Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Alwina Bender is active.

Publication


Featured researches published by Alwina Bender.


Journal of Biomechanics | 2010

Loading of the knee joint during activities of daily living measured in vivo in five subjects

Ines Kutzner; Bernd Heinlein; Friedmar Graichen; Alwina Bender; A. Rohlmann; Andreas Halder; Alexander Beier; G. Bergmann

Detailed knowledge about loading of the knee joint is essential for preclinical testing of implants, validation of musculoskeletal models and biomechanical understanding of the knee joint. The contact forces and moments acting on the tibial component were therefore measured in 5 subjects in vivo by an instrumented knee implant during various activities of daily living. Average peak resultant forces, in percent of body weight, were highest during stair descending (346% BW), followed by stair ascending (316% BW), level walking (261% BW), one legged stance (259% BW), knee bending (253% BW), standing up (246% BW), sitting down (225% BW) and two legged stance (107% BW). Peak shear forces were about 10-20 times smaller than the axial force. Resultant forces acted almost vertically on the tibial plateau even during high flexion. Highest moments acted in the frontal plane with a typical peak to peak range -2.91% BWm (adduction moment) to 1.61% BWm (abduction moment) throughout all activities. Peak flexion/extension moments ranged between -0.44% BWm (extension moment) and 3.16% BWm (flexion moment). Peak external/internal torques lay between -1.1% BWm (internal torque) and 0.53% BWm (external torque). The knee joint is highly loaded during daily life. In general, resultant contact forces during dynamic activities were lower than the ones predicted by many mathematical models, but lay in a similar range as measured in vivo by others. Some of the observed load components were much higher than those currently applied when testing knee implants.


Bio-medical Materials and Engineering | 2010

Realistic loads for testing hip implants

G. Bergmann; Friedmar Graichen; A. Rohlmann; Alwina Bender; Bernd Heinlein; Georg N. Duda; Markus O. Heller; Michael M. Morlock

The aim here was to define realistic load conditions for hip implants, based on in vivo contact force measurements, and to see whether current ISO standards indeed simulate real loads. The load scenarios obtained are based on in vivo hip contact forces measured in 4 patients during different activities and on activity records from 31 patients. The load scenarios can be adapted to various test purposes by applying average or high peak loads, high-impact activities or additional low-impact activities, and by simulating normal or very active patients. The most strenuous activities are walking (average peak forces 1800 N, high peak forces 3900 N), going up stairs (average peak forces 1900 N, high peak forces 4200 N) and stumbling (high peak forces 11,000 N). Torsional moments are 50% higher for going up stairs than for walking. Ten million loading cycles simulate an implantation time of 3.9 years in active patients. The in vitro fatigue properties of cementless implant fixations are exceeded during stumbling. At least for heavyweight and very active subjects, the real load conditions are more critical than those defined by the ISO standards for fatigue tests.


PLOS ONE | 2014

Standardized Loads Acting in Knee Implants

G. Bergmann; Alwina Bender; Friedmar Graichen; Joern Dymke; A. Rohlmann; Adam Trepczynski; Markus O. Heller; Ines Kutzner

The loads acting in knee joints must be known for improving joint replacement, surgical procedures, physiotherapy, biomechanical computer simulations, and to advise patients with osteoarthritis or fractures about what activities to avoid. Such data would also allow verification of test standards for knee implants. This work analyzes data from 8 subjects with instrumented knee implants, which allowed measuring the contact forces and moments acting in the joint. The implants were powered inductively and the loads transmitted at radio frequency. The time courses of forces and moments during walking, stair climbing, and 6 more activities were averaged for subjects with I) average body weight and average load levels and II) high body weight and high load levels. During all investigated activities except jogging, the high force levels reached 3,372–4,218N. During slow jogging, they were up to 5,165N. The peak torque around the implant stem during walking was 10.5 Nm, which was higher than during all other activities including jogging. The transverse forces and the moments varied greatly between the subjects, especially during non-cyclic activities. The high load levels measured were mostly above those defined in the wear test ISO 14243. The loads defined in the ISO test standard should be adapted to the levels reported here. The new data will allow realistic investigations and improvements of joint replacement, surgical procedures for tendon repair, treatment of fractures, and others. Computer models of the load conditions in the lower extremities will become more realistic if the new data is used as a gold standard. However, due to the extreme individual variations of some load components, even the reported average load profiles can most likely not explain every failure of an implant or a surgical procedure.


