Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Marcel van Straten is active.

Publication


Featured researches published by Marcel van Straten.


Medical Physics | 2009

Application- and patient size-dependent optimization of x-ray spectra for CT

Willi A. Kalender; Paul Deak; Markus Kellermeier; Marcel van Straten; Sabrina V. Vollmar

Although x-ray computed tomography (CT) has been in clinical use for over 3 decades, spectral optimization has not been a topic of great concern; high voltages around 120 kV have been in use since the beginning of CT. It is the purpose of this study to analyze, in a rigorous manner, the energies at which the patient dose necessary to provide a given contrast-to-noise ratio (CNR) for various diagnostic tasks can be minimized. The authors used cylindrical water phantoms and quasianthropomorphic phantoms of the thorax and the abdomen with inserts of 13 mm diameter mimicking soft tissue, bone, and iodine for simulations and measurements. To provide clearly defined contrasts, these inserts were made of solid water with a 1% difference in density (DD) to represent an energy-independent soft-tissue contrast of 10 Hounsfield units (HU), calcium hydroxyapatite (Ca) representing bone, and iodine (I) representing the typical contrast medium. To evaluate CT of the thorax, an adult thorax phantom (300 x 200 mm2) plus extension rings up to a size of 460 x 300 mm2 to mimic different patient cross sections were used. For CT of the abdomen, we used a phantom of 360 x 200 mm2 and an extension ring of 460 x 300 mm2. The CT scanner that the authors used was a SOMATOM Definition (Siemens Healthcare, Forchheim, Germany) at 80, 100, 120, and 140 kV. Further voltage settings of 60, 75, 90, and 105 kV were available in an experimental mode. The authors determined contrast for the density difference, calcium, and iodine, and noise and 3D dose distributions for the available voltages by measurements. Additional voltage values and monoenergetic sources were evaluated by simulations. The dose-weighted contrast-to-noise ratio (CNRD) was used as the parameter for optimization. Simulations and measurements were in good agreement with respect to absolute values and trends regarding the dependence on energy for the parameters investigated. For soft-tissue imaging, the standard settings of 120-140 kV were found as adequate choices with optimal values increasing for larger cross sections, e.g., for large abdomens voltages higher than 140 kV may be indicated. For bone and iodine imaging the optimum values were generally found at significantly lower voltages of typically below 80 kV. This offers a potential for dose reduction of up to 50%, but demands significantly higher power values in most cases. The authors concluded that voltage settings in CT should be varied more often than is common in practice today and should be chosen not only according to patient size but also according to the substance imaged in order to minimize dose while not compromising image quality. A reduction from 120 to 80 kV, for example, would yield a reduction in patient dose by more than half for coronary CT angiography. The use of lower voltages has to be recommended for contrast medium studies in cardiac and pediatric CT.


European Radiology | 2008

Validation of a Monte Carlo tool for patient-specific dose simulations in multi-slice computed tomography

Paul Deak; Marcel van Straten; Paul C. Shrimpton; Maria Zankl; Willi A. Kalender

Estimating the dose delivered to the patient in X-ray computed tomography (CT) examinations is not a trivial task. Monte Carlo (MC) methods appear to be the method of choice to assess the 3D dose distribution. The purpose of this work was to extend an existing MC-based tool to account for arbitrary scanners and scan protocols such as multi-slice CT (MSCT) scanners and to validate the tool in homogeneous and heterogeneous phantoms. The tool was validated by measurements on MSCT scanners for different scan protocols under known conditions. Quantitative CT Dose Index (CTDI) measurements were performed in cylindrical CTDI phantoms and in anthropomorphic thorax phantoms of various sizes; dose profiles were measured with thermoluminescent dosimeters (TLD) in the CTDI phantoms and compared with the computed dose profiles. The in-plane dose distributions were simulated and compared with TLD measurements in an Alderson-Rando phantom. The calculated dose values were generally within 10% of measurements for all phantoms and all investigated conditions. Three-dimensional dose distributions can be accurately calculated with the MC tool for arbitrary scanners and protocols including tube current modulation schemes. The use of the tool has meanwhile also been extended to further scanners and to flat-detector CT.


