Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Yifan Feng is active.

Publication


Featured researches published by Yifan Feng.


Ophthalmology | 2013

A Comparison between Scheimpflug Imaging and Optical Coherence Tomography in Measuring Corneal Thickness

Jinhai Huang; Xixia Ding; Giacomo Savini; Chao Pan; Yifan Feng; Dan Cheng; Yanjun Hua; Xiuli Hu; Qinmei Wang

PURPOSE To assess the repeatability and reproducibility of 3 rotating Scheimpflug cameras, the Pentacam (Oculus, Wetzlar, Germany), Sirius (Costruzione Strumenti Oftalmici, Florence, Italy), and Galilei (Ziemer, Biel, Switzerland), and 1 Fourier-domain optical coherence tomography (FD-OCT) system, the RTvue-100 OCT (Optovue Inc., Fremont, CA), in measuring corneal thickness. DESIGN Evaluation of diagnostic test. PARTICIPANTS Sixty-six right eyes of 66 healthy volunteers, whose mean age ± standard deviation (SD) was 35.39±10.06 years (range, 18-55 years). METHODS Corneal thickness measurements obtained by each system included central corneal thickness (CCT), thinnest corneal thickness (TCT), and midperipheral corneal thickness (MPCT), measured at superior, inferior, nasal, and temporal locations at a distance of 1 and 2.5 mm from the corneal apex. In the first session, 3 consecutive measurements were performed by the same operator to assess intraobserver repeatability and by a second operator to assess interobserver reproducibility. Measurements were repeated in the second session scheduled 1 day to 1 week later. The mean values obtained in the 2 sessions by the first operator were used to investigate the intersession reproducibility. MAIN OUTCOME MEASURES Intraobserver repeatability and interobserver and intersession repeatability of corneal thickness measurements, as calculated by means of within-subject SD, test-retest repeatability, coefficient of variation (COV), and intraclass correlation coefficients. RESULTS The precision of CCT, TCT, CT2 mm (midperipheral corneal thickness [MPCT] with a distance of 1 mm from the corneal apex), and CT5 mm (midperipheral corneal thickness [MPCT] with a distance of 2.5 mm from the corneal apex) measurements was high with all 4 systems. The COV was ≤1.16%, 0.94%, and 1.10% for repeatability, interobserver reproducibility, and intersession reproducibility, respectively. The 4 devices offered better interobserver reproducibility than intersession reproducibility for all measurements. The CTsuperior-5 mm (midperipheral corneal thickness [MPCT], measured at superior locations with a distance of 2.5 mm from the corneal apex) measurements showed the poorest repeatability and reproducibility. The Galilei revealed the best precision of CCT, TCT, and CT2 mm measurements. CONCLUSIONS Both Scheimpflug imaging and FD-OCT offer highly repeatable and reproducible measurements of CCT and MPCT. The precision was lower in the midperipheral superior quadrant. FINANCIAL DISCLOSURE(S) The author(s) have no proprietary or commercial interest in any materials discussed in this article.


PLOS ONE | 2012

A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices

Qinmei Wang; Giacomo Savini; Kenneth J. Hoffer; Zhen Xu; Yifan Feng; Daizong Wen; Yanjun Hua; Feng Yang; Chao Pan; Jinhai Huang

