Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Yanjun Hua is active.

Publication


Featured researches published by Yanjun Hua.


Ophthalmology | 2013

A Comparison between Scheimpflug Imaging and Optical Coherence Tomography in Measuring Corneal Thickness

Jinhai Huang; Xixia Ding; Giacomo Savini; Chao Pan; Yifan Feng; Dan Cheng; Yanjun Hua; Xiuli Hu; Qinmei Wang

PURPOSE To assess the repeatability and reproducibility of 3 rotating Scheimpflug cameras, the Pentacam (Oculus, Wetzlar, Germany), Sirius (Costruzione Strumenti Oftalmici, Florence, Italy), and Galilei (Ziemer, Biel, Switzerland), and 1 Fourier-domain optical coherence tomography (FD-OCT) system, the RTvue-100 OCT (Optovue Inc., Fremont, CA), in measuring corneal thickness. DESIGN Evaluation of diagnostic test. PARTICIPANTS Sixty-six right eyes of 66 healthy volunteers, whose mean age ± standard deviation (SD) was 35.39±10.06 years (range, 18-55 years). METHODS Corneal thickness measurements obtained by each system included central corneal thickness (CCT), thinnest corneal thickness (TCT), and midperipheral corneal thickness (MPCT), measured at superior, inferior, nasal, and temporal locations at a distance of 1 and 2.5 mm from the corneal apex. In the first session, 3 consecutive measurements were performed by the same operator to assess intraobserver repeatability and by a second operator to assess interobserver reproducibility. Measurements were repeated in the second session scheduled 1 day to 1 week later. The mean values obtained in the 2 sessions by the first operator were used to investigate the intersession reproducibility. MAIN OUTCOME MEASURES Intraobserver repeatability and interobserver and intersession repeatability of corneal thickness measurements, as calculated by means of within-subject SD, test-retest repeatability, coefficient of variation (COV), and intraclass correlation coefficients. RESULTS The precision of CCT, TCT, CT2 mm (midperipheral corneal thickness [MPCT] with a distance of 1 mm from the corneal apex), and CT5 mm (midperipheral corneal thickness [MPCT] with a distance of 2.5 mm from the corneal apex) measurements was high with all 4 systems. The COV was ≤1.16%, 0.94%, and 1.10% for repeatability, interobserver reproducibility, and intersession reproducibility, respectively. The 4 devices offered better interobserver reproducibility than intersession reproducibility for all measurements. The CTsuperior-5 mm (midperipheral corneal thickness [MPCT], measured at superior locations with a distance of 2.5 mm from the corneal apex) measurements showed the poorest repeatability and reproducibility. The Galilei revealed the best precision of CCT, TCT, and CT2 mm measurements. CONCLUSIONS Both Scheimpflug imaging and FD-OCT offer highly repeatable and reproducible measurements of CCT and MPCT. The precision was lower in the midperipheral superior quadrant. FINANCIAL DISCLOSURE(S) The author(s) have no proprietary or commercial interest in any materials discussed in this article.


PLOS ONE | 2012

A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices

Qinmei Wang; Giacomo Savini; Kenneth J. Hoffer; Zhen Xu; Yifan Feng; Daizong Wen; Yanjun Hua; Feng Yang; Chao Pan; Jinhai Huang