Radiology | 2009

Impact of Heart Rate Frequency and Variability on Radiation Exposure, Image Quality, and Diagnostic Performance in Dual-Source Spiral CT Coronary Angiography

Annick C. Weustink; Lisanne A. Neefjes; Stamatis Kyrzopoulos; Marcel van Straten; Rick Neoh Eu; Willem B. Meijboom; Carlos Van Mieghem; Ermanno Capuano; Marcel L. Dijkshoorn; Filippo Cademartiri; Eric Boersma; Pim J. de Feyter; Gabriel P. Krestin; Nico R. Mollet

PURPOSE To investigate the effect of heart rate frequency (HRF) and heart rate variability (HRV) on radiation exposure, image quality, and diagnostic performance to help detect significant stenosis (> or =50% lumen diameter reduction) by using adaptive electrocardiographic (ECG) pulsing at dual-source (DS) spiral computed tomographic (CT) coronary angiography. MATERIALS AND METHODS Institutional review committee approval and informed consent were obtained. No prescan beta-blockers were applied. Unenhanced CT and CT coronary angiography with adaptive ECG pulsing were performed in 927 consecutive patients (600 men, 327 women; mean age, 60.3 years +/- 11.0 [standard deviation]) divided in three HRF groups: low, intermediate, and high (< or =65, 66-79, and > or =80 beats/min, respectively), and four HRV groups given mean interbeat difference (IBD) during CT coronary angiography: normal, minor, moderate, and severe (IBDs of 0-1, 2-3, 4-10, and >10, respectively). Radiation exposure and image quality were also evaluated. In 444 of these, diagnostic performance was presented as sensitivity, specificity, positive predictive values (PPVs), and negative predictive values and likelihood ratios with corresponding 95% confidence intervals by using quantitative coronary angiography as the reference standard. RESULTS CT coronary angiography yielded good image quality in 98% of patients and no significant differences in image quality were found among HRF and HRV groups. Radiation exposure was significantly higher in patients with low versus high HRF and in patients with severe versus normal HRV. No significant differences among HRF and HRV groups in image quality and diagnostic performance were found. A nonsignificant trend was found toward a lower specificity and PPV in patients with a high HRF or severe HRV when compared with low HRF or normal HRV in patients with a low calcium score (Agatston score <100). CONCLUSION DS spiral CT coronary angiography performed with adaptive ECG pulsing results in preserved diagnostic image quality and performance independent of HRF or HRV at the cost of limited dose reduction in arrhythmic patients.


Medical Physics | 2009

The effect of angular and longitudinal tube current modulations on the estimation of organ and effective doses in x-ray computed tomography

Marcel van Straten; Paul Deak; Paul C. Shrimpton; Willi A. Kalender

PURPOSE Tube current modulation (TCM) is one of the recent developments in multislice CT that has proven to reduce the patient radiation dose without affecting the image quality. Presently established methods and published coefficients for estimating organ doses from the dose measured free in air on the axis of rotation or in the CT dose index (CTDI) dosimetry phantoms do not take into account this relatively new development in CT scanner design and technology. Based on these organ dose coefficients effective dose estimates can be made. The estimates are not strictly valid for CT scanning protocols utilizing TCM. In this study, the authors investigated the need to take TCM into account when estimating organ and effective dose values. METHODS A whole-body adult anthropomorphic phantom (Alderson Rando) was scanned with a multislice CT scanner (Somatom Definition, Siemens, Forchheim, Germany) utilizing TCM (CareDose4D). Tube voltage was 120 kV, beam collimation 19.2 mm, and pitch 1. A voxelized patient model was used to define the tissues and organs in the phantom. Tube current values as a function of tube angle were obtained from the raw data for each individual tube rotation of the scan. These values were used together with the Monte Carlo dosimetry tool IMPACTMC (VAMP GmbH, Erlangen, Germany) to calculate organ dose values both with and without account of TCM. Angular and longitudinal modulations were investigated separately. Finally, corresponding effective dose conversion coefficients were determined for both cases according to the updated 2007 recommendations of the ICRP. RESULTS TCM amplitude was greatest in the shoulder and pelvic regions. Consequently, dose distributions and organ dose values for particular cross sections changed considerably when taking angular modulation into account. The effective dose conversion coefficients were up to 11% lower for a single rotation in the shoulder region and 17% lower in the pelvis when taking angular TCM into account. In the head, neck, thorax, and upper abdominal regions, conversion coefficients changed similarly by only 5% or less. Conversion coefficients for estimating effective doses for scans of complete regions, e.g., chest or abdomen, were approximately 8% lower when taking angular and longitudinal TCMs into account. CONCLUSIONS The authors conclude that for accurate organ and effective dose estimates in individual cross sections in the shoulder or pelvic regions, the angular tube current modulation should be taken into account. In general, using the average of the modulated tube current causes an overestimation of the effective dose.