Purpose To comprehensively assess the precision and agreement of anterior corneal power measurements using 8 different devices. Methods Thirty-five eyes from 35 healthy subjects were included in the prospective study. In the first session, a single examiner performed on each subject randomly measurements with the RC-5000 (Tomey Corp., Japan), KR-8000 (Topcon, Japan), IOLMaster (Carl Zeiss Meditec, Germany), E300 (Medmont International, Australia), Allegro Topolyzer (Wavelight AG, Germany), Vista (EyeSys, TX), Pentacam (Oculus, Germany) and Sirius (CSO, Italy). Measurements were repeated in the second session (1 to 2 weeks later). Repeatability and reproducibility of corneal power measurements were assessed based on the intrasession and intersession within-subject standard deviation (Sw), repeatability (2.77Sw), coefficient of variation (COV), and intraclass correlation coefficient (ICC). Agreement was evaluated by 95% limits of agreement (LoA). Results All devices demonstrated high repeatability and reproducibility of the keratometric values (2.77Sw<0.36D, COV<0.3%, ICC>0.98). Repeated-measures analysis of variance with Bonferroni post test showed statistically significant differences (P<0.01) among mean keratometric values of most instruments; the largest differences were observed between the EyeSys Vista and Medmont E300. Good agreement (i.e., 95%LoA within ±0.5D) was found between most instruments for flat, steep and mean keratometry, except for EyeSys and Medmont. Repeatability and reproducibility of vectors J0 and J45 was good, as the ICCs were higher than 0.9, except J45 of Medmont and Pentacam. For the 95% LoAs of J0 and J45, they were all ≤ ±0.31 among any two paired devices. Conclusions The 8 devices showed excellent repeatability and reproducibility. The results obtained using the RC-5000, KR-8000, IOLMaster, Allegro Topolyzer, Pentacam and Sirius were comparable, suggesting that they could be used interchangeably in most clinical settings. Caution is warranted with the measurements of the EyeSys Vista and Medmont E300, which should not be used interchangeably with other devices due to lower agreement. Trial Registration ClinicalTrials.gov NCT01587287


Journal of Cataract and Refractive Surgery | 2012

Scheimpflug–Placido topographer and optical low-coherence reflectometry biometer: Repeatability and agreement

Wuhe Chen; Colm McAlinden; Konrad Pesudovs; Qinmei Wang; Fan Lu; Yifan Feng; Jie Chen; Jinhai Huang

Purpose To assess the repeatability of common measurements with the Sirius Scheimpflug–Placido topographer and Lenstar LS900 optical low‐coherence reflectometry (OLCR) biometer and the limits of agreement (LoA) between the devices. Setting Eye Hospital of Wenzhou Medical College, Wenzhou, China. Design Comparative evaluation of a diagnostic test or technology. Methods One randomly healthy eye of subjects was scanned 3 times with both devices. The parameters assessed were central corneal thickness (CCT), anterior chamber depth (ACD) from the corneal epithelium and from the endothelium, mean keratometry (K), and white‐to‐white (WTW) corneal diameter. The repeatability of scans was calculated using the within‐subject standard deviation after 1‐way analysis of variance was performed. The agreement between devices was assessed using the Bland‐Altman LoA method, which equals the mean difference between devices ± 1.96 × standard deviation of the differences. The mean of 3 scans of each device was used to assess the LoA. Results Forty subjects were evaluated. The repeatability of the Scheimpflug–Placido topographer and OLCR biometer was 3.10 μm and 3.32 μm for CCT, 0.04 mm and 0.05 mm for WTW corneal diameter, and 0.17 D and 0.10 D for mean K, respectively. The repeatability for both devices was 0.02 mm for the ACD from the corneal epithelium and the ACD from the corneal endothelium. On Bland‐Altman LoA analysis, all parameters were within clinically acceptable limits. Conclusion Both devices had excellent repeatability for all parameters assessed. Good LoAs were found between the 2 devices, indicating they can be used interchangeably for the parameters assessed. Financial Disclosure No author has a financial or proprietary interest in any material or method mentioned.


Optometry and Vision Science | 2012

The effect of cycloplegia on the lenstar and the IOLMaster biometry.

Jinhai Huang; Colm McAlinden; Binbin Su; Konrad Pesudovs; Yifan Feng; Yanjun Hua; Feng Yang; Chao Pan; Huyong Zhou; Qinmei Wang

