Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where B.J. Mijnheer is active.

Publication


Featured researches published by B.J. Mijnheer.


Medical Physics | 2003

Dose–response and ghosting effects of an amorphous silicon electronic portal imaging device

Leah N. McDermott; R. J. W. Louwe; J.J. Sonke; M. van Herk; B.J. Mijnheer

The purpose of this study was to investigate the dose-response characteristics, including ghosting effects, of an amorphous silicon-based electronic portal imaging device (a-Si EPID) under clinical conditions. EPID measurements were performed using one prototype and two commercial a-Si detectors on two linear accelerators: one with 4 and 6 MV and the other with 8 and 18 MV x-ray beams. First, the EPID signal and ionization chamber measurements in a mini-phantom were compared to determine the amount of buildup required for EPID dosimetry. Subsequently, EPID signal characteristics were studied as a function of dose per pulse, pulse repetition frequency (PRF) and total dose, as well as the effects of ghosting. There was an over-response of the EPID signal compared to the ionization chamber of up to 18%, with no additional buildup layer over an air gap range of 10 to 60 cm. The addition of a 2.5 mm thick copper plate sufficiently reduced this over-response to within 1% at clinically relevant patient-detector air gaps (> 40 cm). The response of the EPIDs varied by up to 8% over a large range of dose per pulse values, PRF values and number of monitor units. The EPID response showed an under-response at shorter beam times due to ghosting effects, which depended on the number of exposure frames for a fixed frame acquisition rate. With an appropriate build-up layer and corrections for dose per pulse, PRF and ghosting, the variation in the a-Si EPID response can be reduced to well within +/- 1%.


Medical Physics | 2010

Catching errors with in vivo EPID dosimetry

A. Mans; Markus Wendling; Leah N. McDermott; J.J. Sonke; R. Tielenburg; R.E. Vijlbrief; B.J. Mijnheer; M. van Herk; J. Stroom

The potential for detrimental incidents and the ever increasing complexity of patient treatments emphasize the need for accurate dosimetric verification in radiotherapy. For this reason, all curative treatments are verified, either pretreatment or in vivo, by electronic portal imaging device (EPID) dosimetry in the Radiation Oncology Department of the Netherlands Cancer Institute-Antoni van Leeuwenhoek hospital, Amsterdam, The Netherlands. Since the clinical introduction of the method in January 2005 until August 2009, treatment plans of 4337 patients have been verified. Among these plans, 17 serious errors were detected that led to intervention. Due to their origin, nine of these errors would not have been detected with pretreatment verification. The method is illustrated in detail by the case of a plan transfer error detected in a 5×5Gy intensity-modulated radiotherapy (IMRT) rectum treatment. The EPID reconstructed dose at the isocenter was 6.3% below the planned value. Investigation of the plan transfer chain revealed that due to a network transfer error, the plan was corrupted. 3D analysis of the acquired EPID data revealed serious underdosage of the planning target volume: On average 11.6%, locally up to 20%. This report shows the importance of in vivo (EPID) dosimetry for all treatment plans as well as the ability of the method to assess the dosimetric impact of deviations found.


Radiotherapy and Oncology | 1991

Accuracy in tangential breast treatment set-up: a portal imaging study

G. van Tienhoven; J.H. Lanson; D. Crabeels; S. Heukelom; B.J. Mijnheer

To test the accuracy and reproducibility of the tangential breast treatment set-up used in The Netherlands Cancer Institute, a portal imaging study was performed in 12 patients treated for early stage breast cancer. With an on-line electronic portal imaging device (EPID) images were obtained of each patient in several fractions and compared with simulator films and with each other. In five patients, multiple images (on the average 7) per fraction were obtained to evaluate set-up variations due to respiratory movement. The central lung distance (CLD) and other set-up parameters varied within one fraction about 1 mm (1 SD). The average variation of these parameters between various fractions was about 2 mm (1 SD). The differences between simulator and treatment set-up over all patients and all fractions was on the average 2-3 mm for the central beam edge to skin distance and the central lung distance. It can be concluded that the tangential breast treatment set-up is very stable and reproducible and that respiration does not have a significant influence on treatment volume. The EPID appears to be an adequate tool for studies of treatment set-up accuracy like this.