Medical Physics | 2004

Removal of bone in CT angiography of the cervical arteries by piecewise matched mask bone elimination

Marcel van Straten; Henk W. Venema; Geert J. Streekstra; Charles B. L. M. Majoie; Gerard J. den Heeten; Cornelis A. Grimbergen

In maximum intensity projection (MIP) images of CT angiography (CTA) scans, the arteries are often obscured by bone. A bone removal method is presented that uses an additional, nonenhanced scan to create a mask of the bone by thresholding and dilation. After registration of the CTA scan and the additional scan, the bone in the CTA scan is masked. As the cervical area contains bones that can move with respect to each other, these bones are separated first using a watershed algorithm, and then registered individually. A phantom study was performed to evaluate and quantify the tradeoff between the removal of the bone and the preservation of the arteries contiguous to the bone. The influence of algorithm parameters and scan parameters was studied. The method was clinically evaluated with data sets of 35 patients. Best results were obtained with a threshold of 150 HU and a dilation of 8 in-plane voxels and two out-of-plane voxels. The mean width of the soft tissue layer, which is also masked, was approximately 1 mm. The mAs value of the nonenhanced scan could be reduced from 250 mAs to 65 mAs without a loss of quality. In 32 cases the bones were registered correctly and removed completely. In three cases the bone separation was not completely successful, and consequently the bone was not completely removed. The piecewise matched mask bone elimination method proved to be able to obtain MIP images of the cervical arteries free from overprojecting bone in a fully automatic way and with only a slight increase of radiation dose.


Radiology | 2011

Image Quality and Radiation Exposure Using Different Low-Dose Scan Protocols in Dual-Source CT Coronary Angiography: Randomized Study

Lisan A. Neefjes; Anoeshka S. Dharampal; Alexia Rossi; Koen Nieman; Annick C. Weustink; Marcel L. Dijkshoorn; Gert-Jan R. ten Kate; Admir Dedic; Stella L. Papadopoulou; Marcel van Straten; Filippo Cademartiri; Gabriel P. Krestin; Pim J. de Feyter; Nico R. Mollet

PURPOSE To compare image quality, radiation dose, and their relationship with heart rate of computed tomographic (CT) coronary angiographic scan protocols by using a 128-section dual-source CT scanner. MATERIALS AND METHODS Institutional review board approved the study; all patients gave informed consent. Two hundred seventy-two patients (175 men, 97 women; mean ages, 58 and 59 years, respectively) referred for CT coronary angiography were categorized according to heart rate: less than 65 beats per minute (group A) and 65 beats per minute or greater (group B). Patients were randomized to undergo prospective high-pitch spiral scanning and narrow-window prospective sequential scanning in group A (n = 160) or wide-window prospective sequential scanning and retrospective spiral scanning in group B (n = 112). Image quality was graded (1 = nondiagnostic; 2 = artifacts present, diagnostic; 3 = no artifacts) and compared (Mann-Whitney and Student t tests). RESULTS In group A, mean image quality grade was significantly lower with high-pitch spiral versus sequential scanning (2.67 ± 0.38 [standard deviation] vs 2.86 ± 0.21; P < .001). In a subpopulation (heart rate, <55 beats per minute), mean image quality grade was similar (2.81 ± 0.30 vs 2.94 ± 0.08; P = .35). In group B, image quality grade was comparable between sequential and retrospective spiral scanning (2.81 ± 0.28 vs 2.80 ± 0.38; P = .54). Mean estimated radiation dose was significantly lower (high-pitch spiral vs sequential scanning) in group A (for 100 kV, 0.81 mSv ± 0.30 vs 2.74 mSv ± 1.14 [P < .001]; for 120 kV, 1.65 mSv ± 0.69 vs 4.21 mSv ± 1.20 [P < .001]) and in group B (sequential vs retrospective spiral scanning) (for 100 kV, 4.07 mSv ± 1.07 vs 5.54 mSv ± 1.76 [P = .02]; for 120 kV, 7.50 mSv ± 1.79 vs 9.83 mSv ± 3.49 [P = .1]). CONCLUSION A high-pitch spiral CT coronary angiographic protocol should be applied in patients with regular and low (<55 beats per minute) heart rates; a sequential protocol is preferred in all others.