Purpose To evaluate the effect of cycloplegia on ocular biometry measurements and intraocular lens (IOL) power calculation using the Lenstar LS900 (Haag-Streit AG, Koeniz, Switzerland) and the IOLMaster (Carl Zeiss Meditec AG, Jena, Germany) biometers and to assess the agreement between the devices. Methods Measurements were taken with the Lenstar and the IOLMaster on 43 healthy volunteers with a mean age of 22.1 ± 4.7 years (range, 18 to 37 years). Axial length (AL), anterior chamber depth (ACD), corneal curvature, and horizontal iris width (white-to-white [WTW]) measurements were performed with and without cycloplegia. The IOL powers were calculated using four formulas: Sanders-Retzlaff-Kraff/Theoretical, Holladay 1, Hoffer Q, and Haigis. Results Cycloplegia had no significant effect on AL or corneal curvature. However, ACD and WTW significantly increased postcycloplegia (Lenstar, 0.09 ± 0.06 mm and 0.10 ± 0.17 mm, respectively; IOLMaster, 0.06 ± 0.07 mm and 0.43 ± 0.35 mm, respectively; p <0.001). The Lenstar AL measurements were statistically but not clinically significantly longer than those of the IOLMaster (precycloplegia, 0.03 ± 0.03 mm; postcycloplegia, 0.02 ± 0.03 mm; p < 0.001). For ACD measurements, the 95% limits of agreement were −0.19 to 0.20 mm without cycloplegia and −0.11 to 0.17 mm with cycloplegia. The 95% limits of agreement for WTW measurements were −1.07 to 0.45 mm with cycloplegia. The only significantly different IOL power precycloplegia and postcycloplegia was with the Haigis formula and the Lenstar measurements: 15.12 ± 3.87 diopters and 15.26 ± 3.92 diopters (p < 0.01). Conclusions Cycloplegia affected ACD and WTW but not AL or corneal curvature measurements. Generally, good agreement was found between the Lenstar and the IOLMaster, although not for WTW. Differences between these devices do not produce a clinically significant impact on IOL power.


Journal of Cataract and Refractive Surgery | 2013

Precision of a new Scheimpflug and Placido-disk analyzer in measuring corneal thickness and agreement with ultrasound pachymetry

Jinhai Huang; Giacomo Savini; Liang Hu; Kenneth J. Hoffer; Weicong Lu; Yifan Feng; Feng Yang; Xiuli Hu; Qinmei Wang

Purpose To assess the precision of corneal thickness measurements obtained by a new Scheimpflug camera combined with Placido‐disk corneal topography (Sirius) and compare the measured values with those obtained by ultrasound (US) pachymetry. Setting Eye Hospital of Wenzhou Medical College, Wenzhou, China. Design Comparative evaluation of a diagnostic test or technology. Methods Eyes of healthy subjects were examined with the Scheimpflug–Placido topographer. Central (CCT) and thinnest (TCT) corneal thickness were recorded after 3 consecutive measurements. For US pachymetry, only CCT was measured. Measurements were repeated within 1 week. The within‐subject standard deviation (Sw), test–retest repeatability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to evaluate intrasession repeatability and intersession reproducibility. Interdevice comparison was analyzed with paired t tests and Bland‐Altman plots. Results The intrasession repeatability of Scheimpflug–Placido measurements was high, with test–retest and CoV close to 9 μm and 0.6% for CCT and TCT, respectively. The intersession test–retest and CoV were close to 10 μm and 0.7%, respectively. The ICC was higher than 0.98 for repeatability and reproducibility. High agreement was found between Scheimpflug–Placido and US pachymetry measurements, with narrow 95% limits of agreement. Conclusions The Scheimpflug–Placido instrument showed excellent intrasession repeatability and intersession reproducibility of CCT and TCT measurements in healthy eyes. High agreement and lack of statistically significant difference suggest that the instruments TCT and the US pachymetry–CCT measurements can be used interchangeably in subjects with normal corneal thickness. Financial Disclosure No author has a financial or proprietary interest in any material or method mentioned.