Purpose To comprehensively assess the precision and agreement of anterior corneal power measurements using 8 different devices. Methods Thirty-five eyes from 35 healthy subjects were included in the prospective study. In the first session, a single examiner performed on each subject randomly measurements with the RC-5000 (Tomey Corp., Japan), KR-8000 (Topcon, Japan), IOLMaster (Carl Zeiss Meditec, Germany), E300 (Medmont International, Australia), Allegro Topolyzer (Wavelight AG, Germany), Vista (EyeSys, TX), Pentacam (Oculus, Germany) and Sirius (CSO, Italy). Measurements were repeated in the second session (1 to 2 weeks later). Repeatability and reproducibility of corneal power measurements were assessed based on the intrasession and intersession within-subject standard deviation (Sw), repeatability (2.77Sw), coefficient of variation (COV), and intraclass correlation coefficient (ICC). Agreement was evaluated by 95% limits of agreement (LoA). Results All devices demonstrated high repeatability and reproducibility of the keratometric values (2.77Sw<0.36D, COV<0.3%, ICC>0.98). Repeated-measures analysis of variance with Bonferroni post test showed statistically significant differences (P<0.01) among mean keratometric values of most instruments; the largest differences were observed between the EyeSys Vista and Medmont E300. Good agreement (i.e., 95%LoA within ±0.5D) was found between most instruments for flat, steep and mean keratometry, except for EyeSys and Medmont. Repeatability and reproducibility of vectors J0 and J45 was good, as the ICCs were higher than 0.9, except J45 of Medmont and Pentacam. For the 95% LoAs of J0 and J45, they were all ≤ ±0.31 among any two paired devices. Conclusions The 8 devices showed excellent repeatability and reproducibility. The results obtained using the RC-5000, KR-8000, IOLMaster, Allegro Topolyzer, Pentacam and Sirius were comparable, suggesting that they could be used interchangeably in most clinical settings. Caution is warranted with the measurements of the EyeSys Vista and Medmont E300, which should not be used interchangeably with other devices due to lower agreement. Trial Registration ClinicalTrials.gov NCT01587287


Optometry and Vision Science | 2012

The effect of cycloplegia on the lenstar and the IOLMaster biometry.

Jinhai Huang; Colm McAlinden; Binbin Su; Konrad Pesudovs; Yifan Feng; Yanjun Hua; Feng Yang; Chao Pan; Huyong Zhou; Qinmei Wang

Purpose To evaluate the effect of cycloplegia on ocular biometry measurements and intraocular lens (IOL) power calculation using the Lenstar LS900 (Haag-Streit AG, Koeniz, Switzerland) and the IOLMaster (Carl Zeiss Meditec AG, Jena, Germany) biometers and to assess the agreement between the devices. Methods Measurements were taken with the Lenstar and the IOLMaster on 43 healthy volunteers with a mean age of 22.1 ± 4.7 years (range, 18 to 37 years). Axial length (AL), anterior chamber depth (ACD), corneal curvature, and horizontal iris width (white-to-white [WTW]) measurements were performed with and without cycloplegia. The IOL powers were calculated using four formulas: Sanders-Retzlaff-Kraff/Theoretical, Holladay 1, Hoffer Q, and Haigis. Results Cycloplegia had no significant effect on AL or corneal curvature. However, ACD and WTW significantly increased postcycloplegia (Lenstar, 0.09 ± 0.06 mm and 0.10 ± 0.17 mm, respectively; IOLMaster, 0.06 ± 0.07 mm and 0.43 ± 0.35 mm, respectively; p <0.001). The Lenstar AL measurements were statistically but not clinically significantly longer than those of the IOLMaster (precycloplegia, 0.03 ± 0.03 mm; postcycloplegia, 0.02 ± 0.03 mm; p < 0.001). For ACD measurements, the 95% limits of agreement were −0.19 to 0.20 mm without cycloplegia and −0.11 to 0.17 mm with cycloplegia. The 95% limits of agreement for WTW measurements were −1.07 to 0.45 mm with cycloplegia. The only significantly different IOL power precycloplegia and postcycloplegia was with the Haigis formula and the Lenstar measurements: 15.12 ± 3.87 diopters and 15.26 ± 3.92 diopters (p < 0.01). Conclusions Cycloplegia affected ACD and WTW but not AL or corneal curvature measurements. Generally, good agreement was found between the Lenstar and the IOLMaster, although not for WTW. Differences between these devices do not produce a clinically significant impact on IOL power.