Radiology | 2009

Cystic fibrosis: are volumetric ultra-low-dose expiratory CT scans sufficient for monitoring related lung disease?

Martine Loeve; Maarten H. Lequin; Marleen de Bruijne; Ieneke J. C. Hartmann; Krista Gerbrands; Marcel van Straten; Wim C. J. Hop; Harm A.W.M. Tiddens

PURPOSE To assess whether chest computed tomography (CT) scores from ultra-low-dose end-expiratory scans alone could suffice for assessment of all cystic fibrosis (CF)-related structural lung abnormalities. MATERIALS AND METHODS In this institutional review board-approved study, 20 patients with CF aged 6-20 years (eight males, 12 females) underwent low-dose end-inspiratory CT and ultra-low-dose end-expiratory CT. Informed consent was obtained. Scans were randomized and scored by using the Brody-II CT scoring system to assess bronchiectasis, airway wall thickening, mucus plugging, and opacities. Scoring was performed by two observers who were blinded to patient identity and clinical information. Mean scores were used for all analyses. Statistical analysis included assessment of intra- and interobserver variability, calculation of intraclass correlation coefficients (ICCs), and Bland-Altman plots. RESULTS Median age was 12.6 years (range, 6.3-20.3 years), median forced expiratory volume in 1 second was 100% (range, 46%-127%) of the predicted value, and median forced vital capacity was 99% (range, 61%-123%) of the predicted value. Very good agreement was observed between end-inspiratory and end-expiratory CT scores for Brody-II total score (ICC = 0.96), bronchiectasis (ICC = 0.98), airway wall thickening (ICC = 0.94), mucus plugging (ICC = 0.96), and opacities (ICC = 0.90). Intra- and interobserver agreement were good to very good (ICC range, 0.70-0.98). Bland-Altman plots showed that differences in scores were independent of score magnitude. CONCLUSION In this pilot study, CT scores from end-expiratory and end-inspiratory CT match closely, suggesting that ultra-low-dose end-expiratory CT alone may be sufficient for monitoring CF-related lung disease. This would help reduce radiation dose for a single investigation by up to 75%.


American Journal of Respiratory and Critical Care Medicine | 2014

Monitoring Cystic Fibrosis Lung Disease by Computed Tomography Radiation Risk in Perspective

Wieying Kuo; Pierluigi Ciet; Harm A.W.M. Tiddens; Wei Zhang; R. Paul Guillerman; Marcel van Straten

Computed tomography (CT) is a sensitive technique to monitor structural changes related to cystic fibrosis (CF) lung disease. It detects structural pulmonary abnormalities such as bronchiectasis and trapped air, at an early stage, before they become apparent with other diagnostic tests. Clinical decisions may be influenced by knowledge of these abnormalities. CT imaging, however, comes with risk related to ionizing radiation exposure. The aim of this review is to discuss the risk of routine CT imaging in patients with CF, using current models of radiation-induced cancer, and to put this risk in perspective with other medical and nonmedical risks. The magnitude of the risk is a complex, controversial matter. Risk analyses have largely been based on a linear no-threshold model, and excess relative and excess absolute risk estimates have been derived mainly from atomic bomb survivors. The estimates have large confidence intervals. Our risk estimates are in concordance with previously reported estimates. A large proportion of radiation to which humans are exposed is from natural background sources and varies widely depending on geographical location. The risk differences due to variation in background radiation can be larger than the risks associated with CF lung disease monitoring by CT. We conclude that the risk related to routine usage of CT in clinical care is small. In addition, a life-limiting disease, such as CF, lowers the risk of radiation-induced cancer. Nonetheless, the use of CT should always be justified and the radiation dose should be kept as low as reasonably achievable.