Physics in Medicine and Biology | 1991

Comparison of entrance and exit dose measurements using ionization chambers and silicon diodes

S. Heukelom; J.H. Lanson; B.J. Mijnheer

A high precision patient dosimetry method has been developed, based on the use of p-type diodes. First, entrance as well as exit dose calibration factors have to be determined under reference irradiation conditions. Secondly, a set of correction factors must be added for situations deviating from the reference conditions, i.e. for different source-skin distances, phantom (patient) thicknesses, field sizes or for insertion of a wedge into the photon beam. Finally some other detector characteristics such as the temperature dependence of the response have to be taken into account. For most irradiation conditions this procedure is sufficiently accurate to allow entrance as well as exist dose determinations with a diode to be in good agreement with dose values measured by an ionization chamber. The main factors effecting the value of the correction factors, the dependence of the diode sensitivity on the energy and the dose per pulse, have been investigated to explain some of the observed phenomena. Despite a strong energy dependence of the sensitivity, the correction factors are, for a particular type of diode, the same for 4 and 8 MV x-ray beams. The variation in the values for the correction factors with integrated dose received by the diode is small. These findings indicate that the correction factors, once available, can be applied under a number of circumstances. Due to the difference in behaviour of various diodes, even from the same batch, it is, however, necessary to determine the characteristics for each diode individually.


Medical Physics | 2006

Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification

Leah N. McDermott; Markus Wendling; B. van Asselen; J. Stroom; J.J. Sonke; M. van Herk; B.J. Mijnheer

The aim of this study was to demonstrate how dosimetry with an amorphous silicon electronic portal imaging device (a-Si EPID) replaced film and ionization chamber measurements for routine pre-treatment dosimetry in our clinic. Furthermore, we described how EPID dosimetry was used to solve a clinical problem. IMRT prostate plans were delivered to a homogeneous slab phantom. EPID transit images were acquired for each segment. A previously developed in-house back-projection algorithm was used to reconstruct the dose distribution in the phantom mid-plane (intersecting the isocenter). Segment dose images were summed to obtain an EPID mid-plane dose image for each field. Fields were compared using profiles and in two dimensions with the y evaluation (criteria: 3%/3 mm). To quantify results, the average gamma (gamma avg), maximum gamma (gamma max), and the percentage of points with gamma < 1(P gamma < 1) were calculated within the 20% isodose line of each field. For 10 patient plans, all fields were measured with EPID and film at gantry set to 0 degrees. The film was located in the phantom coronal mid-plane (10 cm depth), and compared with the back-projected EPID mid-plane absolute dose. EPID and film measurements agreed well for all 50 fields, with (gamma avg) =0.16, (gamma max)=1.00, and (P gamma < 1)= 100%. Based on these results, film measurements were discontinued for verification of prostate IMRT plans. For 20 patient plans, the dose distribution was re-calculated with the phantom CT scan and delivered to the phantom with the original gantry angles. The planned isocenter dose (plan(iso)) was verified with the EPID (EPID(iso)) and an ionization chamber (IC(iso)). The average ratio, (EPID(iso)/IC(iso)), was 1.00 (0.01 SD). Both measurements were systematically lower than planned, with (EPID(iso)/plan(iso)) and (IC(iso)/plan(iso))=0.99 (0.01 SD). EPID mid-plane dose images for each field were also compared with the corresponding plane derived from the three dimensional (3D) dose grid calculated with the phantom CT scan. Comparisons of 100 fields yielded (gamma avg)=0.39, gamma max=2.52, and (P gamma < 1)=98.7%. Seven plans revealed under-dosage in individual fields ranging from 5% to 16%, occurring at small regions of overlapping segments or along the junction of abutting segments (tongue-and-groove side). Test fields were designed to simulate errors and gave similar results. The agreement was improved after adjusting an incorrectly set tongue-and-groove width parameter in the treatment planning system (TPS), reducing (gamma max) from 2.19 to 0.80 for the test field. Mid-plane dose distributions determined with the EPID were consistent with film measurements in a slab phantom for all IMRT fields. Isocenter doses of the total plan measured with an EPID and an ionization chamber also agreed. The EPID can therefore replace these dosimetry devices for field-by-field and isocenter IMRT pre-treatment verification. Systematic errors were detected using EPID dosimetry, resulting in the adjustment of a TPS parameter and alteration of two clinical patient plans. One set of EPID measurements (i.e., one open and transit image acquired for each segment of the plan) is sufficient to check each IMRT plan field-by-field and at the isocenter, making it a useful, efficient, and accurate dosimetric tool.