PLOS ONE | 2014

Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes

Xiangjun Chen; Aleksandar Stojanovic; Yanjun Hua; Jon Roger Eidet; Di Hu; Jingting Wang; Tor Paaske Utheim

Purpose To determine the measurement reliability of CorVis ST, a dynamic Scheimpflug analyser, in virgin and post-photorefractive keratectomy (PRK) eyes and compare the results between these two groups. Methods Forty virgin eyes and 42 post-PRK eyes underwent CorVis ST measurements performed by two technicians. Repeatability was evaluated by comparing three consecutive measurements by technician A. Reproducibility was determined by comparing the first measurement by technician A with one performed by technician B. Intraobserver and interobserver intraclass correlation coefficients (ICCs) were calculated. Univariate analysis of covariance (ANCOVA) was used to compare measured parameters between virgin and post-PRK eyes. Results The intraocular pressure (IOP), central corneal thickness (CCT) and 1st applanation time demonstrated good intraobserver repeatability and interobserver reproducibility (ICC≧0.90) in virgin and post-PRK eyes. The deformation amplitude showed a good or close to good repeatability and reproducibility in both groups (ICC≧0.88). The CCT correlated positively with 1st applanation time (r = 0.437 and 0.483, respectively, p<0.05) and negatively with deformation amplitude (r = −0.384 and −0.375, respectively, p<0.05) in both groups. Compared to post-PRK eyes, virgin eyes showed longer 1st applanation time (7.29±0.21 vs. 6.96±0.17 ms, p<0.05) and lower deformation amplitude (1.06±0.07 vs. 1.17±0.08 mm, p<0.05). Conclusions CorVis ST demonstrated reliable measurements for CCT, IOP, and 1st applanation time, as well as relatively reliable measurement for deformation amplitude in both virgin and post-PRK eyes. There were differences in 1st applanation time and deformation amplitude between virgin and post-PRK eyes, which may reflect corneal biomechanical changes occurring after the surgery in the latter.


Journal of Cataract and Refractive Surgery | 2015

Anterior chamber depth measurements using Scheimpflug imaging and optical coherence tomography: Repeatability, reproducibility, and agreement

Qinmei Wang; Xixia Ding; Giacomo Savini; Hao Chen; Yifan Feng; Chao Pan; Yanjun Hua; Jinhai Huang

Purpose To determine the repeatability, reproducibility, and agreement of anterior chamber depth (ACD) measurements obtained with 3 Scheimpflug cameras and an anterior segment optical coherence tomography (AS‐OCT) device. Setting Eye Hospital of Wenzhou Medical University, Zhejiang, China. Design Observational cross‐sectional study. Methods Two observers took 3 consecutive measurements in healthy right eyes using each device to assess intraoperator repeatability. The mean values obtained at different sessions by the first operator were used to determine the intersession reproducibility. Three consecutive measurements obtained by the first operator at the first session were averaged and used to assess agreement. Results The ACD measurements obtained by 2 observers in 71 eyes were highly repeatable using the 4 devices, with a test–retest repeatability of 0.04 to 0.07 mm for intraoperator repeatability. The interoperator and intersession reproducibility of ACD measurements were high, and the test–retest repeatability ranges of interoperator and intersession reproducibility were 0.06 to 0.07 mm and 0.05 to 0.08 mm, respectively. The ACD measurements of the 4 systems were sorted from the thickest to the thinnest (Galilei G2 > Visante > Sirius > Pentacam). The differences in the measurements were statistically significant except between the ACD measurements obtained by the Sirius device and the Visante device. However, good agreement with narrow 95% limits of agreement was found between these devices. Conclusions The 4 devices provided high intraoperator repeatability and interoperator and intersession reproducibility for ACD measurements. Good agreement in ACD measurements was found between the devices in healthy eyes. Financial Disclosure No author has a financial or proprietary interest in any material or method mentioned.


