstein; A Molineu; P Alvarez; P Summers; Stephen F. Kry

PURPOSE The RPC, established in 1968 has contributed to the development, conduct, and QA of NCI funded multi-institutional cooperative group clinical trials and institutions, primarily in the USA/Canada and 242 other countries, participating in trials. METHODS The RPC QA program components were designed to audit the radiation dose calculation chain from the NIST traceable reference beam calibration, to inclusion of dosimetry parameters used to calculate tumor doses, to the delivery of the radiation dose. The QA program included: 1) remote TLD/OSLD audit of machine output, 2) on-site dosimetry review visits, 3) credentialing for advanced technologies, and 4) review of patient treatment records. The RPC presented and published their findings to the radiation oncology community. RESULTS The number of institutions monitored by the RPC increased from around 1200 in the late 90s, to ∼2000 in 2013. There were over 4000 megavoltage therapy machines and ∼28,000 therapy beams in the 1991 institutions monitored by the RPC by the end of 2013. Within the 14,000 photon, electron and proton beam outputs remotely monitored with TLD/OSLD annually, between 10-20% of the institutions have one or more beams outside the RPC 5% criterion. Dosimetry site visits to photon and proton centers continue to result in 2-4 recommendations affecting key dosimetry parameters that impact patient treatment times. One in four patient treatment records reviewed by the RPC have their dose data corrected by >5% before trial groups use them for outcomes analysis. Twelve of fourteen clinically active proton centers are approved to participate in NCI funded clinical trials. The RPC published 222 peer reviewed articles since 1972. CONCLUSION Findings from the RPC suggest that human errors continue to play a role in radiotherapy discrepancies and without the RPC independent QA program, the number of undetected errors and time elapsed before their discovery would have been greater. Work supported by MGH C06 CA059267 and grants CA10953, CA081647 awarded by NCI, DHHS.


Medical Physics | 2014

TU-C-BRE-03: Aggregation of Linac Measurement Data

J Kerns; P Alvarez; D Followill; J Lowenstein; A Molineu; P Summers; Stephen F. Kry

PURPOSE Accurate data of linear accelerator radiation characteristics is important for treatment planning system commissioning as well as regular quality assurance of the machine. The RPC has performed site visits of numerous machines . Data gathered from Varian machines from the past 15 years are presented. The data collected can be used as a secondary check or when commissioning a new machine to verify that values are reasonable. METHODS Data from the past 15 years of RPC site visits was compiled and analyzed. Data was composed from measurements from approximately 400 Varian machines. Each dataset consists of several point measurements at various locations in a water phantom to measure percentage depth dose, output factors, including small MLC fields, off-axis factors, and wedge factors if applicable. Common statistical values are presented for each machine type. Where applicable, data was compared to other reference data given by the vendor or a select number of previous researchers. RESULTS Data is separated by energy and parameter and then analyzed by machine class. Data distributions of the parameter data were normal except occasionally at the tails. Distributions of the data for each class and parameter are tabulated to give not simply a singular reference value, but metrics about the distribution: 5th and 95th percentile values and the standard deviation as well as the median. CONCLUSION The RPC has collected numerous data on Varian linacs and presented the finding of the past 15 years. The data can be used as a reference data set for physicists to compare against. A linac that deviates from the values does not necessarily indicate there is a problem as long as the treatment planning system correlates to the machine. Comparison of linac and treatment planning system data to external reference data can prevent serious treatment errors.


Medical Physics | 2013

SU‐E‐T‐403: Intensity Modulated Proton Therapy Plans with Multiple Fields for Prostate Cancer

S Dhanesar; Narayan Sahoo; M Kerr; M Taylor; P Summers; P Alvarez; R Wu; F Poenisch; X Zhu; M Gillin

PURPOSE To investigate the possible improvements in the dose conformity in intensity modulated proton therapy (IMPT) plans with multiple fields as compared to widely used single field optimized (SFO) plans with two lateral fields for spot scanned proton therapy (SSPT) and to verify the accuracy of the delivered dose using the Radiological Physics Center Proton Prostate Phantom (RPCPPP). METHODS The CT images of RPCPPP were used to design multiple IMPT plans using both SFO and multi field optimization (MFO) methods in the Varian Eclipse treatment planning system (TPS). The standard SFO clinical plan used two lateral fields going through the femoral heads. Non-standard plans were generated using three or more oblique fields based on the SFO and MFO options of the TPS. The angles were carefully chosen to avoid bladder and rectum as much as possible. The RPC phantom containing films and TLDs was irradiated using the standard clinical SFO plan and the four field MFO plan to compare the planned and delivered dose. RESULTS Based on the DVHs, MFO plan with multiple oblique fields is found to have superior target conformity with reduced dose to organs at risk (OAR) and normal tissue, especially in the high dose regions, as compared to SFO plans. The point doses measured by TLDs and 2-D dose distribution measured by GafChromic films in the RPCPPP irradiation passed the commonly used RPC acceptance criteria, namely +/- 7% agreement in point does and more than 85% of the voxels meeting the gamma index criteria of 7% dose or 4 mm distance agreement. CONCLUSION The MFO SSPT plans for prostatic targets with multiple fields lead to better target conformity with lower doses to OARs as compared to SFO plans. The results of irradiations of the RPCPPP show that these plans can be delivered with acceptable accuracy.

Collaboration


Dive into the P Summers's collaboration.

Top Co-Authors

Avatar

D Followill

University of Texas MD Anderson Cancer Center

View shared research outputs
Top Co-Authors

Avatar

Geoffrey S. Ibbott

University of Texas MD Anderson Cancer Center

View shared research outputs
Top Co-Authors

Avatar

Narayan Sahoo

University of Texas MD Anderson Cancer Center

View shared research outputs
Top Co-Authors

Avatar

P Alvarez

University of Texas MD Anderson Cancer Center

View shared research outputs
Top Co-Authors

Avatar

Stephen F. Kry

University of Texas MD Anderson Cancer Center

View shared research outputs
Top Co-Authors

Avatar

M Gillin

University of Texas MD Anderson Cancer Center

View shared research outputs
Top Co-Authors

Avatar

F Poenisch

University of Texas MD Anderson Cancer Center

View shared research outputs
Top Co-Authors

Avatar

J Lowenstein

University of Texas MD Anderson Cancer Center

View shared research outputs
Top Co-Authors

Avatar

R Grant

University of Texas MD Anderson Cancer Center

View shared research outputs
Top Co-Authors

Avatar

A Molineu

University of Texas MD Anderson Cancer Center

View shared research outputs
Researchain Logo
Decentralizing Knowledge