Journal of Biomechanics | 2007

In vivo measurement of shoulder joint loads during activities of daily living

P. Westerhoff; Friedmar Graichen; Alwina Bender; Andreas Halder; Alexander Beier; A. Rohlmann; G. Bergmann

Until recently the contact loads acting in the glenohumeral joint have been calculated using musculoskeletal models or measured in vitro. Now, contact forces and moments are measured in vivo using telemeterized shoulder implants. Mean total contact forces from four patients during eight activities of daily living are reported here. Lifting a coffee pot (1.5kg) with straight arm caused an average force of 105.0%BW (%body weight) (range: 90-124.6%BW), while setting down the coffee pot in the same position led to higher forces of 122.9%BW on the average (105.3-153.4%BW). The highest joint contact forces were measured when the straight arm was abducted or elevated by 90 degrees or more, with a weight in the hand. Lifting up 2kg from a board up to head height caused a contact force of 98.3%BW (93-103.6%BW); again, setting it down on the board led to higher forces of 131.5%BW (118.8-144.1%BW). In contrast to previously calculated high loads, the contact force during passive holding of a 10kg weight laterally was only 12.3%BW (9.2-17.9%BW), but when lifting it up to belt height it increased to 91.5%BW (87-95%BW). The moments transferred inside the joint at our patients varied much more than did the forces both inter and intra-individually. Our data suggest that patients with shoulder problems or during the first post-operative weeks after shoulder fractures or joint replacements should avoid certain activities encountered during daily living e.g. lifting or holding a weight with an outstretched arm. Some energy-related optimization criteria used in the literature for analytical musculoskeletal shoulder models must now be reconsidered.


Journal of Biomechanics | 2011

In vivo gleno-humeral joint loads during forward flexion and abduction.

G. Bergmann; Friedmar Graichen; Alwina Bender; A. Rohlmann; Andreas Halder; Alexander Beier; P. Westerhoff

To improve design and preclinical test scenarios of shoulder joint implants as well as computer-based musculoskeletal models, a precise knowledge of realistic loads acting in vivo is necessary. Such data are also helpful to optimize physiotherapy after joint replacement and fractures. This is the first study that presents forces and moments measured in vivo in the gleno-humeral joint of 6 patients during forward flexion and abduction of the straight arm. The peak forces and, even more, the maximum moments varied inter-individually to a considerable extent. Forces of up to 238%BW (percent of body weight) and moments up to 1.74%BWm were determined. For elevation angles of less than 90° the forces agreed with many previous model-based calculations. At higher elevation angles, however, the measured loads still rose in contrast to the analytical results. When the exercises were performed at a higher speed, the peak forces decreased. The force directions relative to the humerus remained quite constant throughout the whole motion. Large moments in the joint indicate that friction in shoulder implants is high if the glenoid is not replaced. A friction coefficient of 0.1-0.2 seems to be realistic in these cases.


Journal of Biomechanics | 2011

The effect of valgus braces on medial compartment load of the knee joint – in vivo load measurements in three subjects

Ines Kutzner; Steffen Küther; Bernd Heinlein; Jörn Dymke; Alwina Bender; Andreas Halder; G. Bergmann

Knee osteoarthritis occurs predominately at the medial compartment. To unload the affected compartment, valgus braces are used which induce an additional valgus moment in order to shift the load more laterally. Until now the biomechanical effect of braces was mainly evaluated by measuring changes in external knee adduction moments. The aim of this study was to investigate if and to which extent the medial compartment load is reduced in vivo when wearing valgus braces. Six components of joint contact load were measured in vivo in three subjects, using instrumented, telemeterized knee implants. From the forces and moments the medio-lateral force distribution was calculated. Two braces, MOS Genu (Bauerfeind AG) and Genu Arthro (Otto Bock) were investigated in neutral, 4° and 8° valgus adjustment during walking, stair ascending and descending. During walking with the MOS brace in 4°/8° valgus adjustment, medial forces were reduced by 24%/30% on average at terminal stance. During walking with the GA in the 8° valgus position, medial forces were reduced by only 7%. During stair ascending/descending significant reductions of 26%/24% were only observed with the MOS (8°). The load reducing ability of the two investigated valgus braces was confirmed in three subjects. However, the load reduction depends on the brace stiffness and its valgus adjustment and varies strongly inter-individually. Valgus adjustments of 8° might, especially with the MOS brace, not be tolerated by patients for a long time. Medial load reductions of more than 25% can therefore probably not be expected in clinical practise.


Medical Engineering & Physics | 2009

An instrumented implant for in vivo measurement of contact forces and contact moments in the shoulder joint.

P. Westerhoff; Friedmar Graichen; Alwina Bender; A. Rohlmann; G. Bergmann

To improve implant design, fixation and preclinical testing, implant manufacturers depend on realistic data of loads acting on the shoulder joint. Furthermore, these data can help to optimize physiotherapeutic treatment and to advise patients in their everyday living conditions. Calculated shoulder joint loads vary extremely among different authors [Anglin C, Wyss UP, Pichora DR. Glenohumeral contact forces. Proc Inst Mech Eng [H] 2000;214:637-44]. Additionally the moments acting in the joint caused by friction or incongruent articular surfaces, for example, are not implemented in most models. An instrumented shoulder joint implant was developed to measure the contact forces and the contact moments acting in the glenohumeral joint. This article provides a detailed description of the implant, containing a nine-channel telemetry unit, six load sensors and an inductive power supply, all hermetically sealed inside the implant. The instrumented implant is based on a clinically proven BIOMET Biomodular shoulder replacement and was calibrated before implantation by using complex mathematical calculation routines in order to achieve an average measuring precision of approximately 2%.