Journal of Cataract and Refractive Surgery | 2013

Repeatability, reproducibility, and agreement of corneal power measurements obtained with a new corneal topographer

Xinjie Mao; Giacomo Savini; Zuopao Zhuo; Yifan Feng; Jian Zhang; Qinmei Wang; Hao Chen; Jinhai Huang

Purpose To study the repeatability and reproducibility of corneal power measurements obtained with a new corneal topographer (Keratograph 4) and assess their agreement with those obtained by a rotating Scheimpflug camera (Pentacam HR) and an automated keratometer (IOLMaster). Setting Eye Hospital, Wenzhou Medical University, Wenzhou, China. Design Observational cross‐sectional study. Methods One eye of normal subjects was measured 3 times using all devices. Another operator performed an additional 3 consecutive scans using the corneal topographer. One week later, the first operator repeated the corneal power measurements using the corneal topographer. Parameters were flat meridian power, steep meridian power, average power, J0, and J45. The repeatability and reproducibility of measurements were assessed by the within‐subject standard deviation (Sw), repeatability (2.77 Sw), coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement between devices was assessed using 95% limits of agreement (LoA). Results Intraoperator repeatability and interoperator and intersession reproducibility of all measured parameters showed a CoV of less than 0.3%, a 2.77 Sw of 0.32 diopter or less, and an ICC of more than 0.97. No significant differences in the parameters were noted between the corneal topographer and Scheimpflug camera. Statistically significant differences existed between the parameters analyzed by the corneal topographer and the automated keratometer, except J45. The mean differences between the corneal topographer and the other 2 devices were small, and the 95% LoA were narrow for all measurements. Conclusion The new corneal topographer had excellent reliability and high agreement with the other 2 devices in corneal power measurements in normal subjects. Financial Disclosure No author has a financial or proprietary interest in any material or method mentioned.


Clinical and Experimental Ophthalmology | 2013

Phacoemulsification versus manual small-incision cataract surgery for age-related cataract: meta-analysis of randomized controlled trials

Jia‐yu Zhang; Yifan Feng; Jian‐qiu Cai

To compare outcomes of phacoemulsification (PE) with manual small‐incision cataract surgery (MSICS) for age‐related cataract.


Journal of Cataract and Refractive Surgery | 2015

Anterior chamber depth measurements using Scheimpflug imaging and optical coherence tomography: Repeatability, reproducibility, and agreement

Qinmei Wang; Xixia Ding; Giacomo Savini; Hao Chen; Yifan Feng; Chao Pan; Yanjun Hua; Jinhai Huang

Purpose To determine the repeatability, reproducibility, and agreement of anterior chamber depth (ACD) measurements obtained with 3 Scheimpflug cameras and an anterior segment optical coherence tomography (AS‐OCT) device. Setting Eye Hospital of Wenzhou Medical University, Zhejiang, China. Design Observational cross‐sectional study. Methods Two observers took 3 consecutive measurements in healthy right eyes using each device to assess intraoperator repeatability. The mean values obtained at different sessions by the first operator were used to determine the intersession reproducibility. Three consecutive measurements obtained by the first operator at the first session were averaged and used to assess agreement. Results The ACD measurements obtained by 2 observers in 71 eyes were highly repeatable using the 4 devices, with a test–retest repeatability of 0.04 to 0.07 mm for intraoperator repeatability. The interoperator and intersession reproducibility of ACD measurements were high, and the test–retest repeatability ranges of interoperator and intersession reproducibility were 0.06 to 0.07 mm and 0.05 to 0.08 mm, respectively. The ACD measurements of the 4 systems were sorted from the thickest to the thinnest (Galilei G2 > Visante > Sirius > Pentacam). The differences in the measurements were statistically significant except between the ACD measurements obtained by the Sirius device and the Visante device. However, good agreement with narrow 95% limits of agreement was found between these devices. Conclusions The 4 devices provided high intraoperator repeatability and interoperator and intersession reproducibility for ACD measurements. Good agreement in ACD measurements was found between the devices in healthy eyes. Financial Disclosure No author has a financial or proprietary interest in any material or method mentioned.