PLOS ONE | 2014

Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography

Jinhai Huang; Xixia Ding; Giacomo Savini; Zhengxuan Jiang; Chao Pan; Yanjun Hua; Fan Wu; Yifan Feng; Ye Yu; Qinmei Wang

Purpose To compare corneal thickness measurements using Pentacam (Oculus, Germany), Sirius (CSO, Italy), Galilei (Ziemer, Switzerland), and RTVue-100 OCT (Optovue Inc., USA). Methods Sixty-six eyes of 66 healthy volunteers were enrolled. Three consecutive measurements were performed with each device. The mean value of the three measurements was used for subsequent analysis. Central corneal thickness (CCT), thinnest corneal thickness (TCT), and midperipheral corneal thickness (MPCT; measured at superior, inferior, nasal, and temporal locations with a distance of 1 mm (CT2mm) or 2.5 mm (CT5mm) from the corneal apex) were analyzed. Differences and agreement between measurements were assessed using the repeated-measures analysis of variance (ANOVA) and Bland-Altman analyses, respectively. Results Statistically significant differences (p<0.001) among the four devices were revealed in CCT, TCT and CT2mmmeasurements. The CCT, TCT, and CT2mm values were ranked from the thickest to the thinnest as follows: Galilei>Sirius>Pentacam>RTVue OCT. For these measurements, agreement between measurements by Sirius and Pentacam was good, whereas Galilei overestimated and RTVue underestimated corneal thickness compared to Sirius and Pentacam. As regards CT5mm measurements, Pentacam provided the largest values, whereas RTVue OCT yielded the smallest values. Agreement of the CT5mm measurements was good between the Pentacam, Sirius, moderate between Galilei and the other two Scheimpflug systems, and poor between the RTVue OCT and the remaining devices. Conclusions The Pentacam and Sirius can be used interchangeably for CCT measurements, while the Galilei and RTVue systematically over- and underestimate CCT, respectively. The three Scheimpflug cameras, but not the RTVue, may be used interchangeably for MPCT measurements.


Cornea | 2015

Corneal Power Measurement Obtained by Fourier-Domain Optical Coherence Tomography: Repeatability, Reproducibility, and Comparison With Scheimpflug and Automated Keratometry Measurements.

Qinmei Wang; Yanjun Hua; Giacomo Savini; Hao Chen; Fangjun Bao; Shishi Lin; Weicong Lu; Jinhai Huang

Purpose: To assess the repeatability and reproducibility of corneal power values obtained by a Fourier-domain optical coherence tomography (FD-OCT) system (RTVue) and to compare them with the values obtained by a Scheimpflug camera system (Pentacam HR) and by automated keratometry (IOL Master). Methods: Thirty-two eyes from 32 healthy subjects were included in this prospective study. Two experienced observers measured each eye 3 consecutive times with the Pentacam, IOLMaster, and RTVue centered on either the pupil or corneal vertex. The conventional keratometry equivalent (CKE) and anterior (Ka), posterior (Kp), and net (Kn) corneal power values were determined. Results: The corneal power values obtained by the RTVue showed high repeatability (all intraclass correlation coefficient >0.96) and reproducibility (coefficient of variation <1.0%). Pupil-centered FD-OCT performed slightly better than corneal vertex-centered FD-OCT. Mean corneal values had higher reproducibly than any of the individual values. CKE, Ka, Kp, and Kn obtained by FD-OCT were 0.62 to 0.68 diopters (D), 0.70 to 0.76 D, 0.11 to 0.13 D, and 0.93 to 0.94 D higher than those obtained by the Pentacam HR, respectively. CKE and Ka obtained with the RTVue were also 0.60 to 0.74 D higher than those obtained with the IOLMaster, respectively. Conclusions: The corneal power measurements obtained by the RTVue FD-OCT system showed high repeatability and reproducibility. Measurements obtained by FD-OCT with pupil centration were more reproducible than those obtained by FD-OCT with corneal vertex centration. We recommend that pupil-centered FD-OCT be used in clinical applications. Neither RTVue versus Pentacam HR nor RTVue versus IOLMaster can be used interchangeably.