Spine | 2008

Loads on a telemeterized vertebral body replacement measured in two patients.

A. Rohlmann; F. Graichen; Ralph Kayser; Alwina Bender; G. Bergmann

Study Design. The loads acting on a vertebral body replacement (VBR) were measured in vivo. Objective. To measure the implant loads for different activities within the first 6 months after surgery. Summary of Background Data. Mathematical models exist for predicting spinal loads for various activities. The intradiscal pressure has been measured in vivo for many activities. Loads on internal spinal fixation devices have been measured in 10 patients. However, only little information exists regarding the loads acting on a VBR. Methods. Telemeterized VBRs were implanted into 2 patients with a fractured L1 vertebral body. The implant allows the in vivo measurement of 3 force and 3 moment components acting on the implant. For several activities, implant loads were measured in the first 6 months after surgery. Results. One month after surgery, the resultant force during standing was about 270 N in 1 patient and 300 N in the other. When the patients were lying in relaxed positions, resultant forces were less than 30% of the values during standing. In one patient, implant loads were slightly lower during sitting than during standing whereas in the other patient higher loads were measured during sitting. In both patients, flexion of the upper body and walking upstairs caused implant loads, which were more than twice as high as those during standing. Force direction varied only slightly for forces higher than 100 N. Conclusion. High forces may act on a VBR especially in the first postoperative month. Flexion of the upper body and going upstairs cause high implant loads and should be avoided in the first few months after stabilizing the spine.


Medical Engineering & Physics | 2010

Total hip joint prosthesis for in vivo measurement of forces and moments.

Philipp Damm; Friedmar Graichen; A. Rohlmann; Alwina Bender; G. Bergmann

A new instrumented hip joint prosthesis was developed which allows the in vivo measurement of the complete contact loads in the joint, i.e. 3 force and 3 moment components. A clinically proven standard implant was modified. Inside the hollow neck, 6 semiconductor strain gauges are applied to measure the deformation of the neck. Also integrated are a small coil for the inductive power supply and a 9-channel telemetry transmitter. The neck cavity is closed by a titanium plate and hermetically sealed by electron beam welding. The sensor signals are pulse interval modulated (PIM) with a sampling rate of about 120 Hz. The pulses are transmitted at radio frequencies via a small antenna loop inside the ceramic head, which is connected to the electronic circuit by a two-pin feedthrough. Inductive power supply, calculation of the loads from the measured deformations and real time load display are carried out by the external equipment. The maximum error of the load components is 2% including crosstalk.


PLOS ONE | 2013

Friction in total hip joint prosthesis measured in vivo during walking.

Philipp Damm; Joern Dymke; Robert Ackermann; Alwina Bender; Friedmar Graichen; Andreas Halder; Alexander Beier; G. Bergmann

Friction-induced moments and subsequent cup loosening can be the reason for total hip joint replacement failure. The aim of this study was to measure the in vivo contact forces and friction moments during walking. Instrumented hip implants with Al2O3 ceramic head and an XPE inlay were used. In vivo measurements were taken 3 months post operatively in 8 subjects. The coefficient of friction was calculated in 3D throughout the whole gait cycle, and average values of the friction-induced power dissipation in the joint were determined. On average, peak contact forces of 248% of the bodyweight and peak friction moments of 0.26% bodyweight times meter were determined. However, contact forces and friction moments varied greatly between individuals. The friction moment increased during the extension phase of the joint. The average coefficient of friction also increased during this period, from 0.04 (0.03 to 0.06) at contralateral toe off to 0.06 (0.04 to 0.08) at contralateral heel strike. During the flexion phase, the coefficient of friction increased further to 0.14 (0.09 to 0.23) at toe off. The average friction-induced power throughout the whole gait cycle was 2.3 W (1.4 W to 3.8 W). Although more parameters than only the synovia determine the friction, the wide ranges of friction coefficients and power dissipation indicate that the lubricating properties of synovia are individually very different. However, such differences may also exist in natural joints and may influence the progression of arthrosis. Furthermore, subjects with very high power dissipation may be at risk of thermally induced implant loosening. The large increase of the friction coefficient during each step could be caused by the synovia being squeezed out under load.

Collaboration


Dive into the Alwina Bender's collaboration.

Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

F. Graichen

Free University of Berlin

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Researchain Logo
Decentralizing Knowledge