Medical Physics | 2003

Removal of arterial wall calcifications in CT angiography by local subtraction

Marcel van Straten; Henk W. Venema; Geert J. Streekstra; Jim A. Reekers; Gerard J. den Heeten; Cornelis A. Grimbergen

CT Angiography (CTA) is an established technique for the minimally invasive imaging of arteries. The technique of maximum intensity projection (MIP) is often used to get a comprehensive overview of the vascular anatomy. On a MIP, however, arterial wall calcifications may hinder the visualization of the arterial lumen. These calcifications are in direct contact with the contrast-enhanced blood, which makes removal difficult. We present a local subtraction method for the automatic removal of these calcifications. In our approach a second CT scan has to be made, prior to contrast injection. The calcifications in both scans are registered prior to subtraction to compensate for displacements in between the two scans. Local subtraction results are compared with results obtained by thresholding. The method was tested in a phantom and with data from four patients. The phantom represented an artery with different types of stenosis. Data were used from patients for which CTA of the renal arteries was performed. For two patients the electrocardiogram (ECG) was recorded during the CTA examination, making retrospective cardiac gated reconstructions possible. Both the phantom and the patient study showed that the local subtraction method is capable of removing calcifications and visualizing the residual lumen. In the patient study it appeared that some artifacts remained for higher pitch values. We conclude that the local subtraction method is less subjective and more accurate than thresholding. Best results are obtained by use of a small pitch, at the expense of the volume covered during a single breath hold.


European Journal of Epidemiology | 2014

Leukemia and brain tumors among children after radiation exposure from CT scans: design and methodological opportunities of the Dutch Pediatric CT Study

Johanna M. Meulepas; Cecile M. Ronckers; Anne M. J. B. Smets; Rutger A. J. Nievelstein; Andreas Jahnen; Choonsik Lee; Mariëtte Kieft; J. S. Lameris; Marcel van Herk; Marcel J. W. Greuter; Cécile R. L. P. N. Jeukens; Marcel van Straten; Otto Visser; Flora E. van Leeuwen; Michael Hauptmann

Computed tomography (CT) scans are indispensable in modern medicine; however, the spectacular rise in global use coupled with relatively high doses of ionizing radiation per examination have raised radiation protection concerns. Children are of particular concern because they are more sensitive to radiation-induced cancer compared with adults and have a long lifespan to express harmful effects which may offset clinical benefits of performing a scan. This paper describes the design and methodology of a nationwide study, the Dutch Pediatric CT Study, regarding risk of leukemia and brain tumors in children after radiation exposure from CT scans. It is a retrospective record-linkage cohort study with an expected number of 100,000 children who received at least one electronically archived CT scan covering the calendar period since the introduction of digital archiving until 2012. Information on all archived CT scans of these children will be obtained, including date of examination, scanned body part and radiologist’s report, as well as the machine settings required for organ dose estimation. We will obtain cancer incidence by record linkage with external databases. In this article, we describe several approaches to the collection of data on archived CT scans, the estimation of radiation doses and the assessment of confounding. The proposed approaches provide useful strategies for data collection and confounder assessment for general retrospective record-linkage studies, particular those using hospital databases on radiological procedures for the assessment of exposure to ionizing or non-ionizing radiation.

Collaboration


Dive into the Marcel van Straten's collaboration.

Top Co-Authors

Avatar

Gabriel P. Krestin

Erasmus University Rotterdam

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Marcel L. Dijkshoorn

Erasmus University Rotterdam

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Harm A.W.M. Tiddens

Erasmus University Rotterdam

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Nico R. Mollet

Erasmus University Rotterdam

View shared research outputs
Top Co-Authors

Avatar

Pim J. de Feyter

Erasmus University Rotterdam

View shared research outputs
Top Co-Authors

Avatar

Annick C. Weustink

Erasmus University Rotterdam

View shared research outputs
Top Co-Authors

Avatar
Researchain Logo
Decentralizing Knowledge