Radiotherapy and Oncology | 1991

Quality assurance in conservative treatment of early breast cancer. Report on a consensus meeting of the EORTC Radiotherapy and Breast Cancer Cooperative Groups and the EUSOMA (European Society of Mastology).

Harry Bartelink; G. Garavaglia; Karl-Axel Johansson; B.J. Mijnheer; W. Van den Bogaert; G. van Tienhoven; John Yarnold

A consensus on a quality assurance programme of the treatment of early breast cancer was reached in a multidisciplinary meeting of surgeons, pathologists, radiotherapists, physicists and radiographers. Guidelines for treatment preparation and execution have been set up, including careful location and excision with marking of the primary tumour. The target volumes for irradiation of the whole breast and boost area have been defined. Radiation dose prescription rules, specification and checking procedures are given, together with measures to achieve a homogeneous dose within the target volume. The rules for a quality assurance programme in each clinic are designed for checking equipment and treatment method.


Medical Physics | 2003

Three‐dimensional dose reconstruction of breast cancer treatment using portal imaging

R. J. W. Louwe; E. Damen; M. van Herk; A. Minken; O. Törzsök; B.J. Mijnheer

In this study, we present an algorithm for three-dimensional (3-D) dose reconstruction using portal images obtained with an electronic portal imaging device (EPID). For this purpose an algorithm for 2-D dose reconstruction, which was previously developed in our institution, was adapted. The external contour of the patient was used to correct for absorption of primary photons, but the presence of inhomogeneities was not taken into account. The accuracy of the algorithm was determined by irradiating two anthropomorphic breast phantoms with 6 MV photons. The dose values derived from portal images were compared with results from 3-D dose calculations, which, in turn, were verified with data obtained with an ionization chamber and film dosimetry. It was found that the application of contour information significantly improves the accuracy of 2-D dose reconstruction. If the total dose at the isocenter plane resulting from all treatment beams is reconstructed, the average deviation from the planned dose is 0.1%+/-1.7% (1 SD). If contour information is not available, the differences increase up to +/-20% for the individual beams. In that case, the dose can only be reconstructed with reasonable accuracy when (nearly) opposing beams are used. The average deviation of the 3-D reconstructed dose from the planned dose in the irradiated volume is 1.4%+/-5.4% (1 SD). If the irradiated volume is enclosed by planes less than 5 cm distant from the isocenter plane, then the average deviation is only 0.5%+/-3.4% (1 SD). It can be concluded that the proposed algorithm for a 3-D dose reconstruction allows a determination of the dose at the isocenter plane and the dose-volume histogram with an accuracy acceptable for an independent verification of the treatment.


Medical Physics | 2006

Comparison of ghosting effects for three commercial a-Si EPIDs

Leah N. McDermott; S. Nijsten; J.J. Sonke; Mike Partridge; M. van Herk; B.J. Mijnheer

Many studies have reported dosimetric characteristics of amorphous silicon electronic portal imaging devices (EPIDs). Some studies ascribed a non-linear signal to gain ghosting and image lag. Other reports, however, state the effect is negligible. This study compares the signal-to-monitor unit (MU) ratio for three different brands of EPID systems. The signal was measured for a wide range of monitor units (5-1000), dose-rates, and beam energies. All EPIDs exhibited a relative under-response for beams of few MUs; giving 4 to 10% lower signal-to-MU ratios relative to that of 1000 MUs. This under-response is consistent with ghosting effects due to charge trapping.