PLOS ONE | 2016

Precision (Repeatability and Reproducibility) and Agreement of Corneal Power Measurements Obtained by Topcon KR-1W and iTrace

Yanjun Hua; Zequan Xu; Wei Qiu; Qiang Wu

Purpose To evaluate the repeatability and reproducibility of corneal power measurements obtained by Topcon KR-1W and iTrace, and assess the agreement with measurements obtained by Allegro Topolyzer and IOLMaster. Methods The right eyes of 100 normal subjects were prospectively scanned 3 times using all the 4 devices. Another observer performed additional 3 consecutive scans using the Topcon KR-1W and iTrace in the same session. About one week later, the first observer repeated the measurements using the Topcon KR-1W and iTrace. The steep keratometry (Ks), flat keratometry (Kf), mean keratometry (Km), J0 and J45 were analyzed. Repeatability and reproducibility of measurements were evaluated by the within-subject standard deviation (Sw), coefficient of variation (CoV), test-retest repeatability (2.77Sw), and intraclass correlation coefficient (ICC). Agreements between devices were assessed using Bland-Altman analysis and 95% limits of agreement (LoA). Results Intraobserver repeatability and interobserver and intersession reproducibility of the Ks, Kf and Km showed a CoV of no more than 0.5%, a 2.77Sw of 0.70 D or less, and an ICC of no less than 0.99. However, J0 and J45 showed poor intraobserver repeatability and interobserver and intersession reproducibility (all ICCs not greater than 0.446). Statistically significant differences existed between Topcon KR-1W and IOLMaster, Topcon KR-1W and iTrace, Topcon KR-1W and Topolyzer, iTrace and Topolyzer, iTrace and IOLMaster for Ks, Kf and Km measurements (all P < 0.05). The mean differences between Topcon KR-1W, iTrace, and the other 2 devices were small. The 95% LoA were approximately 1.0 D to 1.5 D for all measurements. Conclusions The Ks, Kf and Km obtained by Topcon KR-1W and iTrace showed excellent intraobserver repeatability and interobserver and intersession reproducibility in normal eyes. The agreement between Topcon KR-1W and Topolyzer, Topcon KR-1W and IOLMaster, iTrace and Topolyzer, iTrace and IOLMaster, Topcon KR-1W and iTrace were not so good, they should not be interchangeable in clinical application. Given that the intraobserver repeatability and interobserver and intersession reproducibility of corneal astigmatism measurements obtained by Topcon KR-1W and iTrace were poor, it should be cautious that Topcon KR-1W and iTrace were applied for the preparation of toric lens implantation.


PLOS ONE | 2016

Evaluation of Equivalent Keratometry Readings Obtained by Pentacam HR (High Resolution)

Yanjun Hua; Xiaolan Zhang; Tor Paaske Utheim; Jinhai Huang; Chao Pan; Weina Tan; Qinmei Wang