PLOS ONE | 2014

Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography

Jinhai Huang; Xixia Ding; Giacomo Savini; Zhengxuan Jiang; Chao Pan; Yanjun Hua; Fan Wu; Yifan Feng; Ye Yu; Qinmei Wang

Purpose To compare corneal thickness measurements using Pentacam (Oculus, Germany), Sirius (CSO, Italy), Galilei (Ziemer, Switzerland), and RTVue-100 OCT (Optovue Inc., USA). Methods Sixty-six eyes of 66 healthy volunteers were enrolled. Three consecutive measurements were performed with each device. The mean value of the three measurements was used for subsequent analysis. Central corneal thickness (CCT), thinnest corneal thickness (TCT), and midperipheral corneal thickness (MPCT; measured at superior, inferior, nasal, and temporal locations with a distance of 1 mm (CT2mm) or 2.5 mm (CT5mm) from the corneal apex) were analyzed. Differences and agreement between measurements were assessed using the repeated-measures analysis of variance (ANOVA) and Bland-Altman analyses, respectively. Results Statistically significant differences (p<0.001) among the four devices were revealed in CCT, TCT and CT2mmmeasurements. The CCT, TCT, and CT2mm values were ranked from the thickest to the thinnest as follows: Galilei>Sirius>Pentacam>RTVue OCT. For these measurements, agreement between measurements by Sirius and Pentacam was good, whereas Galilei overestimated and RTVue underestimated corneal thickness compared to Sirius and Pentacam. As regards CT5mm measurements, Pentacam provided the largest values, whereas RTVue OCT yielded the smallest values. Agreement of the CT5mm measurements was good between the Pentacam, Sirius, moderate between Galilei and the other two Scheimpflug systems, and poor between the RTVue OCT and the remaining devices. Conclusions The Pentacam and Sirius can be used interchangeably for CCT measurements, while the Galilei and RTVue systematically over- and underestimate CCT, respectively. The three Scheimpflug cameras, but not the RTVue, may be used interchangeably for MPCT measurements.


PLOS ONE | 2015

Axial Length Measurement Failure Rates with the IOLMaster and Lenstar LS 900 in Eyes with Cataract

Colm McAlinden; Qinmei Wang; Konrad Pesudovs; Xin Yang; Fangjun Bao; A-Yong Yu; Shishi Lin; Yifan Feng; Jinhai Huang

Purpose To evaluate axial length (AL) measurement failure rate with the IOLMaster (Carl Zeiss AG, Germany) and Lenstar LS 900 (Haag-Streit AG, Switzerland) in eyes with cataract. Methods Two hundred and ninety-six eyes of 170 patients with cataract were enrolled. Cataract type and severity were graded using the Lens Opacities Classification System III (LOCS III) and AL measurements were attempted with IOLMaster (version 5.4) and Lenstar LS 900 (version 1.1). Chi-squared analysis was used to assess if the difference in AL measurement acquisition rate was statistically significant between the two devices. The association of the different cataract types and severity with the AL measurement acquisition rate was evaluated with logistic regression analysis. Results AL measurements were obtained in 184 eyes (62.16%) using the IOLMaster and 191 eyes (64.53%) using the Lenstar, which corresponds to a failure rate of 37.84% and 35.47% respectively. Chi-square analysis indicated no significant difference between the Lenstar and IOLMaster for AL measurement failure rate (x2 = 0.356, P = 0.550). Logistic regression analysis indicated no association between acquisition rates and cortical or nuclear cataracts with either device. There was a statistically significant association between acquisition rates and increasing severity of posterior subcapsular cataracts with the IOLMaster (β = -1.491, P<0.001) and Lenstar LS 900 (β = -1.507, P<0.001). Conclusion The IOLMaster and Lenstar LS 900 have similar AL measurement failure rates (35–38%) for Chinese public hospital cataract patients. Increasing severity of posterior subcapsular cataracts was problematic for both devices.

Collaboration


Dive into the Yifan Feng's collaboration.

Top Co-Authors

Avatar

Jinhai Huang

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar

Qinmei Wang

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Chao Pan

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar

Yanjun Hua

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Feng Yang

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar

Hao Chen

Wenzhou Medical College

View shared research outputs
Top Co-Authors

Avatar

Shishi Lin

Wenzhou Medical College

View shared research outputs
Researchain Logo
Decentralizing Knowledge