Radiotherapy and Oncology | 1992

In vivo dosimetry during pelvic treatment

S. Heukelom; J.H. Lanson; B.J. Mijnheer

High precision in vivo entrance and exit dose measurements have been performed with p-type diodes on patients during 8 MV X-ray irradiation of the pelvis, to investigate the accuracy of dose calculations in this region. Based on phantom measurements the accuracy of the p-type diode measuring system itself, i.e. the agreement with ionisation chamber dose measurements, was shown to be better than 0.7% while the reproducibility in the dose determination was 1.1%, 1.5% and 1.6% (1 S.D.) at the entrance point, isocentre and exit point, respectively, for the wedged lateral fields. Patient movement and the uncertainty in the diode position increased these values to 1.7%, 1.5% and 3.1% (1 S.D.) for dose determinations on patients. From the entrance and exit in vivo dose values the dose actually delivered to the isocentre was determined. For the anterior-posterior beams a good correspondence for most patients was observed at the entrance and exit point and at the isocentre between the in vivo and calculated dose values. For the wedged lateral beams a systematic deviation of about 3% was observed. In addition to the in vivo dose measurements phantom dose measurements have been performed to quantify the accuracy of the dose calculation algorithms including the computation of the number of monitor units. These measurements also served to quantify the effects of the actual patient on the dose delivery. The measurements showed that accurate calculation of the dose requires a separation of the head and phantom scatter contribution of the output of the treatment machine. The dependence of the wedge factor on field size, depth and source-skin-distance has also to be considered for accurate dose calculations. The effect of the patient on the dose calculation is mainly related to the actual electron densities of fat and bone structures compared to water: neglecting these densities in the dose computation could yield deviations up to 8.5% for the exit point in wedged beams. Based on these results, improvements in the dose calculation algorithms and monitor unit calculation including the use of the actual electron densities will be implemented in the treatment planning procedure.


Radiotherapy and Oncology | 1991

Quality assurance of the EORTC trial 22881/10882: “assessment of the role of the booster dose in breast conserving therapy”: the Dummy Run

G. van Tienhoven; N.A.M. van Bree; B.J. Mijnheer; Harry Bartelink

The EORTC trial 22881/10882 is a randomised trial with the aim to assess the role of the boost dose in breast conserving therapy in stage I and II breast cancer. In order to detect potential protocol deviations concerning irradiation technique and in the dose specification procedure of participating institutions before actual patient accrual, a Dummy Run was performed. Three transverse sections of a patient were sent to 16 participating institutions with a request to make a three-plane treatment plan according to the protocol prescriptions. A treatment chart and beam data were also requested for recalculation of the dose. Additional information was asked in a questionnaire. On evaluation, the techniques differed considerably with respect to photon beam energy, varying between 60Co gamma-rays and 8 MV X-rays, and the use of wedge filters. Two institutions did not apply wedges, whereas wedge angles in the other institutions varied between 6 degrees and 45 degrees. Twelve institutions used collimator rotation and/or a table wedge to diminish the amount of irradiated lung volume. The dose was specified in a point according to the protocol prescription in 11 institutions and to the 90, 95 or 100% isodose curve in four. Twelve institutions applied lung density corrections during treatment planning, while nine reported problems with their planning system in off-axis dose distribution calculation and/or the simulation of collimator rotation. Recalculation of the dose at the isocentre showed agreement within 2% compared with the stated dose. The dose reported in the tumour excision area varied between 93 and 100%.(ABSTRACT TRUNCATED AT 250 WORDS)

Collaboration


Dive into the B.J. Mijnheer's collaboration.

Top Co-Authors

Avatar

M. van Herk

Netherlands Cancer Institute

View shared research outputs
Top Co-Authors

Avatar

A. Mans

Netherlands Cancer Institute

View shared research outputs
Top Co-Authors

Avatar

R. Rozendaal

Netherlands Cancer Institute

View shared research outputs
Top Co-Authors

Avatar

I. Olaciregui-Ruiz

Netherlands Cancer Institute

View shared research outputs
Top Co-Authors

Avatar

J.J. Sonke

Netherlands Cancer Institute

View shared research outputs
Top Co-Authors

Avatar

J.H. Lanson

Netherlands Cancer Institute

View shared research outputs
Top Co-Authors

Avatar

J. Stroom

Netherlands Cancer Institute

View shared research outputs
Top Co-Authors

Avatar

Markus Wendling

Netherlands Cancer Institute

View shared research outputs
Top Co-Authors

Avatar

H. Spreeuw

Netherlands Cancer Institute

View shared research outputs
Top Co-Authors

Avatar

Leah N. McDermott

Netherlands Cancer Institute

View shared research outputs
Researchain Logo
Decentralizing Knowledge