Purpose To assess the repeatability of Equivalent Keratometry Readings (EKRs) obtained by the Pentacam HR (high resolution) in untreated and post-LASIK eyes, and to compare them with the keratometry (K) values obtained by other algorithms. Methods In this prospective study, 100 untreated eyes and 71 post-LASIK eyes were included. In the untreated group, each eye received 3 consecutive scans using the Pentacam HR, and EKR values in all central corneal zone, the true net power (Knet) and the simulated K (SimK) were obtained for each scan. In the post-LASIK group, each eye received subjective refraction and 3 consecutive scans with the Pentacam HR preoperatively. During the 3-month post-surgery exam, the same examinations and the use of an IOLMaster were conducted for each eye. The EKRs in all zone, the Knet, the mean K (Km) by IOLMaster and the K values by clinical history method (KCHM) were obtained. The repeatability of the EKRs was assessed by the within-subject standard deviation (Sw), 2.77Sw, coefficient of variation (CVw) and intraclass correlation coefficient (ICC). The bonferroni corrected multiple comparisons were performed to analyze the differences among the EKRs and K values calculated by other algorithms within the 2 groups. The 95% limits of agreement (LoA) were calculated. Results The EKR values in all central corneal zone were repeatable in both the untreated group (Sw≦0.19 D, 2.77Sw≦0.52 D, CVw≦1%, ICC≧0.978) and the post-LASIK group (Sw≦0.22 D, 2.77Sw≦0.62 D, CVw≦1%, ICC≧0.980). In the untreated group, the EKR in 4mm zone was close to SimK (P = 1.000), and the 95% LoA was (-0.13 to 0.15 D). The difference between Knet and SimK was -1.30±0.13 D (95% LoA -1.55 to -1.55 D, P<0.001). In the post-LASIK group, all the EKRs were significantly higher than KCHM (all P<0.001). The differences between the EKR in 4mm zone and KCHM, the EKR in 7mm zone and KCHM, Knet and KCHM, Km and KCHM, SimK and Knet were 0.64±0.50 D (95% LoA, -0.33 to 1.62 D), 1.77±0.88 D (95% LoA, 0.04 to 3.51 D), -0.98±0.48 D (95% LoA, -1.92 to -0.04 D), 0.64±0.53 D (95% LoA, -0.40 to 1.68 D), and 1.73±0.20 D (95% LoA, 1.33 to 2.13 D), respectively. Conclusions The EKRs obtained by the Pentacam HR were repeatable in both untreated eyes and post-LASIK eyes. Compared to the total corneal power obtained by the clinical history method, the EKR values generally overestimated the total corneal power in post-LASIK eyes. So, further calibrations for the EKR values should be conducted, before they were used for the total corneal power assessment in post-LASIK eyes.


PLOS ONE | 2018

Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes

Yanjun Hua; Wei Qiu; Qiuyi Xiao; Qiang Wu

Purpose To assess the precision (repeatability and reproducibility) of ocular parameters measured by the Tomey OA-2000 biometer, and to compare them with those measured by the IOLMaster. Methods In this prospective study, the right eyes of 108 healthy subjects were included. Three consecutive scans were obtained by 2 observers using the Tomey OA-2000, and in the same session one observer used the IOLMaster (version 5.4.4.0006) for the measurements. About 1 week later, 3 scans were obtained by one observer using the Tomey OA-2000. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), keratometer readings, pupil diameter (PD) and corneal diameter (CD) values measured by the Tomey OA-2000 and IOLMaster were analyzed. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within subject standard deviation (Sw) and 2.77Sw were calculated to assess the repeatability and reproducibility. The paired t test and Bland-Altman plots were used to analyze the differences and agreements of parameters measured by the two devices, respectively. Results Intraobserver repeatability, and interobserver and intersession reproducibility of the AL, CCT, ACD, LT, Kf, Ks, Km, PD and CD values measured by the Tomey OA-2000 biometer showed a CoV of less than 1% except that for PD, and an ICC of more than 0.97 except that for PD and CD. The AL, Kf, Ks, Km and CD values measured by the Tomey OA-2000 were 0.058 ± 0.094 mm, 0.088± 0.150 diopters (D), 0.163 ± 0.170 D, 0.127 ± 0.117 D and 0.171 ± 0.217 mm lower than those measured by the IOLMaster, respectively (all Ps < 0.05). However, the ACD values from the two devices were comparable (P = 0.169). The 95% linite of agreement (LoA) of the AL, ACD, CD and all keratometer readings were no more than 0.24 mm, 0.14 mm 0.60 mm and 0.5 D, respectively. Conclusion Except for the PD and CD, the ocular parameters measured by the Tomey OA-2000 were highly repeatable and reproducible. Except for the CD value, there was good agreement of ocular parameters measured by the Tomey OA-2000 and the IOLMaster in healthy eyes.

Collaboration


Dive into the Yanjun Hua's collaboration.

Top Co-Authors

Avatar

Jinhai Huang

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar

Qinmei Wang

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar

Chao Pan

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar

Yifan Feng

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Xixia Ding

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Feng Yang

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar

Hao Chen

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar

Qiang Wu

Shanghai Jiao Tong University

View shared research outputs
Researchain Logo
Decentralizing Knowledge